Skip to main content

Table 3 Results of the LAMP optimization

From: RLEP LAMP for the laboratory confirmation of leprosy: towards a point-of-care test

 

Cq value [cycles]

Melting temperature [°C]

Mean

SD

CV

Mean

SD

CV

Test 1: master mix

 ISO-DR-001

21.78

0.13

0.006

91.90

0.00

0.000

 ISO-DR-004

14.59

0.11

0.007

91.95

0.07

0.001

Test 2: primer ratio

 1:1

17.05

0.01

0.001

92.00

0.00

0.000

 2:1

12.77

0.51

0.040

92.00

0.00

0.000

Test 3: annealing temperature (°C)

 68.6

18.64

3.42

0.183

92.20

0.28

0.003

 68.1

19.21

92.00

 67.0

19.15

2.42

0.126

92.15

0.21

0.002

 65.0

15.97

0.74

0.047

92.00

0.00

0.000

 62.6

19.24

0.89

0.046

91.95

0.07

0.001

 60.6

21.76

0.98

0.045

92.00

0.14

0.002

 59.3

24.26

0.59

0.024

91.95

0.07

0.001

 58.6

28.41

1.81

0.064

91.85

0.07

0.001