Skip to main content

Table 7 Comparison of the semi-quantitative results of RLEP DRB LAMP formats (in-house and ready-to-use)

From: RLEP LAMP for the laboratory confirmation of leprosy: towards a point-of-care test

 

High positive samples of MB leprosy patients

Medium positive samples of MB leprosy patients

Low positive samples of MB leprosy patients

Mean

SD

CV

Mean

SD

CV

Mean

SD

CV

In-house RLEP DRB LAMP

 Tp [mm:ss]

16:30

02:00

0.112

18:00

01:15

0.075

21:45

01:45

0.085

Ready-to-use RLEP DRB LAMP

 Tp [mm:ss]

24:15

03:00

0.113

27:00

02:00

0.068

28:45

00:45

0.026