Evaluation of different DNA extraction methods and loop-mediated isothermal amplification primers for the detection of Mycobacterium ulcerans in clinical specimens

Abstract

Background

Early diagnosis and treatment of Buruli ulcer is critical in order to avoid the debilitating effects of the disease. In this regard, the development of new diagnostic and point of care tools is encouraged. The loop-mediated isothermal amplification for the detection of Mycobacterium ulcerans represents one of the new tools with a good potential of being developed into a point of care test. There is however the need to standardize the assays, reduce sample preparation times, improve the detection/visualization system and optimize them for high-throughput screening, adaptable to low resourced laboratories.

Methods

In this study, we assessed two DNA extraction protocols (modified Boom and EasyNAT methods), three previously published LAMP primer sets (BURULI, MU 2404 and BU-LAMP), and compared the sensitivity and specificity of LAMP assays on three DNA amplification platforms.

Results

Our results show that Buruli ulcer diagnosis using primers targeting IS2404 for the LAMP method is sensitive (73.75–91.49%), depending on the DNA extraction method used. Even though the modified Boom DNA extraction method provided the best results, its instrumentation requirement prevent it from being field applicable. The EasyNAT method on the other hand is simpler and may represent the best method for DNA extraction in less resourced settings.

Conclusions

For further work on the development and use of LAMP tests for Buruli diagnosis, it is recommended that the BURULI sets of primers be used, as these yielded the best results in terms of sensitivity (87.50–91.49%) and specificity (89.23–100%), depending on the DNA extraction methods used.

Peer Review reports

Background

Buruli ulcer (BU) caused by Mycobacterium ulcerans is a necrotizing skin disease endemic in several rural communities in sub-Saharan Africa, particularly in Cote d’Ivoire, Ghana and Benin. There are also other foci of BU in the Americas, as well as in Asia and Oceania, with a very sharp recent increase in the number of cases in Australia in recent years. BU is the third most prevalent mycobacterial disease after tuberculosis [1,2,3]. Due to poor understanding of its epidemiology and transmission, there are no effective primary preventive measures for control of the disease. Control efforts focus on early case detection and prompt treatment to stop progression to the ulcerative stages, and thus avoid the associated disabilities [4].

Absence of effective diagnostic services accounts for the poor surveillance of the disease [5, 6]. In rural areas where majority of cases are detected, samples are often sent to reference laboratories for confirmatory testing, which takes 1–2 weeks. District hospitals rely on clinical diagnosis, but this is often inaccurate and complicated by other disease conditions with similar presentations, leading to misdiagnosis. Microscopic examination detects 29–78% of clinically suspected BU cases and is currently the only rapid and affordable test available for BU diagnosis in many endemic areas [6]. The detection rate of culture is between 34 and 79% and takes an average of 9–12 weeks to yield positive results but this too is only available in reference labs [6]. Culture therefore cannot be used for rapid laboratory confirmation. Histopathological examinations reportedly detect 30% additional cases than other confirmatory tests, however these techniques are restricted to external reference laboratories and are unavailable in peripheral health centres, district or regional hospitals [6]. These challenges; including poor conditions for sample transportation, and long turnaround times, contribute to late treatment and management of the disease. The WHO’s Second International Conference on Buruli Ulcer Control and Research resolved to strengthen the capacity of national laboratories to confirm cases of the disease but warned that “efforts are still needed to develop simple diagnostic tools usable in the field as well as disability prevention methods” [7]. Therefore, new field applicable techniques for the diagnosis of BU, paralleling PCR in sensitivity and specificity, need to be developed [2, 6].

IS2404 PCR has close to 96% sensitivity and has been recommended by WHO for laboratory confirmation of BU [8,9,10,11,12,13]. However, PCR is labour intensive, expensive and unsuited for use in rural African settings with poorly resourced labs [14]. Isothermal amplification of nucleic acids in recent times has been relied upon to circumvent the limitations of traditional PCR and provides a platform that can support the development of rapid Point of Care (POC) molecular diagnostic tests [15]. Loop-mediated isothermal amplification (LAMP) is a method that amplifies DNA with high specificity, sensitivity and rapidity at a constant temperature. The method utilizes the strand displacement ability of the Bacillus stearothermophilus DNA polymerase (Bst polymerase) to loop 4–6 primer pairs, including inner and outer primers, to the target loci of interest. Thus, it yields several stem-loop amplicons, which sustains the reaction cycling at 60–65 °C. The main advantage of LAMP over PCR is that it does not require thermal cyclers and gel imaging equipment, and thus, amplification can be performed within an hour or less in a laboratory with a heat block or water bath [8, 16]. LAMP reaction results can be visualised with the naked eye from colorimetric changes following addition of intercalating fluorescent dyes such as SYBR green or hydroxynapthol blue (HNB) to reaction tubes, pre- or post-reaction [17]. LAMP is also less inhibited by contaminants within the sample and the closed tube approach reduces problems of cross contamination. It has been widely developed for detection of various disease pathogens, including viral, bacterial and protozoan [18,19,20,21,22]. The superiority of LAMP, cost-effectiveness and ease of use by potential ‘unskilled’ personnel makes it an attractive molecular diagnostic tool to further develop for clinical use, particularly in resource-limited settings where PCR cannot be afforded.

Previous studies developed LAMP assays for the detection of Mycobacterium ulcerans in clinical samples [8, 16, 23]. Separately, two of the primers (Table 1) amplified fragments of the multi-copy insertion sequence, IS2404, of which 214 copies are present in the genome of the bacterium [8, 23], while the other amplified another insertion sequence, IS2606, with about 60 genomic copies present [16]. The results showed superior sensitivity of LAMP to PCR [8, 16, 23]. However, the sample sizes used in these investigations were small and different methods were used, which did not allow accurate comparisons of the data. Hence, there was a need to standardize the assays, reduce sample preparation times, improve the detection/visualization system, and optimize them for high-throughput screening, adaptable to low resourced laboratories. To reduce the turnaround times, we utilized and compared two rapid DNA extraction protocols for sample preparation. Using several clinical samples from endemic communities, we compared the sensitivity and specificity of LAMP assays on three DNA amplification platforms. Three recommended fluorescent dyes were also optimized and used to visualize test results. We present comparative genomic DNA extraction and primer-specific LAMP data that provide suitable alternatives with remarkable sensitivity for rapid diagnosis of BU.

Methods

Work flow

The flow diagram of the study is shown in Fig. 1.

Fig. 1
figure 1

Flow diagram of study procedure showing assays performed and the total number of samples tested

Samples

The samples used in this study were archived as well as fresh patient’s samples. Fresh samples were collected from consenting patients attending Buruli ulcer Clinic at selected district hospitals and health centres in Ghana, using the Fine Needle Aspiration (FNA) and Swabs methods [24, 25]. The Noguchi Memorial Institute for Medical Research (NMIMR) is one of the two reference centres in Ghana where patients’ samples are sent for confirmation. As such, these samples were also included in the study. Swabs and FNAs were respectively put into sterile 15 ml falcon tubes and phosphate buffered saline, and transported in cool box to NMIMR.

DNA extractions

Previous results indicated that crude DNA preparations yield low sensitivity compared with purified extracts in the LAMP assay [8]. Consequently, crude DNA extracts were not included in this work. In order to simplify DNA extraction for the LAMP test in field settings, two simple and rapid extraction protocols were used; a commercially available syringe silica membrane method (EasyNAT) for the extraction of M. tuberculosis DNA from sputum, and a modification of the syringe silica method (modified EasyNAT).

Modified Boom extraction

Portions (250 μl) of the sample suspension were lysed with 250 μl of lysis buffer (1.6 M guanidinum HCl, 60 mM Tris (pH 7.4), 1% Triton X-100, 60 mM EDTA, 10% Tween-20), 50 μl of Proteinase K (20 mg/ml) and 500 μl glass beads in a sterile 1.5 ml tube. The mixture was incubated horizontally in an orbital shaker (200 rpm) at 60 °C for 12 h. To capture the DNA, 40 μl of diatomaceous earth solution was added to the lysed sample and incubated at 37 °C for 1 h. The mixture was then centrifuged at 14000 rpm for 1 min and the supernatant discarded. Pelleted DNA bound to the diatomaceous earth was washed twice with 900 μl of ice cold 70% ethanol and 900 μl of absolute acetone, pulsed vortexed and centrifuged at 14000 rpm for 60s. The supernatant was discarded and DNA bound to diatomaceous earth was dried at 50 °C for 20 min using a heat block. DNA was re-suspended in 80 μl of nuclease free water, incubated at 68 °C for 20 min to release the DNA from the diatomaceous earth. The DNA extract was centrifuged at 14000 rpm for 2 min and about 60 μl of the resultant supernatant transferred to a sterile tube and kept at -20 °C. Negative and positive controls were included in each batch of DNA extraction. The negative control included all the reagents for DNA extraction, but excluding a clinical specimen, while the positive control included 250 μl of M. ulcerans culture in suspension.

Genomic DNA extraction using disposable silica syringe DNA extraction device (EasyNAT)

DNA was extracted from suspension of clinical samples following the instructions provided by the manufacturer (USTAR Biotechnologies Ltd., Hangzhou, China). Briefly, 500 μl of clinical specimen was pipetted into 1.5 mL tube, then 600 μl of lysis buffer and 50 μl Proteinase K were added. The mixture was incubated at 60 °C for 30 min and allowed to cool to room temperature. 1400 μl of absolute ethanol was added to the lysed clinical specimen, and the tube was inverted 3–4 times. The mixture was aspirated through the nozzle of an assembled silica syringe device (SSD), with the aid of the attached plunger. The lysed sample was slowly sucked back into the tube, leaving the DNA bound to the silica membrane at the base of the syringe. The DNA bound to the silica membrane was washed twice by sucking 60% of absolute ethanol through the nozzle of the syringe and slowly expelled. At this step, the DNA remained bound to the silica membrane. The 5 ml syringe barrel was detached from the SSD and replaced with a 1 ml syringe barrel (included in the kit). 100 μl of nuclease free water was sucked and allowed to soak the silica membrane for 2 min. The dissolved DNA was eluted into a 1.5 ml tube.

Genomic DNA extraction using a modified EasyNAT protocol

Suspension of clinical sample (150 μl) was lysed by adding 600 μl of lysis buffer (1.6 M guanidinum HCl, 60mMTris (pH 7.4), 1% Triton X-100, 60 mM EDTA, 10% Tween20) and 50 μl Proteinase K (20 mg/ml). The mixture was incubated at 60 °C for 30 min and allowed to cool to room temperature. Absolute ethanol (1400 μl) was added to the lysed sample, vortexed and sucked through the assembled silica syringe device (SSD) with the aid of the plunger. The lysed sample was then filtered back to the tube by pushing the plunger slowly. DNA bound to the silica membrane at the base of the SSD was washed twice with 2 ml of 60% alcohol using air pressure generated from pushing the plunger. The 5 ml syringe barrel was detached from the SSD and replaced with 1 ml syringe barrel for DNA elution. Nuclease free water (100 μl) was then sucked through the nozzle of the syringe. With the aid of the plunger, nuclease free water was sucked over the silica membrane 3–4 times and finally eluted in a 1.5 ml eppendorf tube.

Nucleic acid amplifications

All DNA samples were tested with IS2404 PCR reference method [13]. The primer sequences for the IS2404 PCR is shown in Table 1. All published LAMP primer sets (Table 2) were tested on the different DNA extracts, in order to determine their sensitivity. LAMP assays were performed using the Loopamp DNA amplification kit (Eiken Chemical). For specificity, the primers were also tested on M. nonchromogenicum, M. fluoranthenivorans, M. fortuitum, M. porcinum, M. mucogenicum, M. ulcerans, and M. tuberculosis.

Table 1 IS2404 nested PCR primers
Table 2 LAMP primers from the three different studies

Preparation of suspensions containing different copy numbers of IS2404 for determination of the detection limit of the LAMP tests

A single colony of M. ulcerans growing on Lowenstein-Jensen media slant was harvested in nuclease free water. DNA was extracted from this culture using the QIAGEN DNeasy Blood & Tissue Kits (QIAGEN, California). The concentration of the final extract was 0.2 ng/μl (stock concentration in a final volume of 20 μl), which was equivalent to 300,000 copies of IS2404. The latter was serially diluted in a volume of 20 μl to contain 30,000, 3000, 300, 30 and 3 copies of IS2404. LAMP assays were performed on these dilutions to measure the detection limit of the three primer sets.

Optimization of visualization system for M. ulcerans

The detection of LAMP amplicons can utilize several intercalating fluorescent dyes [17], which suggests that an optimum dye concentration will increase the detection limit. In this study, we also compared two dyes; Calcein and HNB using protocols described elsewhere [17, 26]. This was done in closed tubes to prevent possible cross-contamination. Calcein turns from yellow (negative test) to fluorescent green (positive test) and hydroxynapthol blue turns from violet (negative test) to sky blue (positive test).

Data analysis

For data analysis, 2 × 2 contingency tables were generated for the evaluation of diagnostic accuracy. The data was analysed for diagnostic test evaluation using MedCalc Software (Version 18.6). The disease prevalence, sensitivity, specificity, positive and negative predictive values as well as accuracy, were expressed as percentages, with their corresponding 95% confidence intervals.

Results

Comparison of DNA extraction protocols

Results of IS2404 PCR performed on the three different DNA extracts is shown in Table 3 below. A total of 136 extracts were analysed. The modified Boom DNA extraction method gave the highest PCR positivity (69.12%), followed by the EasyNAT DNA extraction method (58.82%).

Table 3 PCR analysis of 136 clinical samples. Values in brackets represent the 95% CI

Sensitivity and specificity of LAMP assays in combination with different DNA extracts

The results of the various tests were compared for their sensitivity, specificity and predictive values and accuracy (Table 4). For this, the IS2404 PCR was used as the reference standard. For a particular LAMP assay and with a particular DNA extract, more positives were detected with the BURULI primer set designed by Ablordey [8], followed by the MU 2404 primers of Njiru [23] and then MU LAMP primers of de Souza [16] (Table 4). Specificity however did not follow a similar pattern. Across corresponding tests, the LAMP test on Boom extracts yielded the most number of positives followed by EasyNAT and then the modified EasyNAT extracts. However positivity of the LAMP test for modified and unmodified extracts where the same. For the modified Boom, EasyNAT and modified DNA extracts, the BURULI primers yielded sensitivities of 91.49, 87.50 and 90.14%, and specificities of 97.26, 100.00 and 89.23% respectively. The MU 2404 primers yielded sensitivities of 86.12, 73.75, 83.10% and specificities of 88.10, 98.21 and 93.85% respectively. The BU-LAMP primers on the other hand yielded the lowest sensitivities of 57.45, 47.50, 35.21% and highest specificities of 100.00, 98.21 and 95.38% respectively. Thus, comparing the trade off between sensitivity and specificity, coupled with the predictive values and the accuracies, the BURULI primers gave the best results.

Table 4 Comparison of sensitivity and specificity of three LAMP tests based on calcein visualization with different DNA extracts. Values in brackets represent the 95% CI

Across corresponding tests, sensitivity was highest for the LAMP test on Boom extracts (57.45–91.49%), followed by the modified EasyNAT (35.21–90.14%) and then unmodified EasyNAT extracts (47.50–87.50%). Specificity did not follow a specific pattern, but the values were high (88.1–100%). The results also showed the EasyNAT method to have a high specificity (98.2–100%) i.e. the probability of screening negative if the disease is truly absent. On the other hand, the modified EasyNAT method had the lowest sensitivities (35.21–90.14%) i.e. the probability of screening positive if the disease is truly present. The predictive values of the various tests were also estimated. The positive predictive value, which estimates the probability that an individual actually has the disease given a positive test, varied according to the methods.

The detection limit of the LAMP test was 30 copies of IS2404. However half and a quarter of the tests were positive for BURULI and MU 2404 primers respectively (Table 5). Concerning the specificity of the tests, the results revealed that, of the eight Mycobacterium species tested with the three sets of primers, only M. ulcerans isolates tested positive in the LAMP test.

Table 5 Detection limit of the LAMP tests with DNA extracted using QIAGEN, and visualized based on calcein detection system

Comparison of fluorescence detection dyes

The optimization of a fluorescent dye system was undertaken using the BURULI and MU 2404 primers. For a given test (BURULI or MU 2404), no significant difference was found between tests performed using calcein or HNB. (Table 6).

Table 6 Comparison of LAMP reactions using two different dyes, on DNA extracted using QIAGEN. Values in brackets represent the 95% CI

Discussions

This study revealed that the efficiency of the LAMP reaction depends on the quality of extracted DNA and the efficiency of the primers used. We compared the sensitivities and specificities of the various tests, as well as their predictive values, using IS2404 PCR as the gold standard. In deciding upon the criteria for acceptable levels of specificity and sensitivity, it is important to consider the consequences of leaving cases undetected (false negatives) against the false classification of healthy individuals as infected (false positives) [27]. In the context of BU diagnosis, in the absence of other options, the sensitivity of a diagnostic method should be increased at the expense of specificity, since the penalty associated with missing a case is high. Our results indicated that the modified Boom extraction method had the highest sensitivities. Thus, it is a good method for use in reference laboratories. However, the appropriateness of this method for confirmation of Buruli ulcer cases in resource limited settings should be cautiously explored, due to the requirements associated with it. It is not very practical for use in resource limited settings as it requires complex, lengthy procedures and sophisticated equipment. Thus, in terms of programmatic feasibility, testing requirements, time and cost, the modified EasyNAT method appears to be the best for field monitoring and evaluation programs, with sensitivities of 90.14 and 83.10% for the BURULI and MU 2404 primers respectively. These results also showed the BURULI to be best suited for analysis, since these gave the highest sensitivities, best predictive values and accuracies.

The aforementioned primers are both based on IS2404, while BU-LAMP primers are based on IS2606. PCR assays based on IS2404 and IS2606 revealed that IS2606 PCR was 10 times less sensitive, compared to IS2404 PCR [13]. The copy numbers for the IS2404 and IS2606 target sequences in the genome of M. ulcerans strain Agy99 are 213 and 91 respectively [28], representing a little over twice as many copies of IS2404 compared to IS2606. These differences in copy numbers may explain why the IS2404 based primers resulted in a higher sensitivity.

Conclusions

The need for confirmation of Buruli ulcer cases requires tools that are sensitive. The results of this work would permit the development of improved tools for the diagnosis for Buruli ulcer. These assays, if adapted for field settings and poorly resourced laboratories, have the robustness and good turnaround times, to be used by ‘semi-skilled’ lab technicians with little supervision. For the development and use of future tests for Buruli diagnosis using the LAMP method, it is recommended that the BURULI primers [8] be used.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

DNA:

Deoxyribonucleic acid

LAMP:

Loop mediated isothermal amplification

MU:

Mycobacterium ulcerans

BU:

Buruli ulcer

PCR:

Polymerase chain reaction

SSD:

Silica syringe device

HNB:

Hydroxynapthol blue

EDTA:

Ethylenediaminetetraacetic acid

NMIMR:

Noguchi memorial institute for medical research

References

  1. World Health Organization. Buruli ulcer disease: Mycobacterium ulcerans infection: an overview of reported cases globally. Wkly Epidemiol Rec. 2004;79:194–9 https://apps.who.int/iris/handle/10665/232478.

    Google Scholar 

  2. World Health Organization. Buruli ulcer: progress report, 2004-2008. Wkly Epidemiol Rec. 2008;83:145–54 https://apps.who.int/iris/handle/10665/241134.

    Google Scholar 

  3. Johnson PDR. Buruli Ulcer in Australia. In: Pluschke G, Röltgen K, editors. Buruli Ulcer: Mycobacterium Ulcerans Disease. Cham: Springer International Publishing; 2019. p. 61–76. https://doi.org/10.1007/978-3-030-11114-4_3.

    Chapter  Google Scholar 

  4. World Health Organization. Buruli ulcer (Mycobacterium ulcerans infection). 2019. https://www.who.int/news-room/fact-sheets/detail/buruli-ulcer-(mycobacterium-ulcerans-infection). Accessed 8 Oct 2020.

  5. Sizaire V, Nackers F, Comte E, Portaels F. Mycobacterium ulcerans infection: control, diagnosis, and treatment. Lancet Infect Dis. 2006;6(5):288–96. https://doi.org/10.1016/S1473-3099(06)70464-9.

    Article  PubMed  Google Scholar 

  6. Beissner M, Herbinger K-H, Bretzel G. Laboratory diagnosis of Buruli ulcer disease. Future Microbiol. 2010;5:363–70. https://doi.org/10.2217/fmb.10.3.

    Article  CAS  PubMed  Google Scholar 

  7. World Health Organization. Cotonou declaration on buruli ulcer. Geneva PP - Geneva: World Health Organization; 2009. https://apps.who.int/iris/handle/10665/329410

    Google Scholar 

  8. Ablordey A, Amissah DA, Aboagye IF, Hatano B, Yamazaki T, Sata T, Ishikawa K, Katano H Detection of Mycobacterium ulcerans by the loop mediated isothermal amplification method. PLoS Negl Trop Dis 2012;6:e1590. https://doi.org/https://doi.org/10.1371/journal.pntd.0001590, 4.

  9. Asiedu K, Wansbrough-Jones M. Mycobacterium ulcerans infection (Buruli or Bairnsdale ulcer): challenges in developing management strategies. Med J Aust. 2007;186:55–6 http://europepmc.org/abstract/MED/17223762.

    Article  Google Scholar 

  10. Bretzel G, Huber KL, Kobara B, Beissner M, Piten E, Herbinger K-H, et al. Laboratory confirmation of Buruli ulcer disease in Togo, 2007-2010. PLoS Negl Trop Dis. 2011;5(7):e1228. https://doi.org/10.1371/journal.pntd.0001228.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Johnson PDR, Azuolas J, Lavender CJ, Wishart E, Stinear TP, Hayman JA, et al. Mycobacterium ulcerans in mosquitoes captured during outbreak of Buruli ulcer, southeastern Australia. Emerg Infect Dis. 2007;13(11):1653–60. https://doi.org/10.3201/eid1311.061369.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Ross BC, Marino L, Oppedisano F, Edwards R, Robins-Browne RM, Johnson PD. Development of a PCR assay for rapid diagnosis of Mycobacterium ulcerans infection. J Clin Microbiol. 1997;35(7):1696–700. https://doi.org/10.1128/JCM.35.7.1696-1700.1997.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Stinear T, Ross BC, Davies JK, Marino L, Robins-Browne RM, Oppedisano F, et al. Identification and characterization of IS2404 and IS2606: two distinct repeated sequences for detection of Mycobacterium ulcerans by PCR. J Clin Microbiol. 1999;37(4):1018–23. https://doi.org/10.1128/JCM.37.4.1018-1023.1999.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Iwamoto T, Sonobe T, Hayashi K. Loop-mediated isothermal amplification for direct detection of Mycobacterium tuberculosis complex, M. avium, and M. intracellulare in sputum samples. J Clin Microbiol. 2003;41(6):2616–22. https://doi.org/10.1128/jcm.41.6.2616-2622.2003.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Notomi T, Okayama H, Masubuchi H, Yonekawa T, Watanabe K, Amino N, et al. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 2000;28(12) 63e, 663:E63. https://doi.org/10.1093/nar/28.12.e63.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. de Souza DK, Quaye C, Mosi L, Addo P, Boakye DA. A quick and cost effective method for the diagnosis of Mycobacterium ulcerans infection. BMC Infect Dis. 2012;12(1):8. https://doi.org/10.1186/1471-2334-12-8.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Goto M, Honda E, Ogura A, Nomoto A, Hanaki K-I. Colorimetric detection of loop-mediated isothermal amplification reaction by using hydroxy naphthol blue. Biotechniques. 2009;46(3):167–72. https://doi.org/10.2144/000113072.

    Article  CAS  PubMed  Google Scholar 

  18. Aonuma H, Yoshimura A, Perera N, Shinzawa N, Bando H, Oshiro S, et al. Loop-mediated isothermal amplification applied to filarial parasites detection in the mosquito vectors: Dirofilaria immitis as a study model. Parasit Vectors. 2009;2(1):15. https://doi.org/10.1186/1756-3305-2-15.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Njiru ZK, Mikosza ASJ, Matovu E, Enyaru JCK, Ouma JO, Kibona SN, et al. African trypanosomiasis: sensitive and rapid detection of the sub-genus Trypanozoon by loop-mediated isothermal amplification (LAMP) of parasite DNA. Int J Parasitol. 2008;38(5):589–99. https://doi.org/10.1016/j.ijpara.2007.09.006.

    Article  CAS  PubMed  Google Scholar 

  20. Pandey BD, Poudel A, Yoda T, Tamaru A, Oda N, Fukushima Y, et al. Development of an in-house loop-mediated isothermal amplification (LAMP) assay for detection of Mycobacterium tuberculosis and evaluation in sputum samples of Nepalese patients. J Med Microbiol. 2008;57(Pt 4):439–43. https://doi.org/10.1099/jmm.0.47499-0.

    Article  CAS  PubMed  Google Scholar 

  21. Parida M, Posadas G, Inoue S, Hasebe F, Morita K. Real-time reverse transcription loop-mediated isothermal amplification for rapid detection of West Nile virus. J Clin Microbiol. 2004;42(1):257–63. https://doi.org/10.1128/jcm.42.1.257-263.2004.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Poon LLM, Wong BWY, Ma EHT, Chan KH, Chow LMC, Abeyewickreme W, et al. Sensitive and inexpensive molecular test for falciparum malaria: detecting plasmodium falciparum DNA directly from heat-treated blood by loop-mediated isothermal amplification. Clin Chem. 2006;52(2):303–6. https://doi.org/10.1373/clinchem.2005.057901.

    Article  CAS  PubMed  Google Scholar 

  23. Njiru ZK, Yeboah-Manu D, Stinear TP, Fyfe JAM. Rapid and sensitive detection of Mycobacterium ulcerans by use of a loop-mediated isothermal amplification test. J Clin Microbiol. 2012;50(5):1737–41. https://doi.org/10.1128/JCM.06460-11.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Eddyani M, Fraga AG, Schmitt F, Uwizeye C, Fissette K, Johnson C, et al. Fine-needle aspiration, an efficient sampling technique for bacteriological diagnosis of nonulcerative Buruli ulcer. J Clin Microbiol. 2009;47(6):1700–4. https://doi.org/10.1128/JCM.00197-09.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Yeboah-Manu D, Danso E, Ampah K, Asante-Poku A, Nakobu Z, Pluschke G. Isolation of <span class="named-content genus-species" id="named-content-1">Mycobacterium ulcerans</span> from swab and fine-needle-aspiration specimens. J Clin Microbiol. 2011;49:1997 LP – 1999. doi:https://doi.org/10.1128/JCM.02279-10.

  26. Suebsing R, Kampeera J, Sirithammajak S, Withyachumnarnkul B, Turner W, Kiatpathomchai W. Colorimetric method of loop-mediated isothermal amplification with the pre-addition of Calcein for detecting Flavobacterium columnare and its assessment in Tilapia farms. J Aquat Anim Health. 2015;27(1):38–44. https://doi.org/10.1080/08997659.2014.966212.

    Article  CAS  PubMed  Google Scholar 

  27. Hennekens CH, Buring JE, Mayrent SL. Epidemiology in medicine. Boston, Massachusetts: Little, Brown; 1987.

    Google Scholar 

  28. Stinear TP, Seemann T, Pidot S, Frigui W, Reysset G, Garnier T, et al. Reductive evolution and niche adaptation inferred from the genome of Mycobacterium ulcerans, the causative agent of Buruli ulcer. Genome Res. 2007;17(2):192–200. https://doi.org/10.1101/gr.5942807.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The Authors will like to thank the Medical Doctors and Nurses for taking samples from Patients. We also thank the Patients for consenting to participate in the study.

Funding

This study was supported by the Swiss Agency for Development and Cooperation and the UBS Optimus Foundation through FIND, Geneva-Switzerland. The funders had no role in the design of the study and collection, analysis, and interpretation of data and in writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

Methodology, formal analysis, and writing—original draft preparation, review and editing, AA, EA, CAN, SAK, DKD; investigation and data creation, AA, DKD; conceptualization, project administration, and supervision, JMN, IC. All authors have read and approved the manuscript.

Corresponding author

Correspondence to Anthony Ablordey.

Ethics declarations

Ethics approval and consent to participate

This study was carried out at the Noguchi Memorial Institute for Medical Research (NMIMR), University of Ghana. Clinical samples were taken at the hospital or health centres as part of the routine diagnostic procedure and case confirmation recommended by WHO, and sent to the NMIMR for analysis. Ethical approval for collection and use of patients’ samples was obtained from the Institutional Review Board of the NMIMR (Study number: NMIMR-IRB CPN 024/18–19). Patients were required to sign an informed consent. To ensure anonymity, personal information such as names, and other personal identifiers associated with patient’s samples were replaced with codes.

Consent for publication

Not applicable

Competing interests

The authors declare no conflict of interest.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ablordey, A., Ahotor, E., Narh, C.A. et al. Evaluation of different DNA extraction methods and loop-mediated isothermal amplification primers for the detection of Mycobacterium ulcerans in clinical specimens. BMC Infect Dis 21, 598 (2021). https://doi.org/10.1186/s12879-021-06308-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12879-021-06308-z

Keywords