RLEP LAMP for the laboratory confirmation of leprosy: towards a point-of-care test

Abstract

Background

Nucleic acid-based amplification tests (NAAT), above all (q)PCR, have been applied for the detection of Mycobacterium leprae in leprosy cases and household contacts with subclinical infection. However, their application in the field poses a range of technical challenges. Loop-mediated isothermal amplification (LAMP), as a promising point-of-care NAAT does not require sophisticated laboratory equipment, is easy to perform, and is applicable for decentralized diagnosis at the primary health care level. Among a range of gene targets, the M. leprae specific repetitive element RLEP is regarded as highly sensitive and specific for diagnostic applications. 

Methods

Our group developed and validated a dry-reagent-based (DRB) RLEP LAMP, provided product specifications for customization of a ready-to-use kit (intended for commercial production) and compared it against the in-house prototype. The assays were optimized for application on a Genie® III portable fluorometer. For technical validation, 40 “must not detect RLEP” samples derived from RLEP qPCR negative exposed and non-exposed individuals, as well as from patients with other conditions and a set of closely related mycobacterial cultures, were tested together with 25 “must detect RLEP” samples derived from qPCR confirmed leprosy patients. For clinical validation, 150 RLEP qPCR tested samples were analyzed, consisting of the following categories: high-positive samples of multibacillary (MB) leprosy patients (> 10.000 bacilli/extract), medium-positive samples of MB leprosy patients (1.001–10.000 bacilli/extract), low-positive samples of MB leprosy patients (1–1.000 bacilli/extract), endemic controls and healthy non-exposed controls; each n = 30. 

Results

Technical validation: both LAMP formats had a limit of detection of 1.000 RLEP copies, i.e. 43–27 bacilli, a sensitivity of 92% (in-house protocol)/100% (ready-to-use protocol) and a specificity of 100%. Reagents were stable for at least 1 year at 22 °C. Clinical validation: Both formats showed a negativity rate of 100% and a positivity rate of 100% for high-positive samples and 93–100% for medium positive samples, together with a positive predictive value of 100% and semi-quantitative results. The positivity rate for low-positive samples was 77% (in-house protocol)/43% (ready-to-use protocol) and differed significantly between both formats. 

Conclusions

The ready-to-use RLEP DRB LAMP assay constitutes an ASSURED test ready for field-based evaluation trials aiming for routine diagnosis of leprosy at the primary health care level.

Peer Review reports

Background

Leprosy caused by Mycobacterium leprae is a neglected, chronic infectious disease predominantly affecting the skin and peripheral nerves which is transmitted through the aerial route [1, 2].

The disease is spectral and categorized according to the Ridley–Jopling classification based on the type of lesions and the bacterial load into indeterminate, tuberculoid, lepromatous and three borderline forms. Alternatively, a simplified, field-based classification introduced by the WHO distinguishes paucibacillary (PB, up to five lesions and/or only one nerve trunk involved) and multibacillary (MB, more than five lesions and/or more than one nerve trunk involved) forms. The clinical diagnosis of leprosy can be challenging and the simplified, field-based WHO classification is widely used without applying bacteriological analysis [3]. In the absence of bacteriological evidence misclassification and misdiagnosis leading to inappropriate treatment can occur [4, 5].

To bring back the laboratory diagnostic component into routine practice [5], beyond acid-fast bacilli (AFB) microscopy with its limitations regarding expertise, sensitivity, and specificity [5,6,7], a range of PCR-based molecular diagnostic tests targeting various gene markers have been applied to the diagnosis of leprosy from skin biopsies and slit skin smears with a positivity rate approaching 50% in PB and 80% in MB cases [5].

Real-time qPCR is considered at least 20 times more sensitive than microscopy [8] and among a range of possible gene targets, the M. leprae specific repetitive element RLEP is regarded as highly sensitive, 100% specific and therefore a suitable target for diagnostic applications [5, 6, 9,10,11,12].

RLEP belongs to a family of dispersed repeats with unknown function in the genome of M. leprae, namely: LEPRPT, LEPREP, REPLEP and RLEP (in ascending order of copy numbers) [13, 14]. Four slightly different RLEP sequences were published: RLEP 1, RLEP 2, RLEP 3 and RLEP 4 (GenBank [pubmed, NCBI], accession numbers FM211192.1, X17151.1, X17153.1 and X17152.1), with RLEP 1 showing most nucleotide differences among these sequences. An M. leprae specific central portion (545 bp [15] to 488 bp [13]) is common to all copies of RLEP, and flanked by variable additional sequences (44–100 bp [15] to 113–587 bp [13]). Published copy numbers vary between 28 copies (cp) [15] and 37 cp [13], according to mutations in the primer binding sites. For the genome of Togolese strains as used in this study, the copy number was averaged to 30 cp, determined through two different qPCRs detecting RLEP and 16S rRNA using logarithmic dilutions of plasmid standards as published by Beissner et al. [16]. For one patient with Pakistani origin but living for decades in Munich, Germany, diagnosed and treated at DITM (Division of Infectious Diseases and Tropical Medicine, Munich), the amplifiable RLEP copy number was 23 cp [17].

Our group developed an RLEP qPCR on nasal swabs constituting a noninvasive sampling technique facilitating early diagnosis of leprosy cases as well as laboratory assessment of contact persons [16]. Nasal swabs have been effectively applied for PCR-based diagnosis and are commonly used to investigate nasal carriage leading to subclinical infection of M. leprae in household contacts (HHC) [12, 18,19,20,21,22,23,24,25,26].

Although recognizing the superiority of PCR compared to other diagnostic techniques, previous WHO recommendations did not include PCR for laboratory confirmation of cases or screening of contacts in the field as PCR assays would be difficult to perform in field settings, lack standardization, are not commercially available and require technical and laboratory expertise. Recent policy developments however, such as the scheduled implementation of laboratory diagnostics for the period of 2026–2030 in the upcoming WHO Global Leprosy Strategy 2021–2030, the appraisal of the necessity for effective and affordable diagnostic tests as expressed by the Technical Advisory Group (TAG) and its support for improvement of laboratory capacity including molecular diagnostics, as well as the recent assessment of research priorities by the Global Partnership for Zero Leprosy giving top priority to the development of diagnostic tests constitute a new phase for the role of molecular biology for the diagnosis of leprosy [1, 27,28,29].

Loop-mediated isothermal amplification (LAMP), a promising nucleic acid-based (NAAT) point-of-care (POC) technology applicable for decentralized diagnosis at the primary health care level, has the potential to change the situation. To overcome PCR limitations LAMP was developed as a field-friendly and cost-effective diagnostic tool providing advantages such as no requirements for sophisticated laboratory equipment and skills, the potential for simplified DNA extraction methods, naked-eye detection of amplicons, and the use of (freeze-) dried reagents [30].

Several studies have shown LAMP as a useful tool for the rapid detection of bacterial, parasitic and viral pathogenic agents of infectious diseases including assays for 16 neglected tropical diseases such as leishmaniasis, schistosomiasis, Buruli ulcer, helminthic diseases or yaws [31,32,33,34,35,36,37,38,39]. Besides a range of in-house assays, commercial LAMP diagnostic kits developed by Eiken (Eiken Chemical Co., Tokyo, Japan) for human African trypanosomiasis (HAT) [40], tuberculosis (TB) [41], malaria [42,43,44], and leishmaniasis [45, 46], as well as the centrifugation-free Illumigene assay for malaria genus-level detection developed by Meridian Biosciences (Cincinnati, USA) are available [47, 48].

The LAMP technique employs a Bacillus stearothermophilus (Bst) derived DNA polymerase and a set of four to six primers (two inner and two outer primers at minimum, and to speed up the reaction and augment sensitivity, if designable, a pair of loop-primers) that recognize six distinct sequences of the target DNA, which makes them highly specific to the target. The LAMP primers can be designed via the user-friendly online platform Primer Explorer V4 software [49] by Eiken. LAMP amplifies a few copies of DNA to 109 cp in less than 30 min with high efficiency [30]. The mechanism behind the LAMP reaction involves three major steps: initiation, cycling amplification, and elongation. Typically, the reaction begins with the initial step, i.e. the binding of the inner primers, followed by strand displacement DNA synthesis by the outer primers. Subsequently, the cyclical amplification step and elongation occur [30]. The auto strand displacement properties of the Bst polymerase enable using a heating block or a normal water bath maintained at a constant temperature for the amplification reaction instead of thermal cycling. The appearance of e.g. magnesium pyrophosphate precipitate is indicative for amplification of the target DNA, but amplicon binding fluorophores are also available. Real-time turbidimetry facilitates the quantification of the template DNA in the reaction and allows the analysis of minute quantities of DNA. Furthermore, LAMP amplicons can be analyzed using agarose gel electrophoresis and/or simple colorimetric naked-eye visualization closed detection systems or real-time fluorometry. As a result of these properties, the major advantage of LAMP is its application in the field or other resource-limited settings [30, 34, 50].

Four LAMP assays were recently designed for detecting M. leprae: Garg et al., Jiang et al. and Joshi et al. each designed RLEP or 16S LAMP assays, respectively, that required only standard consumables and a constant temperature water bath or heating block. Naked-eye detection was achieved via DNA intercalating dyes (colorimetric detection)—in case of Joshi et al. with a closed tube technology to minimize the risk of cross-contamination—and additional visual detection of turbidity and bridge flocculation assay as used by Garg et al. For further confirmation also agarose gel electrophoresis was used by Garg et al. and Jiang et al. [51,52,53]. Joshi et al. additionally designed a RLEP multiplex LAMP (m-LAMP) for differential detection of M. leprae and Leishmania donovani using the portable real-time Genie® II fluorometer from OptiGene (Horsham, United Kingdom) for amplification and detection of the LAMP products [54].

To the best of our knowledge no LAMP detecting M. leprae was described that uses dried reagents not requiring a continuous cold or freezing chain. Based on own experience with the development of a dry-reagent-based (DRB) IS2404 LAMP for decentralized diagnosis of Buruli ulcer in cooperation with the Foundation for Innovative Diagnostics (FIND) which currently awaits clinical validation in Togo [55], our group designed and validated the first RLEP DRB LAMP for the diagnosis of leprosy at point-of-care, provided product specifications for customization of a ready-to-use kit and compared it against the in-house prototype.

Methods

This study aimed to establish, optimize and validate a leprosy-specific RLEP-detection-based LAMP assay employing lyophilized reagents (i.e. in-house RLEP DRB LAMP) transformable into a commercially produced ready-to-use POC kit (i.e. ready-to-use RLEP DRB LAMP).

Ethical approval

The study involves human tissue from Togolese, Ghanaian and German patients.

For Togolese and Ghanaian patients, samples were collected in the course of two studies, approved by the national Togolese (Université de Lomé, and Ministry of Health, 012/2012/CBRS) and the Ghanaian KNUST (Kwame Nkrumah University of Science and Technology; NMIMR-IRB CPN 116/12-13) Ethics Committees. Written informed consent (IC) was obtained from all Togolese and Ghanaian study participants and/or their legal representatives if below 18 years of age.

All German patients sought medical advice at our outpatient department and signed a consent form approving the collection of data and samples before they were subjected to routine clinical and laboratory procedures. The form has been approved by the institutional ethics review board of the Medical Centre of the University of Munich (LMU).

LAMP—primer set

A set of RLEP LAMP primers was pre-designed with PrimerExplorer V4 from Eiken Chemical, Tokyo, Japan [49]. On this basis and analyzing DNA sequences of the M. leprae specific repetitive element (RLEP, GenBank, NCBI) LAMP primers for the detection of RLEP sequences from M. leprae derived from different clinical samples of various geographic origins were manually optimized for a customized application. The primer set consisted of a pair of inner and outer primers named RLEP FIP, RLEP BIP and RLEP F3, RLEP B3 specific for the RLEP sequence in the M. leprae genome. The specificity of the primers was confirmed with the basic local alignment search tool (BLAST, GenBank, NCBI) [56].

RLEP F3 has a forward complementary sequence, RLEP B3 has a reverse complementary sequence and RLEP FIP, as well as RLEP BIP, show a more complex sequence construction binding each at two different positions which results in loop structures typical for LAMP.

Primer positions within the RLEP sequence of the leprosy genome (Mycobacterium leprae Br4923, GenBank Accession No.: FM211192.1) are displayed in Fig. 1.

Fig. 1
figure 1

Primer set and location in the RLEP sequence of M. leprae Br4923. Primers RLEP FIP and RLEP BIP contain two sequences each binding at different positions ([position 1] and [position 2])

Primer sequences:

RLEP F3: 5ʹCGCACCTGATGTTATCCCTTʹ3

RLEP FIP: 5ʹATGCCTGCTTGCTGGCTGAG [1] CACCATTTCTGCCGCTGG [2]ʹ3

RLEP BIP: 5ʹCAGTGCATCGATGATCCGGCC [1] GTGTGGGTGGTTGATCTGC [2]ʹ3

RLEP B3: 5ʹGGTTTGGGTGGTGTTGTGGʹ3

LAMP—optimizing the protocol

After primer design, a protocol for the newly designed LAMP was established and optimized on a qPCR device (CFX 96; BioRad Laboratories, Hercules, USA) using duplicate measurements of RLEP standard 107 (GenExpress, Berlin, Germany; described below in the samples section).

As a dry-reagent-based NAAT depends crucially on a freeze-dryable master mix, based on our experience from previous projects we compared two master mixes, namely Isothermal Mastermix ISO-DR-001 and Isothermal Mastermix ISO-DR-004 from OptiGene in a first test. For a second test, as our own experience differed slightly from the manufacturer’s protocol, different ratios of the inner (RLEP FIP/BIP [20 µM]) and outer primers (RLEP F3/B3 [5 µM]) were compared (inner primers: outer primers 1:1 against 2:1). For a third test, eight different annealing temperatures were compared using a temperature gradient: 68.6 °C, 68.1 °C, 67.0 °C, 65.0 °C, 62.6 °C, 60.6 °C, 59.3 °C and 58.6 °C. For all three tests each the mean, standard deviation (SD) and coefficient of variation (CV) were used for comparison. For the resulting protocol please refer to Additional file 1—in-house “wet” RLEP LAMP run protocol.

DRB LAMP—freeze-drying

The optimized LAMP protocol was freeze-dried in our department (in-house lyophilization protocol). Then, based on our experience with the development of a DRB IS2404 LAMP for the diagnosis of Buruli ulcer, according to our specifications the same LAMP protocol was lyophilized by Amplex (Amplexdiagnostics GmbH, Gars am Inn, Germany; ready-to-use lyophilization protocol), to generate simple ready-to-use LAMP test strips for comparison with the in-house lyophilization protocol and eventually future application at point-of-care [55].

DRB LAMP—reaction mix for lyophilization

One DRB LAMP reaction mix contained 54 µl primer mix (i.e. 18 µl RLEP FIP [20 µM], 18 µl RLEP BIP [20 µM], 9 µl RLEP F3 [5 µM] and 9 µl RLEP B3 [5 µM]; TibMolBiol, Berlin, Germany) and 90 µl master mix (Isothermal Mastermix ISO-DR-004 [OptiGene] re-suspended with PCR grade water).

In-house lyophilization protocol

The reaction mix for the in-house RLEP DRB LAMP was lyophilized in-house into 0.5 ml screw cap microtubes (SARSTEDT, Nümbrecht, Germany) with the freeze dryer Alpha 1–2 LDplus (Christ Martin, Osterode, Germany) for 20 h at 0.31 mbar. For the whole protocol please refer to Additional file 2—in-house lyophilization protocol. All reaction mixes for the in-house RLEP DRB LAMP were stored at 4 °C in the dark.

Ready-to-use lyophilization protocol

The reaction mix for the ready-to-use RLEP DRB LAMP was commercially lyophilized by Amplex directly in the wells of the LAMP strips (8-well Genie® Strips, OptiGene). The test is currently for research use only; CE certification can be sought upon demand.

DRB LAMP—run preparation and protocol

For the in-house as well as the ready-to-use lyophilization protocol the reaction mixes for the in-house as well as the ready-to-use RLEP DRB LAMP were resolved as described below, and the samples and controls were added to run the test on a Genie® III fluorometer (OptiGene), which is a hand-held, battery-driven device for isothermal amplification and real-time fluorescence detection at the point-of-care [57]. Please also refer to the Additional file 3—in-house RLEP DRB LAMP run protocol and Additional file 4—ready-to-use RLEP DRB LAMP run protocol.

In-house RLEP DRB LAMP—run preparation and materials

For testing of samples with the in-house RLEP DRB LAMP, 135 µl buffer ISO-DR-004 (OptiGene) and 72 µl PCR grade water were added to the lyophilized in-house RLEP DRB LAMP reaction mix to dissolve it to the total amount of 207 µl. In each well of an 8-well Genie® Strip, 23 µl dissolved reaction mix was pipetted and 2 µl template was added to a final reaction volume of 25 µl.

Ready-to-use RLEP DRB LAMP—run preparation and materials

For testing of samples with the ready-to-use RLEP DRB LAMP 12.5 µl buffer ISO-DR-004 and 7.5 µl water were pipetted directly in the wells of the 8-well Genie® Strip to the ready-to-use RLEP DRB LAMP reaction mix. Then, 5 µl template was added to reach a final reaction volume of 25 µl.

DRB LAMP—run protocol

The Genie® III fluorometer provides space for one 8-well Genie® Strip, so up to six samples can be processed together with 1 negative and 1 positive control.

Overall, one test had a turn-around time of 45 min, consisting of 10 min hands-on for preparation and 35 min for the run protocol on the Genie® III fluorometer (30 min at 65 °C for amplification and detection, then heating to 98 °C for 1 min and reducing the temperature in steps of 0.05 °C/s to 80 °C for modified melting curve analysis). In Fig. 2 an example of a run is provided.

Fig. 2
figure 2

Example of a Genie® III run including the corresponding modified melting curve analysis. 953.16–1204.16 = samples (tested positive); PC = positive control; NC = negative control

Storage testing

The shelf life of the in-house RLEP DRB LAMP was calculated based on accelerated ageing, where the effect of ageing is simulated in a heating cabinet. The calculation of the presumed storage time is based on Arrhenius’ equation, which simply states that a 10 °C increase in temperature doubles the rate of a chemical reaction. To give an example: if a kit is stable for 60 days at 48 °C it is most likely also stable for 1 year at ambient temperature (22 °C). The accelerated ageing calculator applied can be found at iso-inc.com [58].

For testing of accelerated ageing, in-house RLEP DRB LAMP tubes were stored for up to 120 days at 48 °C dry and dark in a heating cabinet (Heraeus BS042E, Thermo Electron Corp. Waltham, USA), to calculate the shelf-life at ambient temperature. After 15, 30, 60 and 120 days in-house RLEP DRB LAMP tubes were tested with a standard series RLEP Std 107–103 (GenExpress, Berlin, Germany; described below in “Samples” section). If the standard samples were tested positive, the experiment was continued; if one or more of the standard samples were tested negative, the accelerated ageing time was defined as exceeded, and the experiment aborted.

Additionally, the shelf life at 4 °C was tested directly. Therefore in-house RLEP DRB LAMP tubes were stored for up to 1 year at 4 °C dry and dark in a laboratory refrigerator. After 3, 6 and 12 months each in-house RLEP DRB LAMP tube was tested as described for the accelerated ageing.

The storage of the ready-to-use RLEP DRB LAMP was assessed by Amplex using an internal protocol.

Samples

For technical and clinical validation RLEP standards, culture extracts and a wide range of clinical samples from various patients and control groups were tested.

RLEP plasmid standards

Lyophilized RLEP plasmid standards used for optimization, storage testing and technical validation (RLEP Std 108; GenExpress, Berlin, Germany) were resuspended with DNA hydration solution (Qiagen, Hilden, Germany), according to the manufacturer’s instructions (GenExpress). A tenfold serial dilution (RLEP Std 107–101) was prepared by successive diluting of 10 µl RLEP Std in 90 µl DNA hydration solution. The 108 dilutions were aliquoted and stored at − 20 °C for a maximum of 30 days. Thawed aliquots and tenfold serial dilutions were stored at 4 °C and used within 1 week.

Clinical samples

A wide range of clinical samples (nasal and buccal swab samples, slit skin smears, biopsies, fine needle aspirates [FNAs], skin scrapings) from non-exposed and endemic controls as well as patients with different diseases were tested.

Collection of clinical samples

Clinical samples from untreated MB leprosy patients diagnosed according to the WHO classification and treated in Togo after sample collection, MB leprosy patients diagnosed and treated (prior to sample collection) in Germany, endemic controls from Togo and Germany and healthy non-exposed controls from Germany were collected as previously described [16, 59, 60]. Briefly, nasal and buccal swab samples were collected with custom-made swabs (bio-Budget, Krefeld, Germany), slit skin smears were collected from ear lobes and slit skin smears and biopsies were collected from the edges of lesions. All samples were stored in 700 µl cell lysis solution (CLS, Qiagen) and transported at ambient temperature to the respective laboratory (INH [Institut National d’Hygiène, Lomé] in Togo and DITM in Germany).

Clinical samples from patients with Buruli ulcer disease (BUD) originating from Togo and Ghana were collected as previously described [60] and transported to the respective laboratory (INH in Togo and KCCR [Kumasi Centre for Collaborative Research in Tropical Medicine, Kumasi] in Ghana). Briefly, swabs were collected by circling the entire undermined edges of ulcerative lesions and transported in 700 µl CLS. FNAs were collected from the centre of non-ulcerative lesions or undermined edges of ulcerative lesions including necrotic tissue and transported in 300 µl CLS.

Skin scrapings and biopsies from patients with cutaneous leishmaniasis diagnosed in the outpatient department of DITM were transported in 300 µl CLS to the laboratories of DITM.

Extraction of samples

All clinical samples from patients with cutaneous leishmaniasis were extracted using a modified protocol for the “Puregene tissue kit” from Qiagen (Protocol: “DNA Purification from Tissue Using the Gentra Puregene Tissue Kit”, Gentra Puregene Handbook 12/2014). Briefly, the tissue was placed into 300 µl CLS and 1.5 µl Puregene proteinase K (20 mg/ml) was added for incubation at 55 °C overnight. Subsequently, 100 µl Protein Precipitation Solution was added on ice and after a centrifugation step, the supernatant was poured to 300 µl isopropanol. After centrifugation the DNA pellet was washed in 70% ethanol, air dried, and 50 µl DNA hydration solution was added to dissolve the pellet.

All other samples (including culture supernatants in 700 µl CLS) were extracted employing the Gentra Puregene method (Qiagen) as previously described [16].

All extractions included extraction controls for the subsequent (q)PCR as routinely conducted in the accredited laboratories of DITM.

Testing of clinical samples

All clinical samples from leprosy patients were tested with RLEP qPCR (as initially designed by Truman et al. [14] and optimized by Beissner et al. [16]).

Additionally, direct DNA sequencing of 21 samples as indicated in Tables 1 and 2 was performed to confirm the presence of M. leprae DNA in the respective samples and to determine the RLEP copy number of M. leprae strains from patients with Togolese and Pakistani origin for exact quantification of the number of bacilli in clinical samples by RLEP qPCR [16].”

Table 1 Number, type and origin of samples used for technical validation of the RLEP DRB LAMP (n = 65)
Table 2 Number, type and origin of samples used for clinical validation of the RLEP DRB LAMP (n = 150)

Determination of bacillary load of clinical samples

Routinely, exact quantification of the number of bacilli per extract, as recently described by Beissner et al. was performed [16]. Briefly, a standard curve of the RLEP qPCR was generated using serial dilutions (107–103 copies) of plasmid standards to estimate the starting quantity (SQ) of clinical samples. The bacillary load (BL) was calculated by BL = (SQ × [volume of DNA extract/volume of template])/copy number. The bacillary load of M. leprae was grouped into three categories: high-positive (i.e. > 10.000 bacilli/extract), medium-positive (i.e. 1.001–10.000 bacilli/extract) and low positive (i.e. 1–1.000 bacilli/extract).

Samples used for technical validation

For technical validation, RLEP plasmid standards as well as 25 “must detect RLEP” samples from Togo (n = 18) and Germany (n = 7), all RLEP qPCR positive and with a bacillary load of at least 3.000 bacilli/extract, were used (Table 1).

Also, 40 “must not detect RLEP” samples were tested. These consisted of a set of culture extracts from closely related mycobacteria (n = 14), clinical samples derived from IS2404 qPCR confirmed BUD patients from Togo (n = 1) and Ghana (n = 3) (PCR as first described by Fyfe et al. [61] and modified by Beissner et al. [62]), clinical samples derived from patients with PCR confirmed cutaneous leishmaniasis (n = 4; PCR as described by Schönian et al. [63]), as well as from endemic controls, i.e. RLEP qPCR negative suspected leprosy cases from Togo (n = 10) and Ghana (n = 2) and healthy non-exposed controls, i.e. DITM laboratory staff (n = 6) (Table 1).

Samples used for clinical validation

For clinical validation, 150 clinical samples were tested: 30 RLEP qPCR high-positive samples of MB leprosy patients from Togo (n = 24) and Germany (n = 6), 30 RLEP qPCR medium-positive samples of MB leprosy patients from Togo (n = 17) and Germany (n = 13), 30 RLEP qPCR low-positive samples of MB leprosy patients from Togo (n = 15) and Germany (n = 15), 30 samples from endemic controls (suspected leprosy cases, RLEP qPCR negative from Togo (n = 28) and Germany (n = 2) and 30 samples from healthy non-exposed controls (non-exposed DITM laboratory staff from Germany, RLEP qPCR negative) (Table 2).

Technical validation

For the technical validation of both the in-house and the ready-to-use RLEP DRB LAMP formats, the limit of detection, sensitivity, specificity as well as the inter- and intra-assay variations were determined following the MIQE guidelines for qPCR, as no guidelines exist for validation criteria of the LAMP format [64].

LOD

The analytical sensitivity was determined as the lower limit of detection (LOD, i.e. lowest concentration with > 95% samples tested positive), using tenfold serial dilutions of RLEP plasmid standards.

The LODs were re-confirmed in the matrix of 10 of the “must not detect RLEP” samples (7 swab samples, 2 slit-skin-smears and 1 biopsy sample; all RLEP qPCR and RLEP DRB LAMP negative), spiked with 1.000 copies of the plasmid standard. Therefore the “must not detect samples” amounted to one-half of the template volume and 2 × 103 copies of the plasmid standard amounted to the other half to get the required dilution of plasmid standard in a semi-natural “sample”.

Sensitivity

The sensitivity rate was defined as the number of “must detect RLEP” clinical samples with a positive RLEP DRB LAMP divided by the number of all “must detect RLEP” clinical samples (n = 25).

Specificity

The specificity rate was defined as the number of “must not detect RLEP” samples with a negative RLEP DRB LAMP divided by the number of all “must not detect RLEP” samples (n = 40).

Inter-assay variation

Inter-assay variability was assessed by testing an RLEP plasmid standard (diluted 106) on 3 subsequent days to calculate the mean, standard deviation (SD) and coefficient of variation (CV) of the inter-assay variability. Calculations were conducted in accordance with MIQE guidelines for qPCR in previously described modifications for other NAAT assays [1, 62, 64]. Variability was judged low if the coefficient of variation (CV) was ≤ 0.05, medium CV ≤ 0.2, high CV > 0.2.

Intra-assay variation

Seven samples with Std 106 were analyzed in one run to calculate the mean, SD and CV of the intra-assay variability. Variability was judged low if the coefficient of variation (CV) was ≤ 0.1, medium CV ≤ 0.5, high CV > 0.5.

Semiquantitative results

The correlation between the bacillary load and the time to positivity (Tp) was assessed.

Clinical validation

For clinical validation of both the in-house and the ready-to-use RLEP DRB LAMP formats, positivity and negativity rates, as well as positive and negative prediction values, were determined as follows:

Positivity rate

The positivity rate was defined as the number of RLEP qPCR positive samples with a positive RLEP DRB LAMP divided by the number of all RLEP qPCR positive samples (n = 90).

Negativity rate

The negativity rate was defined as the number of RLEP qPCR negative samples with a negative RLEP DRB LAMP divided by the number of all RLEP qPCR negative samples (n = 60).

Positive predictive value (PPV)

The PPV was defined as the number of all RLEP qPCR positive samples with a positive RLEP DRB LAMP divided by the number of all RLEP DRB LAMP positive samples.

Negative predictive value (NPV)

The NPV was defined as the number of all RLEP qPCR negative samples with a negative RLEP DRB LAMP divided by the number of all RLEP DRB LAMP negative samples.

Statistical analysis

For comparison of the two LAMP formats, an estimation of the standard error of proportion (95-per cent confidence intervals [95%-CI]) was conducted with Microsoft Excel 2010 (Microsoft cooperation, Redmond, USA). Significant differences were defined as not overlapping of 95%-CI of proportions.

Results

After optimizing the protocol, the RLEP DRB LAMP formats (in-house and ready-to-use) passed through storage testing and technical as well as clinical validation.

LAMP—optimizing the protocol

Regarding the optimization of the “wet” RLEP LAMP protocol (the preliminary stage of both RLEP DRB LAMP formats), the first test revealed that the Isothermal Mastermix ISO-DR-004 proved to be more sensitive and rapid than the ISO-DR-001. For the second test, inner and outer primers at a ratio of 2:1 proved to be more sensitive than at a ratio of 1:1. For the third test, the temperature 65.0 °C proved to be the most sensitive. Based on these test results the optimized protocol as outlined in the “Methods” section was determined. Data are provided in Table 3.

Table 3 Results of the LAMP optimization

Storage testing

In-house storage performed well at 48 °C for up to 60 days, but standard samples were tested negative after storage at 48 °C for 120 days resulting in shelf life for 1 year at ambient temperature (22 °C) as calculated utilizing an accelerated ageing calculation [58]. All in-house RLEP DRB LAMP tubes stored at 4 °C performed well, resulting in the minimum shelf life of 1 year.

Ready-to-use RLEP DRB LAMP storage testing resulted in the shelf life of 18 months at room temperature according to the manufacturers’ instructions (15–30 °C).

Technical validation for RLEP DRB LAMP (in-house and ready-to-use)

LOD

The lower limit of detection for both RLEP DRB LAMP formats was 1.000 RLEP copies, with an amplifiable copy number varying between 23 and 37 that constitute 43–27 genome equivalents (for Togolese patients 33 genome equivalents and for one patient with Pakistani origin 43 genome equivalents). The LOD was re-confirmed in the matrix of clinical samples.

Sensitivity

For the in-house RLEP DRB LAMP, 23/25 “must detect RLEP” clinical samples were tested positive, resulting in a sensitivity of 92% (95%-CI 81; 100). For the ready-to-use RLEP DRB LAMP, all 25 “must detect RLEP” clinical samples were tested positive, resulting in a sensitivity of 100%. Data are provided in Table 4.

Table 4 Testing of “must (not) detect RLEP” samples resulting in sensitivity and specificity of both RLEP DRB LAMP formats (in-house and ready-to-use)

Specificity

For both RLEP DRB LAMP formats all of the 40 “must not detect” samples were tested negative, resulting in a specificity of 100% for the in-house as well as for the ready-to-use RLEP DRB LAMP. Data are provided in Table 4.

Inter-assay variation

For both RLEP DRB LAMP formats, the inter-assay variability was judged low concerning annealing temperature. Concerning time to positivity and fluorescence, the inter-assay variability was judged medium for the in-house RLEP DRB LAMP and low for ready-to-use RLEP DRB LAMP. Data are provided in Table 5.

Table 5 Data on inter-and intra-assay variability of the RLEP DRB LAMP formats (in-house and ready-to-use)

Intra-assay variation

For both RLEP DRB LAMP formats, the intra-assay variability was judged low. Data are provided in Table 5.

Clinical validation

Results of the testing of clinical samples are summarized in Table 6.

Table 6 Results of the clinical validation of both RLEP DRB LAMP formats (in-house and ready-to-use)

Positivity rate—in-house RLEP DRB LAMP

For high positive samples, 100% (30/30) were tested positive. For medium positive samples, 100% (30/30) were tested positive. For low positive samples (i.e. samples below the LOD), 77% (23/30; 95%-CI 62; 92) were tested positive. The positivity rate for all samples tested was 92% (83/90, 95%-CI 87; 98). The positivity rate for all samples above the LOD was 100%. Analysis of the different sample types: out of seven false negative samples, one was a buccal swab sample (1/2, i.e. 50% of all positive buccal swab samples tested [95%-CI 0; 100]), five samples were nasal swab samples (5/74, i.e. 7% of all positive nasal swab samples tested [95%-CI 1; 12]) and one sample was a slit skin smear (1/13, i.e. 8% of all positive slit skin smears tested [95%-CI 0; 22]).

Positivity rate—ready-to-use RLEP DRB LAMP

For high positive samples, 100% (30/30) were tested positive. For medium positive samples, 93% (28/30, 95%-CI 84; 100) were tested positive. For low positive samples (i.e. samples below the LOD), 43% (13/30, 95%-CI 26; 61) were tested positive. The positivity rate for all samples tested was 79% (71/90, 95%-CI 70; 87). The positivity rate for all samples above the LOD was 97% (58/60, 95%-CI 92; 100). Analysis of the different sample types: of the 19 false negative samples, two were buccal swab samples (2/2, i.e. 100% of all positive buccal swab samples tested) and 17 were nasal swab samples (17/74, i.e. 23% of all positive nasal swab samples tested [95%-CI 13; 33]). Significant differences between both RLEP DRB LAMP formats were found for the positivity rates of all 90 positive samples tested as well as for the positivity rates for low positive samples.

Negativity rate

For both RLEP DRB LAMP formats, the negativity rate was 100% (30/30 samples from endemic controls and 30/30 samples from healthy non-exposed controls were negative).

Positive predictive value—in-house RLEP DRB LAMP

The positive predictive value was 100% (0/83 results were false positive).

Positive predictive value—ready-to-use RLEP DRB LAMP

The positive predictive value was 100% (0/71 results were false positive).

Negative predictive value—in-house RLEP DRB LAMP

The negative predictive value was 90% (7/67 results were false negative).

Negative predictive value—ready-to-use RLEP DRB LAMP

The negative predictive value was 76% (19/79 results were false negative).

Semi-quantitative results

Both RLEP DRB LAMP formats proved to be semi-quantitative, there is a negative correlation between the time to positivity (Tp) and the bacillary load. Data are provided in Table 7 and Fig. 3.

Table 7 Comparison of the semi-quantitative results of RLEP DRB LAMP formats (in-house and ready-to-use)
Fig. 3
figure 3

Comparison of the semi-quantitative results of RLEP DRB LAMP formats (in-house and ready-to-use). Tp = time to positivity; boxplot with the lowest and highest result shown as whiskers

Discussion

Nucleic acid-based amplification tests (NAAT) are widely used for the laboratory confirmation of TB and BUD in reference laboratories of endemic countries and constitute an integral part of disease control activities for both conditions [65, 66]. As demonstrated in recent publications, NAAT also have great potential for applications in leprosy control. NAAT enable laboratory-based early confirmation of leprosy cases, allow laboratory-based screening of HHC facilitating laboratory confirmation of difficult to diagnose lesions and identification of individuals with nasal carriage of M. leprae eligible for further monitoring, as well as assessing the treatment success [4, 5, 16, 17, 24, 25, 67, 68]. Due to the lack of standardization and commercial availability, the requirement of technical and laboratory expertise, and the difficulty to perform NAAT in most primary health-care settings, current leprosy control strategies have not yet entailed the application of NAAT and the WHO still recommends clinical examination, if available supported by microscopy, as the standard method for diagnosis of leprosy [1, 69]. The Global Leprosy Strategy 2021–2030, however, schedules implementation of laboratory diagnostics from 2026 onwards and the TAG explicitly recommends improving laboratory capacity including molecular diagnostics [27, 28].

Recent leprosy research prioritized early case detection, contact screening and chemoprophylaxis as the most promising tools to interrupt transmission and enhance the effectiveness of leprosy control.

Results from the most comprehensive study on combining contact tracing with prophylactic treatment “Leprosy Post-Exposure Prophylaxis” (LPEP) with single-dose rifampicin (SDR) conducted from 2015 through 2019 with the leprosy control programs of Brazil, India, Indonesia, Myanmar, Nepal, Sri Lanka, and Tanzania, showed that the approach of contact tracing followed by the provision of SDR is feasible as part of routine leprosy control activities. Contacts of clinically diagnosed leprosy index patients were traced and screened for signs of the disease. Those with signs of active leprosy were evaluated according to the routine leprosy control program procedures and, if the diagnosis was confirmed, were administered MDT. Contacts without any evidence of leprosy were assessed for their eligibility to receive SDR. The authors concluded that further research should focus on local epidemiological situations, i.e. the level of endemicity, the ratio of multibacillary to paucibacillary patients and health system characteristics. Following the WHO Global Leprosy Strategy, and the conclusions of other studies, the development of field-friendly diagnostic tests can help to identify early leprosy and those contacts with subclinical infection who would benefit most from SDR post-exposure prophylaxis or further measures, such as modified chemoprophylactic regimens or intensified monitoring [69,70,71,72].

In accordance with the current state of research, the priority to develop a POC test for leprosy is high [73]. A test suitable for the developing world should meet the WHO ASSURED criteria (Affordable, Sensitive, Specific, User-friendly, Rapid and robust, Equipment-free and Delivered) [74, 75]. Classical NAAT such as (q)PCRs typically stand out in the fields of sensitivity and specificity but cannot meet the other criteria. LAMP comes up with several advantages: The technology is less expensive than other NAAT, has an equal sensitivity as classical PCR, a high specificity due to its multiple primers, is rapid (< 1 h turn-around time) and does not require sophisticated equipment [30, 50, 75, 76].

Although perfectly suitable for POC diagnostics, LAMP has a few disadvantages, especially the specificity is questioned: briefly, although multiple primers promote specificity, it may also increase the risk of primer-primer binding and therefore false-positive results. In this validation study, the authors used a modified melting curve analysis of each LAMP product to ensure the specificity of amplicons on the Genie® III fluorometer (as provided by the manufacturer). Nevertheless vast technical and clinical validation studies are required prior to broad-range field-application [77, 78].

Like in other NAAT, unintended contamination through carryover of DNA from well-to-well is also of concern applying LAMP [78,79,80,81]. The extraction and run protocols should therefore be reduced to a minimum of pipetting steps, preferable in a closed, one-step system [36].

Nijiru et al. rated LAMP a potential ASSURED test, given four elements, i.e. template preparation protocols, a lyophilized kit, a reliable power source, and product detection technologies are available [76]. Template preparation was not in the focus of this study but must be subject to further, separate investigation. The other three elements were covered by the formats compared in this study. The formats were dry-reagent based (including master mix and primers) and therefore robust, affordable and worked on a closed amplification and detection unit with a stable power source sufficient for working one whole day in the field (the device only needs a standard power socket for charging). The fluorometric detection of LAMP products allowed the real-time visualization of the amplification curves and, based on the predetermined strand-inherent temperature profile, the application of a melting curve analysis presented a means of quality assurance.

Although coming close to the equipment-free principle put in place by the WHO 2012 [82], our assays were not completely equipment-free which may be considered a drawback. Lateral flow assays (LFAs) are typical examples of equipment-free POC tests, defining them as devices “operable in resource-limited environments, which include those with unreliable electricity, non-sterile conditions and a lack of trained personnel to perform the duties typically reserved for nurses and health workers” [82, 83]. With its stable power source and closed set-up the Genie® III fluorometer meets part of the criteria, but although it is hand-held it represents technical equipment. According to the manufacturer the Genie® III software may be programed based on our validation data to simply display the results as positive or negative, but the personnel working with it will still need at least a short training before fully understanding the LAMP assay. However, due to the option of melting curve analysis, the technical component provides the chance to rule out the above-mentioned disadvantage of primer-primer binding and false-positive results. In the authors’ view, the advantages outweigh the disadvantages.

Four LAMPs for detecting M. leprae were published up to now (Garg et al., Jiang et al., Joshi et al. and Joshi et al. m-LAMP), all 100% specific, with a sensitivity ranging from 83 to 100% (being more sensitive for MB patients and tissue samples and less sensitive for PB patients and slit skin smears) and an analytical sensitivity of 30 bacilli each for both LAMPs from Joshi et al. [51,52,53,54]. With a specificity of 100%, a sensitivity of 92% for minimal invasive samples (nasal swab samples and slit skin smears) and an analytical sensitivity of 27–43 bacilli (depending on the number of RLEP copies per strain) our results are well comparable with the other recently published assays. The three published LAMPs applying standard consumables and easy-to-use nontechnical equipment have two limiting factors: non-specific amplification with different primers can generate false-positive samples, which is a common drawback of LAMP and—as described by Jiang et al.—the fluorescence-based visual detection of results can be ambiguous, leading to the need of confirmation by gel electrophoresis, which would outweigh the benefits of the minimalist technique and can also generate false-positive samples due to contamination [51,52,53]. Joshi et al. solved these problems with their m-LAMP in the same way as we did, i.e. with a melting curve analysis to rule out false-positive samples and a closed and precise detection system [54]. However, in contrast to our DRB assays, the reagents for all M. leprae detecting LAMP formats recently described need to be transported at lower temperatures (mostly at − 20 °C), which hampers the use for POC.

A POC test must be stable at ambient temperature for as long as possible. DRB LAMP assays are mostly stabilized through drying [84,85,86] or lyophilization [87,88,89]. Several LAMPs have been reported to be stable at 4 °C for 2–5 months [37, 87] and at ambient temperature (specified in the literature as 15–30 °C) for 3 weeks to 7 months [37, 84, 87, 88, 90]. To the best of our knowledge, with (more than) 1 year both LAMP assays revealed the longest shelf-life at 4 °C and 22 °C. It is, therefore, possible to e.g. produce the test in Europe, ship it to endemic countries at ambient temperature via normal postal service for storage and application at POC without loss of sensitivity for test results. To the best of our knowledge, this is unique to all previously described M. leprae LAMP assays.

Our two RLEP DRB LAMP formats differ only by the mode of lyophilization (in-house lyophilization vs commercial lyophilization by Amplex directly in the wells of the LAMP strips for the ready-to-use kit). But despite the substantial common features both formats show differences: The in-house RLEP DRB LAMP is significantly superior in detecting samples with a low bacillary load of M. leprae as positive, but requires in-house lyophilization, which is time-consuming and costly (concerning the devices needed). The ready-to-use RLEP DRB LAMP on the other hand is with one less pipetting step easier to prepare and more protected against contamination, and—once CE-certified—may be purchased in large quantities directly from the producer.

Application of the RLEP DRB LAMP assays at POC of course rises the question of the relevance of positive results for classification and clinical management of patients. The WHO classification introduced in 1988 initially employed the lesion count and the bacteriological index (BI) for distinction of PB and MB forms [91]. With the abolishment of bacteriological examination in 1998, misclassification occurred and studies describe that up to 51% of cases clinically diagnosed as PB forms were in fact MB cases—with the consequence of undertreatment and the risk of treatment failure [92, 93]. The limit of detection of microscopy is indicated as 104 bacilli per g tissue which corresponds to approximately 100 bacilli in a 6 mm punch biopsy [94, 95]. The limit of detection of our RLEP DRB LAMP assay has been determined approximately 30 bacilli per sample, the sensitivity of the assay is therefore three times higher as the sensitivity of microscopy. In terms of clinical application the assays can provide a bacteriological confirmation of PB cases, especially those with scanty bacilli (BI 1) with a higher sensitivity than microscopy could achieve.

Furthermore, according to the previous approach of applying lesion numbers and bacteriological examination to classification, clinically diagnosed PB cases with LAMP positivity could be correctly re-classified as MB cases and treated accordingly. We assume that a number of BT cases with scanty bacilli which are in fact correctly classified as PB cases by counting of lesions will be detected as well—but we believe that “overtreatment” in sporadic BT cases is acceptable and less of an issue of concern as “undertreatment” of missed MB cases. However, to put these assumptions on a sound scientific basis, detailed studies on the correlation of clinical classification by lesion count, BI, RLEP DRB LAMP positivity in different sample types (nasal swabs, SSS, biopsies) and histopathological classification are required which use RLEP qPCR as reference standard to determine threshold values for the bacillary load associated with different histological forms of the disease.

An evaluation trial in leprosy endemic regions may be required, especially for answering the question if HHC with subclinical infection are also detected by the assays. Thereby, the ready-to-use RLEP DRB LAMP may serve as large scale screening test for HHC as well as clinically suspected leprosy cases. Due to its augmented positivity rates in medium- and low-positive samples, the in-house RLEP DRB LAMP, which is lyophilized manually and therefore only in low quantities, should be only used as confirmatory test, if the first test reveals a negative result despite strong clinical evidence of leprosy. In this step-wise-approach RLEP qPCR remains the third NAAT, conducted if both LAMP results turn out to be negative. Notwithstanding the capacity of NAAT, clinical diagnosis should remain the key component to administer adequate therapy if WHO criteria are met.

The ready-to-use RLEP DRB LAMP format presented here constitutes an ASSURED test applicable at the primary health care level and can facilitate early case detection of leprosy index cases without the need for a third level reference laboratory [33]. Financing provided, a field-based evaluation trial to determine the clinical performance at peripheral health care level in Togo using the previously established infrastructure [60, 96], will provide ideal conditions for evaluation of the ready-to-use RLEP DRB LAMP prototype for leprosy regarding project logistics and availability of trained personnel.

In collaboration with FIND, a DRB IS2404 LAMP kit for Buruli ulcer using identical laboratory techniques as the ready-to-use RLEP DRB LAMP is currently on the way to commercial production. While the specificity of the IS2404 LAMP equals that of ready-to-use RLEP DRB LAMP for clinical samples, the analytical sensitivity was higher. For M. ulcerans the LOD was 50 copies of the target sequence (i.e. 0.25 M. ulcerans genome equivalents) [55] compared to 1.000 copies for RLEP DRB LAMP (i.e. 27–43 M. leprae genome equivalents according to the geographic origin and occurring RLEP variants). In terms of genome equivalents, this fact is explained by the vastly higher copy number of IS2404 per M. ulcerans genome (i.e. n = 209 [66]) as compared to RLEP per M. leprae genome (n = 23–37) [13,14,15,16,17]. Furthermore, the modified M. ulcerans LAMP [55] employed an additional set of loop-primers as first described by Ablordey et al. enhancing the process of amplification and thus the sensitivity [31]. However, we did not succeed in designing loop-primers encompassing nucleotide regions of all RLEP variants.

The new LAMP assays for laboratory diagnosis of leprosy and BUD expand the existing spectrum of NAAT for neglected tropical diseases from 16 to 18 pathogens and will be of great help for the managing of the respective diseases. The implementation of nearly identical LAMP solutions for two related mycobacterial diseases is a good example of the advantages of the integrated management of skin diseases, which result in the synergistic use of laboratory space, equipment and personnel in settings where both diseases are endemic, as it is already commonplace in Togo.

Conclusions

The priority to develop a POC test for leprosy in line with the ASSURED criteria is high. Based on our own experience with the development of a dry-reagent-based (DRB) LAMP for decentralized diagnosis of Buruli ulcer, our group designed and validated the first RLEP DRB LAMP assay for the diagnosis of leprosy at point-of-care. The DRB LAMP assay is affordable, 100% sensitive, 92–100% specific and has an LOD of 43–27 genome equivalents. It runs on the Genie® III (OptiGene), a user-friendly and rapid (run time 45 min) portable fluorometer. The dry-reagent-based assay is robust and stable for 12 months at ambient temperature. It is coming close to the equipment-free principle put in place by the WHO 2012 and—once CE-certified—can be delivered to point-of-care by simply purchasing the device (OptiGene) with lyophilized LAMP strips and the accompanying buffer (Amplex). The ready-to-use RLEP DRB LAMP assay was finally validated for field-based evaluation trials aiming for routine POC testing of leprosy suspected cases.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AFB:

Acid fast bacilli

BI:

Bacteriological index

BL:

Bacillary load

BLAST:

Basic local alignment search tool

Bst :

Bacillus stearothermophilus

BUD:

Buruli ulcer disease

CE:

“Conformité Européenne”, administrative marking

CI:

Confidence interval

CLS:

Cell lysis solution

Cp:

Copies

CV:

Coefficient of variation

DITM:

Division of Infectious Diseases and Tropical Medicine

DRB:

Dry-reagent-based

FIND:

Foundation for Innovative Diagnostics

FNA:

Fine needle aspirate

HAT:

Human African trypanosomiasis

HHC:

Household contact

IC:

Informed consent

INH:

Institut National d’Hygiène

KCCR:

Kumasi Centre for Collaborative Research in Tropical Medicine

KNUST:

Kwame Nkrumah University of Science and Technology

LAMP:

Loop-mediated isothermal amplification

LFA:

Lateral flow assay

LMU:

Ludwigs-Maximilians-University of Munich

LOD:

Lower limit of detection

MB:

Multibacillary

m-LAMP:

Multiplex LAMP

M. leprae :

Mycobacterium leprae

NAAT:

Nucleic acid-based amplification tests

PB:

Paucibacillary

POC:

Point-of-care

SD:

Standard deviation

SDR:

Single-dose rifampicin

SQ:

Starting quantity

TAG:

Technical and Advisory Group

TB:

Tuberculosis

Tp:

Time to positivity

References

  1. WHO. Guidelines for the diagnosis, treatment and prevention of leprosy. New Delhi: World Health Organization, Regional Office for South-East Asia; 2018.

    Google Scholar 

  2. Chavarro-Portilloa B, Sotob CY, Guerrero MI. Mycobacterium leprae’s evolution and environmental adaptation. Acta Trop. 2019. https://doi.org/10.1016/j.actatropica.2019.105041.

    Article  Google Scholar 

  3. Lastória JC, de Abreu MAMM. Leprosy: review of the epidemiological, clinical, and etiopathogenic aspects—Part 1. An Bras Dermatol. 2014;89:205–18.

    PubMed  PubMed Central  Google Scholar 

  4. Martinez AN, Talhari C, Moraes MO, Talhari S. PCR-based techniques for leprosy diagnosis: from the laboratory to the clinic. PLoS Negl Trop Dis. 2014. https://doi.org/10.1371/journal.pntd.0002655.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Tatipally S, Srikantam A, Kasetty S. Polymerase chain reaction (PCR) as a potential point of care laboratory test for leprosy diagnosis-a systematic review. Trop Med Infect Dis. 2018. https://doi.org/10.3390/tropicalmed3040107.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Soto A, Muñoz PT. Leprosy diagnosis: an update on the use of molecular tools lucrecia. Mol Biol. 2015. https://doi.org/10.4172/2168-9547.1000139.

    Article  Google Scholar 

  7. Scollard D. Leprosy research is a necessity. Lepr Rev. 2019;90:232–6.

    Google Scholar 

  8. Martinez AN, Britto CFPC, Nery JAC, Sampaio EP, Jardim MR, Sarno EN, Moraes MO. Evaluation of real-time and conventional PCR targeting complex 85 genes for detection of Mycobacterium leprae DNA in skin biopsy samples from patients diagnosed with leprosy. J Clin Microbiol. 2006;44:3154–9.

    CAS  PubMed  PubMed Central  Google Scholar 

  9. Martinez AN, Ribeiro-Alves M, Sarno EN, Moraes MO. Evaluation of qPCR-based assays for leprosy diagnosis directly in clinical specimens. PLoS Negl Trop Dis. 2011. https://doi.org/10.1371/journal.pntd.0001354.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Turankar RP, Pandey S, Lavania M, Singh I, Nigam A, Darlong J, et al. Comparative evaluation of PCR amplification of RLEP, 16S rRNA, rpoT and Sod A gene targets for detection of M. leprae DNA from clinical and environmental samples. Int J Mycobacteriol. 2015;4:54–9.

    PubMed  Google Scholar 

  11. Braet S, Vandelannoote K, Meehan CJ, Brum Fontes AN, Hasker E, Rosa PS, et al. The repetitive element RLEP is a highly specific target for detection of Mycobacterium leprae. J Clin Microbiol. 2018. https://doi.org/10.1128/JCM.01924-17.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Pathak VK, Singh I, Turankar RP, Lavania M, Ahuja M, Singh V, Sengupta U. Utility of multiplex PCR for early diagnosis and household contact surveillance for leprosy. Diagn Microbiol Infect Dis. 2019. https://doi.org/10.1016/j.diagmicrobio.2019.06.007.

    Article  PubMed  Google Scholar 

  13. Cole ST, Supply P, Honoré N, et al. Repetitive sequences in Mycobacterium leprae and their impact on genome plasticity. Lepr Rev. 2001;72:449–61.

    CAS  PubMed  Google Scholar 

  14. Truman R, Andrews PK, Robbins NY, Adams LB, Krahenbuhl JL, Gillis TP. Enumeration of Mycobacterium leprae using real-time PCR. PLoS Negl Trop Dis. 2008. https://doi.org/10.1371/journal.pntd.0000328.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Woods SA, Cole ST. A family of dispersed repeats in Mycobacterium leprae. Mol Microbiol. 1990;4(10):1745–51.

    CAS  PubMed  Google Scholar 

  16. Beissner M, Woestemeier A, Saar M, Badziklou K, Maman I, Amedifou C, et al. Development of a combined RLEP/16S rRNA (RT) qPCR assay for the detection of viable M. leprae from nasal swab samples. BMC Infect Dis. 2019. https://doi.org/10.1186/s12879-019-4349-9.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Saar M, Hanus I, Huber K, Beissner M, Löscher T, Bretzel G. Report on an unusual case of leprosy from Germany: just an exception of the rule? Infection. 2019. https://doi.org/10.1007/s15010-019-01349-8.

    Article  PubMed  Google Scholar 

  18. Pattyn SR, Ursi D, Ieven M, Grillone S, Raes V. Detection of Mycobacterium leprae by the polymerase chain reaction in nasal swabs of leprosy patients and their contacts. Int J Lepr Other Mycobact Dis. 1993;61:389–93.

    CAS  PubMed  Google Scholar 

  19. van Beers SM, Izumi S, Madjid B, Maeda Y, Day R, Klatser PR. An epidemiological study of leprosy infection by serology and polymerase chain reaction. Int J Lepr Other Mycobact Dis. 1994;62:1–9.

    PubMed  Google Scholar 

  20. Smith WC, Smith CM, Cree IA, Jadhav RS, Macdonald M, Edward VK, et al. An approach to understanding the transmission of Mycobacterium leprae using molecular and immunological methods: results from the MILEP2 study. Int J Lepr Other Mycobact Dis. 2004;72:269–77.

    PubMed  Google Scholar 

  21. Banerjee S, Sarkar K, Gupta S, Mahapatra PS, Gupta S, Guha S, et al. Multiplex PCR technique could be an alternative approach for early detection of leprosy among close contacts—a pilot study from India. BMC Infect Dis. 2010;10:252–9.

    PubMed  PubMed Central  Google Scholar 

  22. Araujo S, Lobato J, Reis Ede M, Souza DO, Gonçalves MA, Costa AV, et al. Unveiling healthy carriers and subclinical infections among household contacts of leprosy patients who play potential roles in the disease chain of transmission. Mem Inst Oswaldo Cruz. 2012;107:55–9.

    PubMed  Google Scholar 

  23. Cabral BP, Júnior JE, de Macedo AC, Alves AR, Gonçalves TB, Brito e Cabral TC, et al. Anti-PGL1 salivary IgA/IgM, serum IgG/IgM, and nasal Mycobacterium leprae DNA in individuals with household contact with leprosy. Int J Infect Dis. 2013. https://doi.org/10.1016/j.ijid.2013.05.011.

    Article  Google Scholar 

  24. Reis EM, Araujo S, Lobato J, Neves AF, Costa AV, Gonçalves MA, et al. Mycobacterium leprae DNA in peripheral blood may indicate a bacilli migration route and high-risk for leprosy onset. Clin Microbiol Infect. 2014;20:447–52.

    CAS  PubMed  Google Scholar 

  25. Araujo S, Freitas LO, Goulart LR, Goulart IM. Molecular evidence for the aerial route of infection of Mycobacterium leprae and the role of asymptomatic carriers in the persistence of leprosy. Clin Infect Dis. 2016;63:1412–20.

    PubMed  Google Scholar 

  26. Romero-Montoya M, Beltran-Alzate JC, Cardona-Castro N. Evaluation and monitoring of Mycobacterium leprae transmission in household contacts of patients with Hansen’s disease in Colombia. PLoS Negl Trop Dis. 2017. https://doi.org/10.1371/journal.pntd.0005325.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Cooreman E. Global-Leprosy-Strategy-Overview. In: Global consultation of NLP managers, partners and affected persons. 2020. https://www.who.int/docs/default-source/ntds/leprosy/global-consultation-on-global-leprosy-strategy-2021-2030/06-global-leprosy-strategy-overview.pdf. Accessed 26 Aug 2021.

  28. WHO. Global consultation of National Leprosy Programme managers, partners and affected persons on Global Leprosy Strategy 2021–2030. New Delhi: World Health Organization, Regional Office for South-East Asia; 2020.

    Google Scholar 

  29. Steinmann P, Dusenbury C, Addiss D, Mirza F, Smith WCS. A comprehensive research agenda for zero leprosy. Infect Dis Poverty. 2020;9:156–63.

    PubMed  PubMed Central  Google Scholar 

  30. Notomi T, Okayama H, Masubuchi H, Yonekawa T, Watanabe K, Amino N, et al. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 2000. https://doi.org/10.1093/nar/28.12.e63.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Ablordey A, Amissah DA, Aboagye IF, Hatano B, Yamazaki T, Sata T, et al. Detection of Mycobacterium ulcerans by the loop mediated isothermal amplification method. PLoS Negl Trop Dis. 2012. https://doi.org/10.1371/journal.pntd.0001590.

    Article  PubMed  PubMed Central  Google Scholar 

  32. de Souza DK, Quaye C, Mosi L, Addo P, Boakye DA. A quick and cost effective method for the diagnosis of Mycobacterium ulcerans infection. BMC Infect Dis. 2012. https://doi.org/10.1186/1471-2334-12-8.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Beissner M, Phillips RO, Battke F, Bauer M, Badziklou K, Sarfo FS, et al. Loop-mediated isothermal amplification for laboratory confirmation of buruli ulcer disease-towards a point-of-care test. PLoS Negl Trop Dis. 2015. https://doi.org/10.1371/journal.pntd.0004219.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Notomi T, Mori Y, Tomita N, Kanda H. Loop-mediated isothermal amplification (LAMP): principle, features, and future prospects. J Microbiol. 2015;53:1–5.

    CAS  PubMed  Google Scholar 

  35. Knauf S, Lüert S, Smajs D, Strouhal M, Chuma IS, Frischmann S, et al. Gene target selection for loop-mediated isothermal amplification for rapid discrimination of Trepanema pallidum subspecies. PloS Negl Trop Dis. 2018. https://doi.org/10.1371/journal.pntd.0006396.

    Article  PubMed  PubMed Central  Google Scholar 

  36. Deng MH, Zhong LY, Kamolnetr O, Limpanont Y, Lv ZY, et al. Detection of helminths by loop-mediated isothermal amplification assay: a review of updated technology and future outlook. Infect Dis Poverty. 2019. https://doi.org/10.1186/s40249-019-0530-z.

    Article  PubMed  PubMed Central  Google Scholar 

  37. Garcia-Bernalt Diego J, Fernández-Soto P, Crego-Vicente B, Alonso-Castrillejo S, Febrer-Sendra B, Gómez-Sánchez A, et al. Progress in loop-mediated isothermal amplification assay for detection of Schistosoma mansoni: towards a ready-to-use-test. Sci Rep. 2019. https://doi.org/10.1038/s41598-019-51342-2.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Nzelu CO, Kato H, Peters NC. Loop-mediated isothermal amplification (LAMP): an advanced molecular point of care technique for the detection of Leishmania infection. PLoS Negl Trop Dis. 2019. https://doi.org/10.1371/journal.pntd.0007698.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Gilberto SNB, Walter LBJ, Amanda VBV, Amanda TX, Sebastião Da Costa Lima Júnior M, Edeneide MX, et al. Loop-mediated isothermal amplification methods for diagnosis of visceral leishmaniasis (kala-azar)—a systematic review. Expert Rev Mol Diagn. 2020;20:455–65.

    Google Scholar 

  40. Njiru ZK, Mikosza AS, Armstrong T, Enyaru JC, Ndungu JM, Thompson AR. Loop-mediated isothermal amplification (LAMP) method for rapid detection of Trypanosoma brucei rhodesiense. PLoS Negl Trop Dis. 2008. https://doi.org/10.1371/journal.pntd.0000147.

    Article  PubMed  PubMed Central  Google Scholar 

  41. Gray CM, Katamba A, Narang P, Giraldo J, Zamudio C, Joloba M, et al. Feasibility and operational performance of tuberculosis detection by loop-mediated isothermal amplification platform in decentralized settings: results of a multicenter study. J Clin Microbiol. 2016;54:1984–91.

    CAS  PubMed  PubMed Central  Google Scholar 

  42. Hopkins H, Gonález IJ, Polley SD, Angutoko P, Ategeka J, Asilmwe C, et al. Highly sensitive detection of malaria parasitemia in a malaria-endemic setting: performance of a new loop-mediated isothermal amplification kit in a remote clinic in Uganda. J Infect Dis. 2013;208:645–52.

    CAS  PubMed  PubMed Central  Google Scholar 

  43. Polley SD, González IJ, Mohamed D, Daly R, Bowers K, Watson J, et al. Clinical evaluation of a loop-mediated amplification kit for diagnosis of imported malaria. J Infect Dis. 2013;208:637–44.

    CAS  PubMed  PubMed Central  Google Scholar 

  44. Picot S, Cucherat M, Bienvenu AL. Systematic review and meta-analysis of diagnostic accuracy of loop-mediated isothermal amplification (LAMP) methods compared with microscopy, polymerase chain reaction and rapid diagnostic tests for malaria diagnosis. Int J Infect Dis. 2020. https://doi.org/10.1016/j.ijid.2020.07.009.

    Article  PubMed  Google Scholar 

  45. Ibarra-Meneses A, Cruz I, Chicharri C, Sánchez C, Biéler S, Broger T, et al. Evaluation of fluorimetry and direct visualization to interpret results of a loop-mediated isothermal amplification kit to detect Leishmania DNA. Parasit Vectors. 2018. https://doi.org/10.1186/s13071-018-2836-2.

    Article  PubMed  PubMed Central  Google Scholar 

  46. Mukhtar M, Ali SS, Boshara SA, Albertini A, Monnerat S, Bessell P, et al. Sensitive and less invasive confirmatory diagnosis of visceral leishmaniasis in Sudan using loop-mediated isothermal amplification (LAMP). PLoS Negl Trop Dis. 2018. https://doi.org/10.1371/journal.pntd.0006264.

    Article  PubMed  PubMed Central  Google Scholar 

  47. Lucchi NW, Gaye M, Diallo MA, Goldman IF, Ljolje D, Deme AB, et al. Evaluation of the Illumigene malaria LAMP: a robust molecular tool for malaria parasites. Sci Rep. 2016. https://doi.org/10.1038/srep36808.

    Article  PubMed  PubMed Central  Google Scholar 

  48. Rypien C, Chow B, Chan WW, Church DL, Pillai DR. Detection of Plasmodium infection by the illumigene malaria assay compared to reference microscopy and real-time PCR. J Clin Microbiol. 2017;55:3037–45.

    CAS  PubMed  PubMed Central  Google Scholar 

  49. PrimerExplorer: LAMP primer designing software. http://primerexplorer.jp/e. Accessed 26 Aug 2021.

  50. Zhang X, Lowe SB, Gooding JJ. Brief review of monitoring methods for loop-mediated isothermal amplification (LAMP). Biosens Bioelectron. 2014;61:491–9.

    CAS  PubMed  Google Scholar 

  51. Garg N, Sahu U, Kar S, Ahmad FJ. Development of a Loop-mediated isothermal amplification (LAMP) technique for specific and early detection of Mycobacterium leprae in clinical samples. Sci Rep. 2021. https://doi.org/10.1038/s41598-021-89304-2.

    Article  PubMed  PubMed Central  Google Scholar 

  52. Jiang H, Tsang L, Wang H, Liu C. Loop-mediated isothermal amplification (LAMP) assay targeting RLEP for detection of Mycobacterium leprae in leprosy patients. Int J Infect Dis. 2021;107:145–52.

    CAS  PubMed  Google Scholar 

  53. Joshi S, Sharma V, Ramesh V, Singh R, Salotra P. Development of a novel loop-mediated isothermal amplification assay for rapid detection of Mycobacterium leprae in clinical samples. Indian J Dermatol Venereol Leprol. 2021;87:491–7.

    PubMed  Google Scholar 

  54. Joshi S, Dixit KK, Sharma V, Ramesh V, Singh R, Salotra P. Rapid multiplex loop-mediated isothermal amplification (m-LAMP) assay for differential diagnosis of leprosy and Post-Kala-Azar dermal leishmaniasis. Am J Trop Med Hyg. 2021;104:2085–90.

    Google Scholar 

  55. Beissner M, de Souza DK, Maman I, Saar M, Zwirglmaier K, Wiedemann F et al. Dry Reagent-Based Loop-Mediated Isothermal Amplification for Confirmation of Buruli Ulcer Using a Portable Fluorometer. In: ECTMIH 2017—10th European congress on tropical medicine and international health. 16.10.2017–20.10.2017 Oral presentation. 2017.

  56. Benson DA, Karsch-Mizrachi I, Lipman DJ, Ostell J, Sayers EW. GenBank. Nucleic Acids Res. 2008. https://doi.org/10.1093/nar/gkn723.

    Article  PubMed  PubMed Central  Google Scholar 

  57. OptiGene: Genie® III. http://www.optigene.co.uk/instruments/instrument-genie-iii. Accessed 26 Aug 2021.

  58. MPT: Accelerated Aging. https://lso-inc.com/medical-package-testing/services/accelerated-aging. Accessed 26 Aug 2021.

  59. Groenen G, Saunderson P, Ji B. ILEP learning guide three: how to do a skin smear examination for leprosy. London: International Federation of Anti-Leprosy Associations; 2003.

    Google Scholar 

  60. Beissner M, Huber KL, Badziklou K, Halatoko WA, Maman I, Vogel F, et al. Implementation of a national reference laboratory for Buruli ulcer disease in Togo. PLoS Negl Trop Dis. 2013. https://doi.org/10.1371/journal.pntd.0002011.

    Article  PubMed  PubMed Central  Google Scholar 

  61. Fyfe JAM, Lavender CJ, Johnson PDR, Globan M, Sievers A, Azuoloas J, et al. Development and application of two multiplex real-time PCR assays for the detection of Mycobacterium ulcerans in clinical and environmental samples. Appl Environ Microbiol. 2007;73:4733–40.

    CAS  PubMed  PubMed Central  Google Scholar 

  62. Beissner M, Symank D, Phillips RO, Amoako YA, Awua-Boateng N-Y, Sarfo FS, et al. Detection of viable Mycobacterium ulcerans in clinical samples by a novel combined 16S rRNA reverse transcriptase/IS2404 real-time qPCR assay. PLoS Negl Trop Dis. 2012. https://doi.org/10.1371/journal.pntd.0001756.

    Article  PubMed  PubMed Central  Google Scholar 

  63. Schönian G, Nasereddin A, Dinse N, Schweynoch C, Schallig HDFH, Presber W, Jaffe CL. PCR diagnosis and characterization of Leishmania in local and imported clinical samples. Diagn Microbiol Infect Dis. 2003;47:349–58.

    PubMed  Google Scholar 

  64. Taylor S, Wakem M, Dijkman G, Alsarraj M, Nguyen M. A practical approach to RT-qPCR—publishing data that conform to the MIQE guidelines. Methods. 2010. https://doi.org/10.1016/j.ymeth.2010.01.005.

    Article  PubMed  Google Scholar 

  65. Niemz A, Boyle DS. Nucleic acid testing for tuberculosis at the point-of-care in high-burden countries. Expert Rev Mol Diagn. 2014;12:687–701.

    Google Scholar 

  66. Bretzel G, Beissner M. PCR detection of Mycobacterium ulcerans-significance for clinical practice and epidemiology. Expert Rev Mol Diagn. 2018;1:1–12.

    Google Scholar 

  67. Manta FSN, Barbieri RR, Moreira SJM, Santos PTS, Nery JAC, Duppre NC, et al. Quantitative PCR for leprosy diagnosis and monitoring in household contacts: a follow-up study, 2011–2018. Sci Rep. 2019. https://doi.org/10.1038/s41598-019-52640-5.

    Article  PubMed  PubMed Central  Google Scholar 

  68. Gama RS, de Souza MLM, Sarno EN, de Moraes MO, Goncalves A, Stefani MMA, et al. A novel integrated molecular and serological analysis method to predict new cases of leprosy amongst household contacts. PLoS Negl Trop Dis. 2019. https://doi.org/10.1371/journal.pntd.0007400.

    Article  PubMed  PubMed Central  Google Scholar 

  69. WHO. Global leprosy strategy: accelerating towards a leprosy-free world. Geneva: World Health Organization, Regional Office for South-East Asia; 2016.

    Google Scholar 

  70. Smith CS, Noordeen SK, Richardus JH, Sansarricq H, Cole ST, Soares RC, et al. A strategy to halt leprosy transmission. Lancet Infect Dis. 2014;14:96–8.

    PubMed  Google Scholar 

  71. Blok DJ, de Vlas SJ, Richardus JH. Global elimination of leprosy by 2020: are we on track? Parasit Vectors. 2015. https://doi.org/10.1186/s13071-015-1143-4.

    Article  PubMed  PubMed Central  Google Scholar 

  72. Richardus JH, Tiwari A, Barth-Jaeggi T, Arif MA, Banstola NL, Rabindra B, et al. Leprosy post-exposure prophylaxis with single-dose rifampicin (LPEP): an international feasibility programme. Lancet Glob Health. 2020. https://doi.org/10.1016/S2214-109X(20)30396-X.

    Article  PubMed  Google Scholar 

  73. WHO. Ending the neglect to attain the Sustainable Development Goals: a road map for neglected tropical diseases 2021–2030. Geneva: World Health Organization; 2020.

    Google Scholar 

  74. Kettler H, White K, Hawkes S. Mapping the landscape of diagnostics for sexually transmitted infections—key findings and recommendations. World Health Organization on behalf of the Special Programme for Research and Training in Tropical Diseases. 2004.

  75. Mabey D, Peeling RW, Ustianowski A, Perkins MD. Diagnostics for the developing world. Nat Rev Microbiol. 2004;2:231–40.

    CAS  PubMed  Google Scholar 

  76. Loop-Mediated NZK, Technology IA. Towards point of care diagnostics. PLoS Negl Trop Dis. 2012. https://doi.org/10.1371/journal.pntd.0001572.

    Article  Google Scholar 

  77. Lee D, La Mura M, Allnutt TR, Powell W. Detection of genetically modified organisms (GMOs) using isothermal amplification of target DNA sequences. BMC Biotechnol. 2009;9:7.

    PubMed  PubMed Central  Google Scholar 

  78. Wong YP, Othman S, Lau YL, Radu S, Chee HY. Loop-mediated isothermal amplification (LAMP): a versatile technique for detection of microorganisms. J Appl Microbiol. 2018;124:626–43.

    CAS  PubMed  Google Scholar 

  79. Bi A, Nakajima C, Fukushima Y, Tamaru A, Sugawara I, Kimura A, et al. A rapidloop-mediated isothermal amplification assay targeting hspX for the detection of Mycobacterium tuberculosis complex. Jpn J Infect Dis. 2012;65:247–51.

    CAS  PubMed  Google Scholar 

  80. Dhama K, Karthik K, Chakraborty S, Tiwari R, Kapoor S, Kumar A, Thomas P. Loop-mediated isothermal amplification of DNA (LAMP): a new diagnostic tool lights the world of diagnosis of animal and human pathogens: a review. Pak J Biol Sci. 2014;17:151–66.

    CAS  PubMed  Google Scholar 

  81. Fischbach J, Xander NC, Frohme M, Glokler JF. Shining a light on LAMPassays-a comparison of LAMP visualization methods including the noveluse of berberine. Biotechniques. 2015;58:189–94.

    CAS  PubMed  Google Scholar 

  82. Wu G, Zaman MH. Low-cost tools for diagnosing and monitoring HIV infection in low-resource settings. Bull World Health Organ. 2012;90:914–20.

    PubMed  PubMed Central  Google Scholar 

  83. Carrel C, Kava A, Nguyen M, Menger R, Munshi Z, Call Z, et al. Beyond the lateral flow assay: a review of paper-based microfluidics. Microelectron Eng. 2019;206:45–54.

    Google Scholar 

  84. Hayashida K, Kajino K, Hachaambwa L, Namangala B, Sugimoto C. Direct blood dry LAMP: a rapid, stable, and easy diagnostic tool for human African trypanosomiasis. PLoS Negl Trop Dis. 2015. https://doi.org/10.1371/journal.pntd.0003578.

    Article  PubMed  PubMed Central  Google Scholar 

  85. Hayashida K, Kajino K, Simukoko H, Simuunza M, Ndebe J, Chota A, et al. Direct detection of falciparum and non-falciparum malaria DNA from a drop of blood with high sensitivity by the dried-LAMP system. Parasit Vectors. 2017. https://doi.org/10.1186/s13071-016-1949-8.

    Article  PubMed  PubMed Central  Google Scholar 

  86. Salim B, Hayashida K, Mossaad E, Nakao R, Yamagishi J, Sugimoto C. Development and validation of direct dry loop mediated isothermal amplification for diagnosis of Trypanosoma evansi. Vet Parasitol. 2018. https://doi.org/10.1016/j.vetpar.2018.08.009.

    Article  PubMed  Google Scholar 

  87. Chen HW, Weissenberger G, Ching WM. Development of lyophilized loop-mediated isothermal amplification reagents for the detection of leptospira. Mil Med. 2016;181:227–31.

    PubMed  Google Scholar 

  88. Carter C, Akrami K, Hall D, Smith D, Aronoff-Spencer E. Lyophilized visually readable loop-mediated isothermal reverse transcriptase nucleic acid amplification test for detection Ebola Zaire RNA. J Virol Methods. 2017;244:32–8.

    CAS  PubMed  PubMed Central  Google Scholar 

  89. Howson ELA, Armson B, Madi M, Kasanga CJ, Kandusi S, Sallu R, et al. Evaluation of two lyophilized molecular assays to rapidly detect foot-and-mouth disease virus directly from clinical samples in field settings. Transbound Emerg Dis. 2017. https://doi.org/10.1111/tbed.12451.

    Article  PubMed  PubMed Central  Google Scholar 

  90. Engku Nur Syafirah EAR, Najian ABN, ChongFoo P, Ali MRM, Mohamed M, YeanYean C. An ambient temperature stable and ready-to-use loop-mediated isothermal amplification assay for detection of toxigenic Vibrio cholerae in outbreak settings. Acta Trop. 2018;182:223–31.

    CAS  PubMed  Google Scholar 

  91. Scollard D and Gillis TP. International textbook of leprosy. 2019. Online edition. https://www.internationaltextbookofleprosy.org. Accessed 26 Aug 2021.

  92. Pardillo FEF, Fajardo TT, Abalos RM, Scollard D, Gelber RH. Methods for the classification of leprosy for treatment purposes. CID. 2007;44:1096–9.

    Google Scholar 

  93. Parkash O. Classification of leprosy into multibacillary and paucibacillary groups: an analysis. FEMS Immunol Med Microbiol. 2009;55:1–9.

    CAS  PubMed  Google Scholar 

  94. Shepard CC, Mc Rae DH. A method for counting acid-fast bacteria. Int J Lepr. 1968;36:78–82.

    CAS  Google Scholar 

  95. Rudeeaneksin J, Srisungngam S, Sawanpanyalert P, Sittiwakin T, Likanonsakul S, Pasadorn S, Palittapongarnpim P, Brennan PJ, Phetsuksiri B. LightCyclerTM real-time PCR for rapid detection and quantitation of Mycobacterium leprae in skin specimens. FEMS Immunol Med Microbiol. 2008;54:263–70.

    CAS  PubMed  Google Scholar 

  96. Bretzel G, Huber KL, Kobara B, Beissner M, Piten E, Herbinger KH, et al. Laboratory confirmation of Buruli ulcer disease in Togo, 2007–2010. PLoS Negl Trop Dis. 2011. https://doi.org/10.1371/journal.pntd.0001228.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

This study was made possible thanks to the Memento Prize 2014 awarded by Doctors without Borders, the Buko Pharma Campaign and the German Leprosy and Tuberculosis Relief Association. The authors are grateful to Kerstin Helfrich and Simone Lehn (DITM, LMU) for excellent technical assistance. We thank Richard Odame Phillips (KCCR, KNUST) for providing DNA extracts from IS2404 qPCR confirmed BUD patients from Ghana. Furthermore, the authors thank Franz Wiedemann (GLRA-T) for excellent logistic assistance from Togo. The article contains parts of the doctoral dissertation of Malkin Saar.

Funding

Open Access funding enabled and organized by Projekt DEAL. This work was funded through the Memento Prize 2014 awarded by Doctors without Borders, the Buko Pharma Campaign and the German Leprosy and Tuberculosis Relief Association and was also supported by funds from the German Leprosy and Tuberculosis Relief Association, Würzburg, Germany (ILEP Project No.: 7.01.02.54). The funders had no role in the design of the study and collection, analysis and interpretation of data as well as in the writing of the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

GB, MB, MS, IM and KHH were involved in the conceptional design of this study. MB designed the LAMP primers and developed (DRB) LAMP validation protocols together with MS. MS and FG conducted molecular laboratory methods and MS, FG and KHH analyzed the data critically. GB, MB, MS, IM, FG and KHH were involved in the interpretation of data. GB conceptionally designed the manuscript. The manuscript was written by GB, MS and MB. All authors revised the manuscript critically and contributed to the final draft. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Malkin Saar.

Ethics declarations

Ethics approval and consent to participate

All methods were carried out in accordance with relevant guidelines and regulations. The study involves human tissue from Togolese, Ghanaian and German patients.

For Togolese and Ghanaian patients, samples were collected in the course of two studies, approved by the National Togolese (012/2012/CBRS) and the Ghanaian KNUST (NMIMR-IRB CPN 116/12-13) Ethics Committees. Written IC was obtained from all Togolese and Ghanaian study participants and/or their legal representatives if below 18 years of age.

All German patients sought medical advice at our outpatient department and before subjection to routine clinical and laboratory procedures, they signed an informed consent form for the collection of data and samples which has been approved by the institutional ethics review board of the Medical Centre of the University of Munich (LMU).

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

In-house “wet” RLEP LAMP run protocol.

Additional file 2.

In-house lyophilization protocol.

Additional file 3.

In-house RLEP DRB LAMP run protocol.

Additional file 4.

Ready-to-use RLEP DRB LAMP run protocol.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Saar, M., Beissner, M., Gültekin, F. et al. RLEP LAMP for the laboratory confirmation of leprosy: towards a point-of-care test. BMC Infect Dis 21, 1186 (2021). https://doi.org/10.1186/s12879-021-06882-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12879-021-06882-2

Keywords