Evaluation of different platforms for the detection of anti-SARS coronavirus-2 antibodies, Thailand

Abstract

Background

Antibodies against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) help determine previous infection in individuals, regardless of whether they are asymptomatic or symptomatic. The detection of antibodies serves several purposes, including supporting other assays for disease diagnosis, conducting seroepidemiological studies, and evaluating vaccines. Many platforms of immunological methods for anti-SARS-CoV-2 antibody detection and their performance require validation.

Methods

This study evaluated the test performance of three autoanalyzer-based assays (Architect IgG, Vitros IgG, and Vitros total Ig) and one manual ELISA (Wantai total Ig) against a microneutralization (microNT) assay on the detection of SARS-CoV-2 antibodies. Furthermore, an indirect immunofluorescence assay verified the discordant results between the microNT and commercial assays. The test sensitivity, specificity, positive predictive value, and negative predictive value were determined based on four groups of 1005 serum samples: 102 COVID-19 prepandemic sera, 45 anti-SARS-CoV-2 positive sera, 366 sera of people at risk, and 492 sera of citizens returning from countries with a high prevalence of infection.

Results

The analyses as a whole showed that the performance of these commercial assays was comparable. Each group was also analysed separately to gain further insight into test performance. The Architect did not detect two positive sera of people at risk (prevalence of infection 0.55%). The other methods correctly identified these two positive sera but yielded varying false-positive results. The group of returning travellers with an infection rate of 28.3% (139 of 492) better differentiated the test performance of individual assays.

Conclusions

High-throughput Architect and Vitros autoanalyzers appear appropriate for working on large sample sizes in countries that can afford the cost. The Wantai ELISA, while requiring more individual time and technical skill, may provide reliable results at a lower cost. The selection of assays will depend on the laboratory facilities and feasibility.

Peer Review reports

Background

Almost all immunocompetent individuals infected with severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) develop antibodies specific to multiple viral proteins. In particular, antibodies to nucleoprotein (N) and spike (S1 and S2) proteins are of clinical importance [1,2,3,4]. Specific IgM and IgA first appeared 7–14 days after the onset of disease symptoms, followed by IgG at approximately 14 days. IgM peaks at 2–5 weeks and then declines a few weeks later, while IgG may persist longer [3, 5,6,7,8]. Anti-N antibodies developed before the anti-S antibodies [9, 10]. Various immunological methods have demonstrated the binding activities of these immunoglobulin (Ig) isotypes, e.g., enzyme-linked immunosorbent assay (ELISA), chemiluminescence immunological assays (CLIAs), indirect immunofluorescence (IIF) assay, and immunochromatography. Additionally, plaque reduction neutralization (PRNT) or microNT assays detected functional or neutralizing (NT) antibodies. NT antibodies correlate with protective immunity, while binding antibodies may or may not [3, 8, 10,11,12]. Antibodies to the receptor-binding domain (RBD) in the S1 protein correlated well with NT antibody activity [2, 3, 10, 12].

Antibody detection has served many purposes: supporting the diagnosis of SARS-CoV-2 infection when reverse transcription-polymerase chain reaction (RT-PCR) for viral genomes yields an inconclusive result [3]; serosurveillance to estimate the cumulative incidence of SARS-CoV-2 infection [3, 13,14,15]; and vaccine evaluation, particularly by PRNT or microNT assays.

Currently, multiple commercial kits are available globally, and antigenic targets for these tests include the SARS-CoV-2 S, RBD, and N proteins [1,2,3]. Evaluations of most serological test kits used sera from RT-PCR-confirmed cases as the gold standard for comparison with COVID-19 prepandemic sera [16] or RT-PCR-negative sera [17]. Few studies have included assessments of SARS-CoV-2 binding antibodies against functional NT antibodies [8, 18]. Many of these test kits require autoanalyzer machines that are expensive and inaccessible to laboratories in developing countries. Manual ELISAs of comparable performance may have broad utility in lower resource settings.

Therefore, we evaluated four serological assays that used different platforms against the microNT assay as the gold standard method for the detection of anti-SARS-CoV-2 antibodies. The evaluation included two autoanalyzers using three chemiluminescence-based kits (Architech IgG, Vitros IgG, and Vitros total Ig) and a manual ELISA for total Ig (Beijing Wantai). We retested all samples with IIF to validate the discordant results of the microNT and the evaluated test kits. The test sera in this study included COVID-19 prepandemic sera, NT antibody-positive sera from SARS-CoV-2 infected cases, sera of persons at risk of SARS-CoV-2 infection, and sera of travellers returning from countries experiencing SARS-CoV-2 outbreaks at the time of this study.

Methods

Ethical issues and the test sera

This study was conducted following the Declaration of Helsinki and approved by the Mahidol University Central Institutional Review Board (MU-CIRB). The study comprised 1005 convenience serum samples from four groups of participants: (1) Archival COVID-19 prepandemic serum samples collected in 2019 before the emergence of SARS-CoV-2 as the negative controls (N = 102). Participants provided informed consent for using their leftover sera in other research projects under the approval of the MU-CIRB protocol number MU-IRB 2017/180.1210; (2) Anonymized archival sera collected between 10 and 109 days after disease onset from cases with RT-PCR confirmed SARS-CoV-2 infection in 2020 as the positive controls (N = 45). These sera were kindly provided by Rajvithi Hospital. MU-CIRB waived the requirement for informed consent because the sera were the leftover samples from a clinical laboratory investigation under the Protocol number MU-COVID2020.001/2503; (3) Serum samples collected in 2020 from 366 Thai people determined by the Ministry of Public Health (MoPH) to be at increased risk of infection (health personnel, taxi drivers, and workers in entertainment venues). The informed consent in this group of participants was also provided by the MU-CIRB protocol number MU-COVID2020.001/2503. In the enrolment process, survey staff explained the purpose of the study. Participants signed an informed consent form for the interview about demographics and occupational activities as well as the donation of approximately 5–8 ml of blood using venipuncture. Blood specimens were labelled with study ID numbers. (4) Anonymous serum samples from 492 Thai citizens in state quarantines after returning to Thailand in 2020 from extended duty in countries with a known SARS-CoV-2 outbreak at the time of their return. These sera were sent from the Institute for Urban Disease Control and Prevention (IUDC), Department of Disease Control, MoPH for anti-SARS-CoV-2 antibody testing to support active case surveillance by RT-PCR. The outbreak investigation is considered a public health intervention; therefore, the Ethical Review Committee for Research related to COVID-19 Disease or Public Health Emergency, Department of Disease Control, MoPH (DDC ERC) exempted IRB approval and written informed consent due to its urgent nature. Nevertheless, participants received an explanation and gave verbal informed consent on specimen collection for laboratory diagnosis of SARS-CoV-2 infection. We included the laboratory results in our study analysis with permission from IUDC.

Study design

We evaluated the Wantai ELISA total Ig, Architect IgG, Vitros IgG, and Vitros total Ig tests against the microNT test as the reference assay. By its nature, microNT is less sensitive than binding antibody assay techniques. Therefore, IIF was used to retest specimens when the microNT and a commercial assay yielded discordant results. IIF may support the result of microNT or that of the commercial assay. The final result of a sample with discordant results followed that of the IIF assay.

Microneutralization assay

The present study conducted the microNT assay as described previously [19]. Vero cells (African green monkey kidney cells—ATCC, CCL-81) were grown in Eagle’s minimum essential medium (EMEM) (Gibco, NY) supplemented with 10% fetal bovine serum (FBS) (Gibco, NY), antibiotics, and amphotericin B. SARS-CoV-2 (hCoV-19/TH/MUMT-4/2020) was isolated and propagated in Vero cell cultures maintained in EMEM supplemented with 2% FBS. We calculated the virus concentration in the 50% tissue culture infective dose (TCID50) using the Reed–Muench method. We performed the experiments using the infectious viruses in a level 3 biosafety laboratory.

The cytopathic effect (CPE)-based microNT assay employed SARS-CoV-2 at a final concentration of 100 TCID50/100 µl and Vero cells grown in 96-well microculture plates in duplicate. The test serum was heat-inactivated at 56 °C for 30 min and serially diluted twofold starting from an initial serum dilution of 1:10 up to 1:1280. Then, 60 µl of each serum dilution was mixed with 60 µl of the test virus (200 TCID50/100 µl). After incubation at 37 °C for 1 h, 100 µl of the virus serum mixture was added into each well containing the Vero cell monolayer. The reaction plates were further incubated at 37 °C and observed daily for 3 days before the results were obtained. The neutralizing antibody titer was the highest reciprocal serum dilution that inhibited ≥ 50% CPE of the cell monolayer inoculated with the serum–virus mixture compared to the uninfected cultures.

Indirect immunofluorescence assay

Vero cell monolayers were infected with SARS-CoV-2 for 3–4 days and harvested when the infected cultures showed 50–75% CPE. The cells were scraped off the surface of the culture flask, suspended, washed, and spun. Then, the cell pellets were suspended, spotted onto microscopic glass slides, and air-dried before fixing in precooled acetone at − 20 °C for 10 min. The cell deposits were applied with rabbit monoclonal antibodies to SARS-CoV-2 N or spike S1 protein (Sino Biological, Beijing, China) and incubated at 37 °C for 60 min. After washing, goat anti-rabbit IgG (H + L)-conjugated Alexa Fluor® 488 (Invitrogen, Waltham, MA USA) was added, followed by the same staining procedure. The standard cell deposit contained approximately 50–75% antigen-positive cells.

The staining for anti-SARS-CoV-2 antibodies in human sera employed a single serum dilution of 1:10 in a phosphate-buffered saline, and polyclonal rabbit anti-human IgA, IgG, IgM, Kappa, and Lambda conjugated with fluorescein isothiocyanate (Dako, Glostrup, Denmark) in Evans blue diluent were used as the secondary antibodies. The viral antigen localized in the cytoplasm of the infected cells appeared fluorescent apple green under a fluorescence microscope (Fig. 1).

Fig. 1
figure 1

Indirect immunofluorescence (IIF) assay of SARS-CoV-2-infected Vero cells stained with A monoclonal antibody to S epitopes; B monoclonal antibody to N epitopes; and C human serum at a dilution of 1:10. This serum sample was negative by microNT but positive by Wantai ELISA, Vitros total Ig, and IIF; and D human negative serum control. The fluorescent positive cells appeared apple green in the cytoplasm of the infected cells

Wantai ELISA for SARS-CoV-2 total Ig

Sandwich ELISA (Wantai Biological Pharmacy, Beijing, China) for detecting anti-SARS-CoV-2 total antibodies was determined using a plate precoated with the RBD of the SARS-CoV-2 spike protein. The reaction consisted of two steps. First, 100 µl of the serum was incubated with the RBD at 37 °C for 30 min before washing. Second, horseradish peroxidase (HRP)-conjugated RBD was added to each reaction well and further incubated at 37 °C for 30 min before washing and adding a chromogenic solution for 15 min. The amount of the SARS-CoV-2 antibodies was proportional to the optical density values as measured under a spectrophotometer using dual wavelengths of 450 nm and 630 nm. The absorbance value (A) of the test sample was divided by the cut-off value (C.O.) to determine the A/C.O. ratio. Ratios of ≥ 1 were considered positive for SARS-CoV-2 antibodies, and ratios between 0.9 and 1.1 were considered borderline and retested.

Architect SARS-CoV-2 IgG

We used an Architect autoanalyzer (Abbott Laboratories, Inc., USA) to semiquantitatively detect anti-SARS-CoV-2 IgG. The assay is a two-step immunoassay that involves binding between the SARS-CoV-2N antigen coated on the paramagnetic microparticles and human IgG in the test sera, followed by acridinium-conjugated anti-human IgG as the secondary antibody. The chemiluminescence signals emitted were measured as relative light units (RLUs), which correlated with the amount of the specific IgG. The RLU of the sample (S) divided by the calibrator (C) yielded the index S/C ratio. Test sera with S/C ratios ≥ 1.4 were considered positive for SARS-CoV-2 IgG.

Vitros immunodiagnostics

The Vitros Immunodiagnostics autoanalyzer (Ortho-Clinical Diagnostics, Inc., U.S.A.) is a two-step immunoassay for semiquantitatively detecting anti-SARS-CoV-2 IgG or total Ig. The use of a luminogenic substrate and an electron transfer reagent that increased the emitted light level enhanced the chemiluminescence. The S1 protein precoated on the microwells reacted with human IgG or total Ig, which subsequently reacted with either HRP-conjugated mouse anti-human IgG or anti-total Ig. The intensity of the emitted light correlated with the amount of IgG binding or the total Ig in the test sera. Values of ≥ 1 were considered positive for both SARS-CoV-2 IgG and total Ig.

Statistical analysis

We determined the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of each serological assay using all 1005 serum samples as a whole and in certain subgroups, as appropriate. The study also determined the correlations between the index values of each assay and the NT antibody titer to assess whether the index levels correlated with the protective antibodies. We performed all analyses using the Stata program, version 16. R square (R2), mean, and standard deviation (SD) were analysed by GraphPad Prism version 8.4.3 for Windows (GraphPad Software, La Jolla, California, USA). The figures were drawn using GraphPad Prism version 8.4.3 for Windows (GraphPad).

Results

Evaluation of test performance using negative control sera

When run on the 102 COVID-19 prepandemic negative-control sera, all assays (microNT, Wantai ELISA total Ig, Architect IgG, Vitros IgG, and Vitros total Ig) yielded negative results, except one sample that was positive by the Vitros total Ig. Retesting of the one discordant result with IIF suggested that this sample was negative for SARS-CoV-2 antibodies.

Evaluation of test performance using positive control sera

When tested on the 45 microNT antibody-positive serum samples from SARS-CoV-2-infected cases, the Wantai ELISA and Vitros total Ig detected SARS-CoV-2 antibodies in all serum samples (sensitivity = 100%) (45/45). Architect IgG and Vitros IgG had sensitivities of 88.9% (40/45) and 82.2% (37/45), respectively. We used IIF to retest the 13 problem serum samples and found that all were positive, indicating SARS-CoV-2 infection.

Evaluation of test performance using sera of persons from the general community

Among the sera collected from the 366 individuals investigated as being at risk of SARS-CoV-2 infection in 2020, only two sera (0.55%) were true positives (Table 1), suggesting a low prevalence of infection in this population at the time of the study. Architect IgG was the only kit that did not detect either of the two positive serum samples. All other assays detected both positive specimens, showing a sensitivity of 100%. The specificity and NPV of these assays were also all above 99%. Nevertheless, based on IIF verification, each assay produced between one and three false-positive results, leading to PPVs of 40% with Wantai ELISA, 50% with Vitros IgG, 66.7% with Vitros total Ig, and 100% with microNT (Table 1).

Table 1 Test performance of each serological assay among persons investigated as being at risk of SARS-CoV-2 infection in the general population (N = 366)

Evaluation of test performance using sera of returning travellers from countries with a high prevalence of infection

Of the 492 serum samples from Thai citizens returning from foreign countries, 139 (28.3%) were positive for SARS-CoV-2 antibodies. In this higher prevalence population, the sensitivity of the assays ranged from 89.9 to 100%, and the specificities were all above 98.9%. The PPV and NPV of the assays were all over 97.2% and 96.2%, respectively, as shown in Table 2. A single serum sample tested negative by microNT, whereas it tested positive by the Wantai ELISA, Vitros total Ig, and IIF. The fluorescent-positive cells are shown in Fig. 1.

Table 2 Test performance of each serological assay in travellers returning from higher prevalence locations (N = 492)

Evaluation of sera in the overall combined samples

Of the 1005 sera in the combined sample sets, 186 (18.5%) were positive for SARS-CoV-2 antibodies. The microNT and the other four commercial kits yielded discordant results in 55 serum samples, and IIF retested them. All the microNT results, either positive or negative, were supported by IIF. An exception occurred with one sample, which was positive by Wantai ELISA and Vitros total Ig but negative by microNT, resulting in a sensitivity of 99.5%, NPV of 99.9%, and specificity and PPV of 100%. The Wantai ELISA total Ig had the highest sensitivity (100%) and was followed in order by microNT (99.5%), Vitros IgG (94.6%), Vitros total Ig (92.5%), and Architect IgG (89.3%). The specificities of all commercial kits were over 99%; the PPVs and NPVs were over 96% and 97%, respectively (Table 3).

Table 3 Test performance of each serological assay in the overall study (N = 1005)

Correlation between the NT antibody titers and the positive values of each assay

This study determined the correlations between the NT antibody titers and the index values obtained from each commercial assay. The results in Fig. 2 demonstrate a moderate degree of correlation between the microNT assay and the Architect IgG assay, which employed the N antigen (R2 = 0.7781), and between the microNT assay and the Wantai ELISA total Ig, which employed the RBD antigen (R2 = 0.7763). NT antibody titers poorly correlated with the positive values of both Vitros total Ig and Vitros IgG, which employed the S1 antigen (R2 = 0.2987 and 0.3149, respectively).

Fig. 2
figure 2

Correlation between the microNT antibody titers and the index values of each commercial assay: A Architect IgG; B Wantai ELISA total Ig assay; C Vitros IgG; and D Vitros total Ig. R square (R2), mean (–), and SD were analysed by GraphPad Prism version 8.4.3 for Windows (GraphPad Software, La Jolla, California, USA)

Discussion

In Thailand, commercial antibody testing assays must receive approval from the Food and Drug Administration (FDA) based on having a diagnostic sensitivity of ≥ 85%, diagnostic specificity of ≥ 98%, and nonspecificity or cross-reactivity of ≤ 10%, as evaluated using a serum panel from the National Institute of Health, Department of Medical Science, MoPH. The Architect IgG, Vitros IgG, and Vitros total Ig assays have been certified for use and marketing in Thailand. The Wantai ELISA total Ig has been CE (Conformite Europeene) approved for usage in the European Union, but the company has not yet submitted a proposal for its distribution in Thailand. However, Thailand participates in the World Health Organization (WHO) population-based, age-stratified seroepidemiological UNITY study. The study aims to promote standardized epidemiological, molecular, and serological methods to facilitate international comparisons and inform decisions for the COVID-19 response [20], and the WHO offered Wantai ELISA kits to Thailand for this purpose. As a result, Mahidol University received an exemption from the Thai FDA to import these ELISA kits, allowing its inclusion in this evaluation.

The present study evaluated Wantai ELISA total Ig, Architect IgG, Vitros IgG, and Vitros total Ig test kits against microNT (supplemented with IIF verification) in detecting SARS-CoV-2 antibodies. Analysis of sera from 1005 Thai citizens showed that the four serological assays evaluated were relatively comparable in terms of sensitivity (89.3–100%), specificity (99.2–100%), PPV (96.4–100%), and NPV (97.6–100%). While all assays performed well in this evaluation, the analyses stratified by subgroups provided additional insights. As might be expected, the performance of these serological assays diminished when the prevalence of the infection was low, as in the group of 366 Thai community members, among whom only two were positive. In this group, the assays either suffered from a low sensitivity (Architect IgG) or correctly identified the positive sera but produced varying false-positive results (Wantai total Ig, Vitros total Ig, and Vitros IgG). Previous studies have also reported the reduced performance of serological assays when the infection rates were low, which significantly impacted PPV values [21,22,23]. According to the orthogonal testing algorithm, individuals who were initially positive by the first-line assay required retesting with a second-line assay for confirmation [24, 25]. The commercial assays used in this study did not yield false-positive results when evaluated with COVID-19 prepandemic sera (except a single serum sample investigated by Vitros total Ig).

This study sought to determine the correlation between the NT antibody titers and the index values of results from each commercial assay. The S/C ratios of Architect IgG correlated well with the NT antibody titers (R2 = 0.778). This correlation was unexpected because Architect IgG targets the N antigen, which contains no neutralizing epitopes. This finding suggested that both microNT and Architect IgG which target different antigenic domains, can yield comparable results on the detection of previous infection. On the other hand, the NT antibody titers also correlated well with the A/C.O. values of Wantai ELISA total Ig, which targets the RBD (R2 = 0.776), the primary neutralizing domain [2]. Previous investigators demonstrated that Wantai ELISA total Ig had the highest sensitivity compared to the other four commercial serological tests (BIORAD® ELISA total Ig, EUROIMMUN® ELISA IgG, Abbott® CLIA IgG, and LIAISON® CLIA IgG) when assayed in serum samples from nonhospitalized, laboratory-confirmed SARS-CoV-2 infection [26]. After 14 days postsymptom onset, Wantai ELISA total Ig had 98% agreement with microNT [22]. Other investigators also suggested that the Wantai ELISA could be a suitable substitute for the microNT assay [18]. Our study found that the NT antibody titers correlated poorly with the index values of Vitros total Ig (R2 = 0.2987) and Vitros IgG (R2 = 0.3149), which target the S1 protein. We do not have a clear explanation for this finding, as the S1 protein contains multiple antigenic sites in both the RBD and non-RBD that can induce NT antibodies [2, 27,28,29]. Nevertheless, chemiluminescent signals set by Vitros kits were adequate to report the positive or negative results as designed by the manufacturer. Poor correlation between the level of chemiluminescent signals and neutralizing antibodies did not affect the kit quality, because the sensitivities of both Vitros IgG and total Ig were slightly higher than Architect IgG, whereas the specificities were comparable.

In virology laboratories worldwide, microNT is the “gold standard” method for evaluating other antibody assay systems because of its high specificity. MicroNT may not always detect an early infection, as it takes time for antibodies to rise to a level that can neutralize the replicating virus at a concentration of 100 TCID50. Furthermore, some infected individuals may develop a moderate level of binding antibodies without NT antibodies. It is thus possible that serological assays that measure binding antibody activities might be more sensitive than microNT, particularly in the early phase of infection. The advantage of our study was the introduction of IIF (in which the test cells contained both N and S antigens) to verify any discordant results between microNT and a binding antibody assay. We showed that the microNT assay missed one positive sample, which the Wantai ELISA and Vitros total Ig could determine with IIF verification.

Conclusions

Our study found that several commercial serological assays performed well for semiqualitatively determining the presence of anti-SARS-CoV-2 antibodies. However, in low prevalence populations, they may also produce false-positive results. High-throughput Architect and Vitros autoanalyzers may be the most appropriate for working on large sample sizes in countries that can afford the costs. The Wantai manual ELISA, while requiring more individual time and technical skill, may provide reliable results at a lower price. As the COVID-19 pandemic progresses, our findings can aid in selecting an appropriate serological assay for antibody detection to diagnose a past SARS-CoV-2 infection and estimate infection rates or the magnitude of the outbreaks. The assay of choice will ultimately depend on the study population, laboratory facilities, and feasibility.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

microNT:

Microneutralization

SARS-CoV-2:

Severe acute respiratory syndrome-coronavirus-2

COVID-19:

Coronavirus disease 2019

ELISA:

Enzyme-linked immunosorbent assay

CLIAs:

Chemiluminescence immunological assays

IIF:

Indirect immunofluorescence

PRNT:

Plaque reduction neutralization

RBD:

Receptor-binding domain

TCID50:

50% Tissue culture infective dose

PPV:

Positive predictive value

NPV:

Negative predictive value

References

  1. Zeng W, Liu G, Ma H, Zhao D, Yang Y, Liu M, et al. Biochemical characterization of SARS-CoV-2 nucleocapsid protein. Biochem Biophys Res Commun. 2020;527(3):618–23. https://doi.org/10.1016/j.bbrc.2020.04.136.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Premkumar L, Segovia-Chumbez B, Jadi R, Martinez DR, Raut R, Markmann A, et al. The receptor-binding domain of the viral spike protein is an immunodominant and highly specific target of antibodies in SARS-CoV-2 patients. Sci Immunol. 2020;5(48):eabc8413. https://doi.org/10.1126/sciimmunol.abc8413.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Centers for Disease Control and Prevention. Interim guidelines for COVID-19 antibody testing in clinical and public health settings. 2021. https://www.cdc.gov/coronavirus/2019-ncov/lab/resources/antibody-tests-guidelines.html. Updated 17 Mar 2021. Accessed 15 June 2021.

  4. Bohn MK, Loh TP, Wang C-B, Mueller R, Koch D, Sethi S, et al. IFCC Interim guidelines on serological testing of antibodies against SARS-CoV-2. Clin Chem Lab Med. 2020;58(12):2001–8. https://doi.org/10.1515/cclm-2020-1413.

    Article  CAS  PubMed  Google Scholar 

  5. Sethuraman N, Jeremiah SS, Ryo A. Interpreting diagnostic tests for SARS-CoV-2. JAMA. 2020;323(22):2249–51. https://doi.org/10.1001/jama.2020.8259.

    Article  CAS  PubMed  Google Scholar 

  6. Guo L, Ren L, Yang S, Xiao M, Chang D, Yang F, et al. Profiling early humoral response to diagnose novel coronavirus disease (COVID-19). Clin Infect Dis. 2020;71(15):778–85. https://doi.org/10.1093/cid/ciaa310.

    Article  CAS  PubMed  Google Scholar 

  7. Post N, Eddy D, Huntley C, van Schalkwyk MCI, Shrotri M, Leeman D, et al. Antibody response to SARS-CoV-2 infection in humans: a systematic review. PLoS ONE. 2020;15(12):e0244126. https://doi.org/10.1371/journal.pone.0244126.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Wu J, Liang B, Chen C, Wang H, Fang Y, Shen S, et al. SARS-CoV-2 infection induces sustained humoral immune responses in convalescent patients following symptomatic COVID-19. Nat Commun. 2021;12(1):1813. https://doi.org/10.1038/s41467-021-22034-1.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Burbelo PD, Riedo FX, Morishima C, Rawlings S, Smith D, Das S, et al. Sensitivity in detection of antibodies to nucleocapsid and spike proteins of severe acute respiratory syndrome coronavirus 2 in patients with coronavirus disease 2019. J Infect Dis. 2020;222(2):206–13. https://doi.org/10.1093/infdis/jiaa273.

    Article  CAS  PubMed  Google Scholar 

  10. GeurtsvanKessel CH, Okba NMA, Igloi Z, Bogers S, Embregts CWE, Laksono BM, et al. An evaluation of COVID-19 serological assays informs future diagnostics and exposure assessment. Nat Commun. 2020;11(1):3436. https://doi.org/10.1038/s41467-020-17317-y.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Walker GJ, Naing Z, Stella AO, Yeang M, Caguicla J, Ramachandran V, et al. SARS coronavirus-2 microneutralisation and commercial serological assays correlated closely for some but not all enzyme immunoassays. Viruses. 2021;13(2):247. https://doi.org/10.3390/v13020247.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  12. Riepler L, Rössler A, Falch A, Volland A, Borena W, von Laer D, et al. Comparison of four SARS-CoV-2 neutralization assays. Vaccines (Basel). 2020;9(1):13. https://doi.org/10.3390/vaccines9010013.

    Article  CAS  Google Scholar 

  13. Xu X, Sun J, Nie S, Li H, Kong Y, Liang M, et al. Seroprevalence of immunoglobulin M and G antibodies against SARS-CoV-2 in China. Nat Med. 2020;26(8):1193–5. https://doi.org/10.1038/s41591-020-0949-6.

    Article  CAS  PubMed  Google Scholar 

  14. Stadlbauer D, Tan J, Jiang K, Hernandez MM, Fabre S, Amanat F, et al. Repeated cross-sectional sero-monitoring of SARS-CoV-2 in New York City. Nature. 2021;590(7844):146–50. https://doi.org/10.1038/s41586-020-2912-6.

    Article  CAS  PubMed  Google Scholar 

  15. Gudbjartsson DF, Norddahl GL, Melsted P, Gunnarsdottir K, Holm H, Eythorsson E, et al. Humoral immune response to SARS-CoV-2 in Iceland. N Engl J Med. 2020;383(18):1724–34. https://doi.org/10.1056/NEJMoa2026116.

    Article  CAS  PubMed  Google Scholar 

  16. The National SARS-CoV-2 Serology Assay Evaluation Group. Performance characteristics of five immunoassays for SARS-CoV-2: a head-to-head benchmark comparison. Lancet Infect Dis. 2020;20(12):1390–400. https://doi.org/10.1016/S1473-3099(20)30634-4.

    Article  Google Scholar 

  17. Krüttgena A, Cornelissenb CG, Dreher M, Hornef M, Imöhla M, Kleinesa M. Comparison of four new commercial serologic assays for determination of SARS-CoV-2 IgG. J Clin Virol. 2020;128:104394. https://doi.org/10.1016/j.jcv.2020.104394.

    Article  CAS  Google Scholar 

  18. Weidner L, Gänsdorfer S, Unterweger S, Weseslindtner L, Drexler C, Farcet M, et al. Quantification of SARS-CoV-2 antibodies with eight commercially available immunoassays. J Clin Virol. 2020;129:104540. https://doi.org/10.1016/j.jcv.2020.104540.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Lerdsamran H, Prasertsopon J, Mungomklang A, Klinmalai C, Noisamdaeng P, Sangsiriwut K, et al. Seroprevalence of antibodies to enterovirus 71 and coxsackievirus A16 among people of various age groups in a northeast province of Thailand. Virol J. 2018;15(1):158. https://doi.org/10.1186/s12985-018-1074-8.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  20. World Health Organization. The Unity Studies: WHO sero-epidemiological investigations protocols. 2021. https://www.who.int/emergencies/diseases/novel-coronavirus-2019/technical-guidance/early-investigations#:~:text=The%20Unity%20Studies%3A%20WHO%20Sero,evidence%2Dbased%20knowledge%20for%20action.&text=generic%20epidemiological%20investigation%20protocols%20branded%20as%20UNITY%20studies. Accessed 19 June 2021.

  21. Oved K, Olmer L, Shemer-Avni Y, Wolf T, Supino-Rosin L, Prajgrod G, et al. Multi-center nationwide comparison of seven serology assays reveals a SARS-CoV-2 non-responding seronegative subpopulation. EClinicalMedicine. 2020;29:100651. https://doi.org/10.1016/j.eclinm.2020.100651.

    Article  PubMed  Google Scholar 

  22. Nicholson S, Karapanagiotidis T, Khvorov A, Douros C, Mordant F, Bond K, et al. Evaluation of 6 commercial SARS-CoV-2 serology assays detecting different antibodies for clinical testing and serosurveillance. Open Forum Infect Dis. 2021;8(7):ofab239. https://doi.org/10.1093/ofid/ofab239.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  23. Bond K, Nicholson S, Lim SM, Karapanagiotidis T, Williams E, Johnson D, et al. Evaluation of serological tests for SARS-CoV-2: implications for serology testing in a low-prevalence setting. J Infect Dis. 2020;222(8):1280–8. https://doi.org/10.1093/infdis/jiaa467.

    Article  CAS  PubMed  Google Scholar 

  24. Xu G, Emanuel AJ, Nadig S, Mehrotra S, Caddell BA, Curry SR, et al. Evaluation of orthogonal testing algorithm for detection of SARS-CoV-2 IgG antibodies. Clin Chem. 2020. https://doi.org/10.1093/clinchem/hvaa210.

    Article  PubMed  PubMed Central  Google Scholar 

  25. Centers for Disease Control and Prevention. Interim guidelines for COVID-19 antibody testing in clinical and public health settings. 2020. https://www.hsdl.org/?view&did=838930. Updated 1 Aug 2020. Accessed 19 Jun 2021.

  26. Brochot E, Demey B, Handala L, François C, Duverlie G, Castelain S, et al. Comparison of different serological assays for SARS-CoV-2 in real life. J Clin Virol. 2020;130:104569. https://doi.org/10.1016/j.jcv.2020.104569.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Brouwer PJM, Caniels TG, van der Straten K, Snitselaar JL, Aldon Y, Bangaru S, et al. Potent neutralizing antibodies from COVID-19 patients define multiple targets of vulnerability. Science. 2020;369(6504):643–50. https://doi.org/10.1126/science.abc5902.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  28. Chi X, Yan R, Zhang J, Zhang G, Zhang Y, Hao M, et al. A neutralizing human antibody binds to the N-terminal domain of the spike protein of SARS-CoV-2. Science. 2020;369(6504):650–5. https://doi.org/10.1126/science.abc6952.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  29. Keng CT, Zhang A, Shen S, Lip KM, Fielding BC, Tan THP, et al. Amino acids 1055 to 1192 in the S2 region of severe acute respiratory syndrome coronavirus S protein induce neutralizing antibodies: implications for the development of vaccines and antiviral agents. J Virol. 2005;79(6):3289–96. https://doi.org/10.1128/JVI.79.6.3289-3296.2005.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

We thank Abbott Laboratories Inc. and the Ortho-Diagnostics for donating their products and the World Health Organization for donating the Wantai ELISA kits to this study. We are grateful to the Department of Disease Control for facilitating the field investigation. We thank Chinda Kanoksinsombat for coordinating the project. We thank Tipsuda Chanmanee, Worawat Dangsakul, Anusara Jitsatja, Rumporn Kularb, Domrongrit Taonan, and Phanupong Nutthabat for their laboratory assistance. We also thank Dr. Arthur Brown for English editing.

Funding

This study received financial support from the National Research Council of Thailand (NRCT) Grant No. NRCT 11/2563 and the National Science and Technology Development Grant No. P-17-50551.

Author information

Authors and Affiliations

Authors

Contributions

HL performed experiments, analyzed and interpreted data, drafted the manuscript. AM, and SI facilitated the field studies and provided resources. PI, and NS provided samples. JP, NT, and SP performed experiments. SSa performed experiments and data analyses. KP, SSi, and NM assisted in the field investigation and specimen collection. WW supported the BSL-3 laboratory facilities. PP, the grant principal investigator, conceptualized and designed the study, supervised the laboratory work and data analysis, drafted, edited, and finalized the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Pilaipan Puthavathana.

Ethics declarations

Ethics approval and consent to participate

The Mahidol University Central Institutional Review Board (MU-CIRB) approved this study under the protocol numbers MU-IRB 2017/180.1210 for the study sera in group 1, and MU-COVID2020.001/2503 for the study sera in groups 2 and 3. Regarding the study sera in group 4, the Institute for Urban Disease Control and Prevention (IUDC), Department of Disease Control, MoPH got an exemption from the Ethical Review Committee for Research related to COVID-19 Disease or Public Health Emergency, Department of Disease Control, Ministry of Public Health (DDC ERC) for IRB approval and written informed consent on the outbreak investigation and a public health intervention due to its emergency nature. Nevertheless, participants gave verbal informed consent on specimen collection for laboratory diagnosis of SARS-CoV-2 infection.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lerdsamran, H., Mungaomklang, A., Iamsirithaworn, S. et al. Evaluation of different platforms for the detection of anti-SARS coronavirus-2 antibodies, Thailand. BMC Infect Dis 21, 1213 (2021). https://doi.org/10.1186/s12879-021-06921-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12879-021-06921-y

Keywords