The evaluation of a rapid microfluidic immunofluorescence antigen test in detecting the infectiousness of COVID-19 patients

Abstract

Background

A test-based strategy against coronavirus diseaseĀ 2019 (COVID-19) is one of the measures to assess the need for isolation and prevention of infection. However, testing with high sensitivity methods, such as quantitative RT-PCR, leads to unnecessary isolation, whereas the lateral flow antigen test shows low sensitivity and false negative results. The purpose of this study was to evaluate the performance of the LumiraDx SARS-CoV-2 Ag test (Lumira Ag), a rapid microfluidic immunofluorescence method, in assessing infectivity.

Methods

This study was performed from March 2022 to July 2022. A pair of nasopharyngeal swab samples were obtained from each patient with mild COVID-19. One swab was used for Lumira Ag testing, and the other for quantitative RT-PCR testing and virus culture.

Results

A total of 84 patients were included in the study. Among them, PCR, Lumira Ag test, and virus culture indicated positivity for 82, 66, and 24 patients, respectively. When comparing the Lumira Ag test to virus culture, its sensitivity was 100.0% (24/24), specificity, 30.0% (18/60); positive predictive value, 36.3% (24/66); and negative predictive value (NPV), 100.0% (18/18). The positive sample for virus culture was observed until the ninth day from the onset of symptoms, while the Lumira Ag test was observed until day 11.

Conclusions

The Lumira Ag test showed high sensitivity and NPV (100% each) compared to virus culture. A test-based strategy using the Lumira Ag test can effectively exclude COVID-19 infectiousness.

Peer Review reports

Background

Due to the continuation of coronavirus diseaseĀ 2019 (COVID-19) for three years, a uniformly required duration of quarantine for patients is being reconsidered. As an alternative to quarantine, a test-based strategy against COVID-19 is a widely accepted measure to assess the need for isolation and infection prevention [1, 2]. Using this strategy, individuals with negative test results are allowed to end their isolation or remove their masks. For the specific purpose of excluding potential infectiousness, high sensitivity in detecting infectious individuals is an essential characteristic of testing. Furthermore, the infectiousness of the Omicron variant assessed by viral culture has been reported even 5Ā days after the onset of symptoms [3,4,5], reinforcing the need for an efficient test-based strategy.

However, nucleic acid amplification testing, including RT-PCR, has a high sensitivity for SARS-CoV-2 regardless of its viability, leading to continuous positive test results for several weeks [6, 7]. Additionally, maintaining a turn-around time of several hours is difficult; therefore, RT-PCR as point-of-care testing is challenging. In contrast, the lateral flow antigen test shows lower sensitivity for viral culture-positive samples [8,9,10], with potentially false negative results in assessing infectiousness.

The LumiraDx SARS-CoV-2 Ag test (Lumira Ag, LumiraDX UK Ltd., Dumyat, UK) is a rapid antigen test that offers a qualitative test result using a rapid microfluidic immunofluorescence method, which is available as point-of-care testing. Previous studies reported that the Lumira Ag test exhibited higher clinical sensitivity and specificity than RT-PCR [11]. However, the performance of the Lumira Ag test in assessing infectiousness in a test-based strategy is not clear. In this study, we focused on the highly sensitive performance of Lumira Ag testing and aimed to evaluate the performance of the Lumira Ag test for assessing the infectivity of the SARS-CoV-2 Omicron variant.

Methods

Sample collection

A prospective study was conducted from Marchā€“July 2022 to assess the performance of Lumira Ag in detecting SARS-CoV-2 infectivity. According to the surveillance data, the dominance of Omicron variants was greater than 99.67% during the study period [12]. The study was approved by the Institutional Review Board of Nagasaki University Hospital(Approval Number: 22022126). Patients with COVID-19, diagnosed using nucleic acid amplification or antigen tests and staying in recovery accommodation facilities in Nagasaki city, were enrolled in the study. After informed consent was acquired from patients with mild COVID-19, a pair of nasopharyngeal swab samples were obtained and immediately sent to the laboratory. The swabs were obtained from the same naris by a trained physician. One swab was used for Lumira Ag testing, and the other was stored in virus transport media (VTM) and used for quantitative RT-PCR testing and viral culture.

Lumira Ag, RT-PCR, and viral culture

Lumira Ag testing was performed according to the instructions of the manufacturer. Briefly, the swab was placed into the extraction buffer, which was applied to the test strip inserted in the LumiraDx instrument. The qualitative result was obtained within 12Ā min.

Quantitative RT-PCR testing was performed according to the National Institute for Infectious Diseases guidelines [13]. Nucleic acid was extracted and purified from VTM using the MagMAX Viral/Pathogen nucleic acid isolation kit (Thermo Fisher Scientific), following the protocol of the manufacturer. For RT-PCR, 5 Ī¼L of the RNA template was tested using real-time RT-PCR primer/probe sets for 2019-nCoV_N2 [13]. PCR was conducted using the Thunderbird probe one-step qRT-PCR kit (TOYOBO) and the QuantStudio 6 Pro real-time PCR system (Thermo Fisher Scientific). The cycle threshold (Ct) value was used to indicate viral samples.

The virus was isolated to check infectivity using swab samples stored in VTM. A 100 ĀµL aliquot of each VTM was inoculated in Vero E6 cells cultured in Eagle's medium supplemented with 2% fetal calf serum and 1% penicillin or streptomycin solution. Infected cells were cultured at 37Ā Ā°C and observed for cytopathic effects daily. The infected culture fluid (ICF) was collected 5Ā days after infection for the first passage. The second passage of the viral culture was performed using a fresh monolayer of Vero E6 cells following the same procedure used in the first passage of the virus culture. Approximately 200 ĀµL second passage of ICF was used for viral RNA extraction by the Nextractor NX-48 robot using the NX-48S Viral NA Kit (Genolution Inc.). The presence of the virus in ICF was verified using quantitative real-time RT-PCR. The cytopathic effect in viral culture was assessed and referred to as infectivity [8, 14,15,16].

Viral load comparison between the Lumira Ag buffer and VTM as a matrix for PCR testing

To compare the viral load between paired samples and assess the Lumira Ag buffer and VTM as a matrix for the nucleic acid amplification test, the residual buffer from Lumira Ag testing was subsequently used for quantitative RT-PCR testing. The quantitated viral loads were expressed in terms of copies/test and compared on a scatter plot.

Statistical analysis

Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of the Lumira Ag test to viral culture were calculated. The unpaired t-test was performed to compare the viral load between groups. The level of statistical significance was set at Pā€‰<ā€‰0.05.

Results

A total of 84 patients with COVID-19 (mean age: 42.1Ā years, male: 54.8%) who were between 2 and 13Ā days from the onset of symptoms were included in the study. Among them, 82 patients underwent RT-PCR tested positive and subsequently proceeded for further analysis. Lumira Ag showed positive results for 66 of the participants, and viral culture for 24 (Additional file 1). The positive samples for RT-PCR were observed until day 13 from the onset of symptoms, the Lumira Ag test until day 11, and viral culture until day 9 (Fig.Ā 1 and Additional file 2).

Fig. 1
figure 1

The results of Lumira Ag testing, viral culture, and RT-PCR testing. Qualitative results for Lumira Ag testing and viral culture, and quantitative results for RT-PCR testing (Ct value) are shown. The relationships between the test results and the days after the onset of symptoms are shown. The results of Lumira Ag testing are shown as filled circles for positive results and empty circles for negative results. The viral culture results, representing infectiousness, are shown as red for positive and blue for negative. NEG, negative; POS, positive

The comparison of the Lumira Ag test to viral culture is shown in Table 1. The sensitivity was 100% (24/24), specificity was 27.6% (16/58), PPV was 36.3% (24/66), and NPV was 100% (16/16). No samples were positive in viral culture among the samples with negative results in the Lumira Ag test.

Table 1 Comparison of the results of nasopharyngeal swab sample testing by Lumira Ag test and viral culture

Relative to viral culture, the Lumira Ag test sensitivity was 100% (24/24), specificity was 30% (18/60), positive predictive value was 36.3% (24/66); and negative predictive value was 100% (18/18). NEG, negative; POS, positive.

When comparing the samples that tested positive by Lumira Ag with those that tested negative, the Ct value determined using RT-PCR was significantly lower in the positive samples (mean, 22.6 vs. 31.5; Pā€‰<ā€‰0.0001) (Fig.Ā 2a). Among the samples with a Ct valueā€‰ā‰¤ā€‰30, the positivity for Lumira Ag was 93.9% (62/66). In addition, a significantly higher Ct value was observed in viral culture-positive samples than in viral culture-negative ones (19.3 vs. 26.4; Pā€‰<ā€‰0.0001; Fig.Ā 2b). When the cut-off threshold of the Ct value to viral culture was set at 30.0, the positive predictive value was calculated to be 36.3% (24/66), and the negative predictive value was 100.0% (16/16).

Fig. 2
figure 2

Lumira Ag testing and RT-PCR analysis. a Viral load determined using RT-PCR in Lumira Ag-positive and -negative groups. b Viral load determined using RT-PCR was compared in viral culture-positive and -negative groups. A t-test was applied to assess the differences between groups. ****, pā€‰<ā€‰0.0001; ns, not significant; NEG, negative; POS, positive

To assess the variation between the paired swabs, the viral load determined by RT-PCR was compared (see Additional file 3), and the correlation coefficient was 0.8136.

Discussion

In this study, the results of the Lumira Ag test and viral culture were compared and the results were correlated with the viral load determined using RT-PCR. The Lumira Ag test had high sensitivity and NPV relative to viral culture, favoring it as a measure of infectivity.

Nucleic acid amplification testing, including RT-PCR, has been accepted as the gold standard for testing for COVID-19 because of its high sensitivity and specificity. Though almost all clinical research and diagnosis are based on the results of RT-PCR, false-negative results are occasionally observed despite its high analytical sensitivity. A systematic review by Pecoraro et al. [17] reported a wide variation of false-negative rates (2%ā€“58%), and a summary estimate of the overall false-negative rate of 12%. Several possible causes for false-negatives include pre-analytical errors such as quality of clinical specimens [18], sampling procedure, swab type, sample container, as well as the sampling having occurred early in the course of infection, which results in an extremely small amount of virus particles due to the early phase of infection [19]. Thus, clinicians and laboratories need to be aware of the possibility of false-negative results affected by clinical and laboratorial variances, and pre-test probability must be considered in interpreting test results. Another important characteristic of RT-PCR is that it can provide quantitative results based on the cycle threshold (Ct) value, which could be a surrogate of infectivity [20]. However, inter-laboratory variation makes it difficult to apply RT-PCR to assess infectiousness [21]. Additionally, previous studies have reported positive results of RT-PCR several weeks after the diagnosis [22,23,24]. Multiple samples with a high viral load ofā€‰>ā€‰104/test were observed even 10Ā days after symptom onset when no culturable virus was detected. Therefore, RT-PCR has limited practicality in predicting infectiousness.

Rapid antigen testing, such as the lateral flow test, is easy to perform and is widely used as point-of-care testing by healthcare and non-healthcare providers. Therefore, the lateral flow test is a representative test intended to be utilized as part of a test-based strategy, particularly where immediate decisions are required. However, it is well proven that the target population who are to be tested has a great impact on the sensitivity of rapid antigen testing. In the Cochrane review [25], the sensitivity was reported to be 73.0% in symptomatic patients and further increased to 80.9% in those exhibiting symptoms in seven days. On the other hand, the sensitivity remains at 54.7% in asymptomatic patients, with a slightly higher sensitivity of 64.3% in those with previous contact with the patients. To compensate for its low sensitivity, repeated use of rapid antigen testing is suggested and the effectiveness of this method is supported by several clinical studies [26, 27]. However, as an assay format, lateral flow assay shows relatively lower sensitivity (61.4% for alkaline phosphatase-labeled antibodies and 81.3% for latex-conjugated) compared to microfluidic fluorescent immunoassay (89.7%) [25]. Accordingly, for the assessment of infectiousness, a previous study by Kirby et al. [8] reported that the sensitivity of the lateral flow test decreased, which may result in a failure to exclude infectious individuals. The Lumira Ag test is reported to have higher sensitivity than the lateral flow test in detecting SARS-CoV-2 [28, 29], and the results of the present study showed its satisfactory performance in effectively excluding infectious COVID-19 patients. Based on the high NPV of Lumira Ag to viral culture, the exclusion of infectivity in patients with COVID-19 can be proposed as a clinical application. Time-based strategies, which require uniform durations of isolation (e.g. five to ten days), are no longer a practical measure. However, as shown in this study, some patients exhibited infectivity even after five days after the onset of the symptom. To compensate in test-based strategies, highly sensitive diagnostic testing such as Lumira Ag can play an important role in excluding infectiousness and help shorten the isolation period. Since this requires the instrument for testing and must be performed in medical facilities, this application of the use of Lumira Ag testing is suitable for medical staff. By applying this test-based strategy, the isolation period can be shortened and the risk of spreading the infection to the patient and other staff can be minimized. Also, Lumira Ag showed satisfactory performance as a diagnostic test with a positivity of 93.9% for samples with a Ct valueā€‰ā‰¤ā€‰30 (62/66). Consequently, it is an alternative testing device with acceptable sensitivity in clinical settings where nucleic acid amplification tests cannot easily be deployed because of time or manpower constraints.

However, the low specificity for infectiousness observed in this study (30%) raises concerns that positive results of Lumira Ag testing should not be interpreted as evidence for infectivity. The sample with a positive result of Lumira Ag testing was observed on the 11th day after the onset of symptoms and contained no culturable virus.

Viral culture reflects the replication-competent viral particles contained in a clinical specimen and is accepted as a measure to assess infectiousness. Previous clinical studies have defined viral culture as an indicator of infectiousness [16, 30]. However, since viral culture takes several days to report results and requires biosafety level 3 precautions, it is not common to perform viral culture tests in clinical laboratories for the purpose of clinical assessment of infectiousness. The difficulties inherent to viral culture even raise the importance of the evaluation of clinically available testing for assessing infectiousness. Indicators for viral culture include some difficult clinical cases under immunocompromised status with continuous positive viral shedding for weeks and months, when the results of viral culture aid clinical decisions regarding discontinuation of antiviral drug regimens or isolation [21, 22].

The limitations of this study include that the Lumira Ag testing and other tests were performed using different samples obtained from the same patients, which might include some variation in the viral load contained in the samples. To address this issue, the viral load quantitated using RT-PCR was compared between paired samples; a comparable viral load was observed with a correlation coefficient of 0.8136. Therefore, the equivalent quality of paired samples was verified, and the variation was limited.

Conclusion

In conclusion, the Lumira Ag test was compared to viral culture, showing high sensitivity and NPV (100% each) relative to viral culture. These findings imply that the Lumira Ag test can be used as a point-of-care test in test-based strategies, allowing social activities to continue during the pandemic.

Availability of data and materials

The authors confirm that the data supporting the findings of this study are available within the article [and/or] its supplementary materials.

Abbreviations

COVID-19:

Coronavirus diseaseĀ 2019

Ct:

Cycle threshold

ICF:

Infected culture fluid

Lumira Ag:

LumiraDx SARS-CoV-2 Ag test

NPV:

Negative predictive value

PPV:

Positive predictive value

VTM:

Virus transport media

References

  1. Center for Disease Control and Prevention. Isolation and Precautions for People with COVID-19. Vol. 0, Center for Disease Control and Prevention. 2022. p. 7ā€“11. Available from: https://www.cdc.gov/coronavirus/2019-ncov/your-health/isolation.html?CDC_AA_refVal=https%3A%2F%2Fwww.cdc.gov%2Fcoronavirus%2F2019-ncov%2Fyour-health%2Fquarantine-isolation.html. Accessed 16 Nov 2023.

  2. European Centre for Disease Prevention and Control E. Guidance on quarantine of close contacts to COVID-19 cases and isolation of COVID-19 cases, in the current epidemiological situation, 7 January 2022. 2022. 2022;(January 2022). Available from: https://www.ecdc.europa.eu/en/covid-19/prevention-and-control/quarantine-and-isolation.Ā Accessed 16 Nov 2023.

  3. Boucau J, Marino C, Regan J, Uddin R, Choudhary MC, Flynn JP, et al. Duration of Shedding of Culturable Virus in SARS-CoV-2 Omicron (BA.1) Infection. N Engl J Med. 2022;387(3):275ā€“7. Available from: http://www.nejm.org/doi/10.1056/NEJMc2202092.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  4. Keske Ş, GĆ¼ney-Esken G, Vatansever C, Beşli Y, Kuloğlu ZE, Nergiz Z, et al. Duration of infectious shedding of SARS-CoV-2 Omicron variant and its relation with symptoms. Clin Microbiol Infect. 2022;xxxx:7ā€“10.

    Google ScholarĀ 

  5. Takahashi K, Ishikane M, Ujiie M, Iwamoto N, Okumura N, Sato T, et al. Duration of Infectious Virus Shedding by SARS-CoV-2 Omicron Variant-Infected Vaccinees. Emerg Infect Dis. 2022;28(5):998ā€“1001. Available from: https://wwwnc.cdc.gov/eid/article/28/5/22-0197_article.htm.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  6. Xiao AT, Tong YX, Zhang S. Profile of RT-PCR for SARS-CoV-2: A Preliminary Study from 56 COVID-19 Patients. Clin Infect Dis. 2020;71(16):2249ā€“51.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  7. Mina MJ, Parker R, Larremore DB. Rethinking Covid-19 Test Sensitivity ā€” A Strategy for Containment. N Engl J Med. 2020;383(22):e120. Available from: nejm.org.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  8. Kirby JE, Riedel S, Dutta S, Arnaout R, Cheng A, Ditelberg S, et al. Sars-Cov-2 antigen tests predict infectivity based on viral culture: comparison of antigen, PCR viral load, and viral culture testing on a large sample cohort. Clin Microbiol Infect. 2023;29(1):94ā€“100. Available from: https://linkinghub.elsevier.com/retrieve/pii/S1198743X22003743.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  9. Lopera TJ, Alzate-Ɓngel JC, DĆ­az FJ, Rugeles MT, Aguilar-JimĆ©nez W. The Usefulness of Antigen Testing in Predicting Contagiousness in COVID-19. Microbiol Spectr. 2022;10(2):1ā€“9.

  10. Currie DW, Shah MM, Salvatore PP, Ford L, Whaley MJ, Meece J, et al. Relationship of SARS-CoV-2 Antigen and Reverse Transcription PCR Positivity for Viral Cultures. Emerg Infect Dis. 2022;28(3):717ā€“20.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  11. Schneider UV, Forsberg MW, Leineweber TD, Jensen CB, Ghathian K, Agergaard CN, et al. A nationwide analytical and clinical evaluation of 44 rapid antigen tests for SARS-CoV-2 compared to RT-qPCR. J Clin Virol. 2022;153(May):1ā€“8.

  12. Mathieu E, Ritchie H, RodƩs-Guirao L, Appel C, Giattino C, Hasell J, Macdonald B, Dattani S, Beltekian D. EO-O and MR. Coronavirus Pandemic (COVID-19). 2020.

    Google ScholarĀ 

  13. National Institute of Infectious Diseases. Manual for the Detection of Pathogen 2019-nCoV Ver.2.6. 2020. p. 1ā€“16. Available from: https://www.niid.go.jp/niid/images/epi/corona/2019-nCoVmanual20200217-en.pdf.Ā Accessed 16 Nov 2023.

  14. Singanayagam A, Patel M, Charlett A, Bernal JL, Saliba V, Ellis J, et al. Duration of infectiousness and correlation with RT-PCR cycle threshold values in cases of COVID-19, England, January to May 2020. Eurosurveillance. 2020;25(32):1ā€“5.

    ArticleĀ  Google ScholarĀ 

  15. Korenkov M, Poopalasingam N, Madler M, Vanshylla K, Eggeling R, Wirtz M, et al. Evaluation of a rapid antigen test to detect sars-cov-2 infection and identify potentially infectious individuals. J Clin Microbiol. 2021;59(9):1ā€“11.

  16. Bullard J, Dust K, Funk D, Strong JE, Alexander D, Garnett L, et al. Predicting infectious severe acute respiratory syndrome coronavirus 2 from diagnostic samples. Clin Infect Dis. 2020;71(10):2663ā€“6.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  17. Pecoraro V, Negro A, Pirotti T, Trenti T. Estimate false-negative RT-PCR rates for SARS-CoV-2. A systematic review and meta-analysis. Eur J Clin Invest. 2022;52(2):1ā€“16.

  18. Vandenberg O, Martiny D, Rochas O, van Belkum A, Kozlakidis Z. Considerations for diagnostic COVID-19 tests. Nat Rev Microbiol. 2021;19(3):171ā€“83.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  19. Kucirka LM, Lauer SA, Laeyendecker O, Boon D, Lessler J. Variation in false-negative rate of reverse transcriptase polymerase chain reactionā€“based SARS-CoV-2 tests by time since exposure. Ann Intern Med. 2020;173(4):262ā€“8.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  20. Tom MR, Mina MJ. To Interpret the SARS-CoV-2 Test, Consider the Cycle Threshold Value. Clin Infect Dis. 2020;71(16):2252ā€“4.

    ArticleĀ  CASĀ  PubMedĀ  Google ScholarĀ 

  21. Evans D, Cowen S, Kammel M, Oā€™Sullivan DM, Stewart G, Grunert HP, et al. The Dangers of Using Cq to Quantify Nucleic Acid in Biological Samples: A Lesson From COVID-19. Clin Chem. 2021;68(1):153ā€“62.

    ArticleĀ  PubMedĀ  Google ScholarĀ 

  22. Avanzato VA, Matson MJ, Seifert SN, Pryce R, Williamson BN, Anzick SL, et al. Case Study: Prolonged Infectious SARS-CoV-2 Shedding from an Asymptomatic Immunocompromised Individual with Cancer. Cell. 2020;183(7):1901-1912.e9. Available from: https://doi.org/10.1016/j.cell.2020.10.049.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  23. Irifune S, Ashizawa N, Takazono T, Mutantu P, Nabeshima T, Ngwe Tun MM, et al. Discrepancy of SARS-CoV-2 PCR results due to the sample collection sites and possible improper sampling. J Infect Chemother. 2021;27(10):1525ā€“8. Available from: https://doi.org/10.1016/j.jiac.2021.07.008.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  24. Chu VT, Schwartz NG, Donnelly MAP, Chuey MR, Soto R, Yousaf AR, et al. Comparison of Home Antigen Testing with RT-PCR and Viral Culture during the Course of SARS-CoV-2 Infection. JAMA Intern Med. 2022;182(7):701ā€“9.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  25. Dinnes J, Sharma P, Berhane S, van Wyk SS, Nyaaba N, Domen J, et al. Rapid, point-of-care antigen tests for diagnosis of SARS-CoV-2 infection. Cochrane Database Syst Rev. 2022;2022(7):31ā€“41. Available from: http://doi.wiley.com/10.1002/14651858.CD013705.pub3.

    Google ScholarĀ 

  26. Love NK, Ready DR, Turner C, Verlander NQ, French CE, Martin AF, et al. Daily use of lateral flow devices by contacts of confirmed COVID-19 cases to enable exemption from isolation compared with standard self-isolation to reduce onward transmission of SARS-CoV-2 in England: a randomised, controlled, non-inferiority trial. Lancet Respir Med. 2022;10(11):1074ā€“85.

    ArticleĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  27. Smith RL, Gibson LL, Martinez PP, Ke R, Mirza A, Conte M, et al. Longitudinal Assessment of Diagnostic Test Performance over the Course of Acute SARS-CoV-2 Infection. J Infect Dis. 2021;224(6):976ā€“82.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

  28. Karon BS, Donato LJ, Bridgeman AR, Blommel JH, Kipp B, Anthony Maus, et al. Analytical sensitivity and specificity of four point of care rapid antigen diagnostic tests for SARS-CoV-2 using real-time quantitative PCR, quantitative droplet digital PCR, and a mass spectrometric antigen assay as comparator methods. Clin Chem. 2021;67:1545ā€“53.

  29. KrĆ¼ger LJ, Klein JAF, Tobian F, Gaeddert M, Lainati F, Klemm S, et al. Evaluation of accuracy, exclusivity, limitof-detection and ease-of-use of LumiraDxTM: an antigen-detecting point-of-care device for SARS-CoV-2. Infection. 2022;50:395ā€“406.

  30. Pekosz A, Parvu V, Li M, Andrews JC, Manabe YC, Kodsi S, et al. Antigen-Based Testing but Not Real-Time Polymerase Chain Reaction Correlates with Severe Acute Respiratory Syndrome Coronavirus 2 Viral Culture. Clin Infect Dis. 2021;73(9):E2861ā€“6.

    ArticleĀ  CASĀ  PubMedĀ  PubMed CentralĀ  Google ScholarĀ 

Download references

Acknowledgements

Not applicable.

Funding

This study was supported by Shionogi & Co., Ltd., Osaka, Japan, and MEXT KAKENHI [Grant Number JP21418576], Grant-in-Aid for Early-Career Scientists.

Author information

Authors and Affiliations

Authors

Contributions

KO obtained the samples, acquired, analysed, and interpreted the data, and was one of the major contributors to manuscript writing. HK, YK, and DS performed the Lumira Ag and RT-PCR tests and interpreted the data. NM and KM performed the viral culture assay. FK, KS, KK, HH, TT, KI, and HM made substantial contributions to the conception and design of the study, drafted the manuscript, and substantively revised it. KY contributed to the funding, management, and supervision of the study. All authors have approved the submitted version (and any substantially modified version that involves the author's contribution to the study) and have agreed to be personally accountable for the author's own contributions and to ensure that questions related to the accuracy or integrity of any part of the work, even those in which the author was not personally involved, are appropriately investigated, resolved, and the resolution documented in the literature.

Corresponding author

Correspondence to Kenji Ota.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Institutional Review Board of Nagasaki University Hospital (Approval Number: 22022126). Informed consent was obtained from all patients. All methods were carried out in accordance with relevant guidelines and regulations in the declaration. No illiterates or minors were involved in the study.

Consent for publication

Not applicable.

Competing interests

KO received financial support from the Government of Japanā€™s Ministry of Education, Culture, Sports, Science, and Technology. KY received financial support from Shionogi & Co., Ltd. No competing was declared by other Authors.

Additional information

Publisherā€™s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

SARS-CoV-2 isolation from swab sample. (A) Vero E6 cell control and (B) Vero E6 cells inoculated with swab sample showing cytopathic effect (CPE) at 4 days post inoculation. Magnification is 10x for all images.

Additional file 2.

The results of Lumira Ag testing, viral culture, and RT-PCR testing shown in copies/test.

Additional file 3.

The viral load of the paired samples as determined by RT-PCR. The X-axis shows the viral load of the samples stored in virus transport media (VTM) used for RT-PCR and viral culture. The Y axis shows the viral load of the samples used for Lumira Ag testing. The viral load was expressed as copies/test. Y = 0.8174x-0.2559, the correlation coefficient was 0.8136.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ota, K., Kodama, H., Kawamoto, Y. et al. The evaluation of a rapid microfluidic immunofluorescence antigen test in detecting the infectiousness of COVID-19 patients. BMC Infect Dis 23, 823 (2023). https://doi.org/10.1186/s12879-023-08821-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12879-023-08821-9

Keywords