Evaluation of a pulsed xenon ultraviolet disinfection system to decrease bacterial contamination in operating rooms

  • Lynn El Haddad1,
  • Shashank S. Ghantoji1,
  • Mark Stibich1, 2,
  • Jason B. Fleming3,
  • Cindy Segal4,
  • Kathy M. Ware1 and
  • Roy F. Chemaly1Email author
BMC Infectious DiseasesBMC series – open, inclusive and trusted201717:672

https://doi.org/10.1186/s12879-017-2792-z

Received: 27 June 2017

Accepted: 4 October 2017

Published: 10 October 2017

Abstract

Background

Environmental cleanliness is one of the contributing factors for surgical site infections in the operating rooms (ORs). To decrease environmental contamination, pulsed xenon ultraviolet (PX-UV), an easy and safe no-touch disinfection system, is employed in several hospital environments. The positive effect of this technology on environmental decontamination has been observed in patient rooms and ORs during the end-of-day cleaning but so far, no study explored its feasibility between surgical cases in the OR.

Methods

In this study, 5 high-touch surfaces in 30 ORs were sampled after manual cleaning and after PX-UV intervention mimicking between-case cleaning to avoid the disruption of the ORs’ normal flow. The efficacy of a 1-min, 2-min, and 8-min cycle were tested by measuring the surfaces’ contaminants by quantitative cultures using Tryptic Soy Agar contact plates.

Results

We showed that combining standard between-case manual cleaning of surfaces with a 2-min cycle of disinfection using a portable xenon pulsed ultraviolet light germicidal device eliminated at least 70% more bacterial load after manual cleaning.

Conclusions

This study showed the proof of efficacy of a 2-min cycle of PX-UV in ORs in eliminating bacterial contaminants. This method will allow a short time for room turnover and a potential reduction of pathogen transmission to patients and possibly surgical site infections.

Keywords

Operating rooms Environment cleanliness Pulsed xenon ultraviolet Between-cases mimicking

Background

About 400,000 surgical site infections (SSIs) are documented annually in the United States, with associated costs of around $21,000 per case [1, 2]. Prevalent organisms associated with SSIs, such as Staphylococcus aureus, Enterococcus species, Klebsiella spp., Pseudomonas aeruginosa, and Escherichia coli, can persist on surfaces from 1.5 h to more than 30 months [3].

Standard manual cleaning alone is not sufficient to eliminate these pathogens; only around 47% of surfaces are appropriately disinfected during between-case and end-of-day terminal manual cleaning [4]. Implementation of efficient environmental disinfection methods as a supplement to manual cleaning may aid in reducing the risk of wound contamination and subsequent infection, thus eliminating the possible transmission of pathogens to patients [5, 6].

The portable ultraviolet light germicidal device employing pulsed xenon lamps (PX-UV) has been shown to be a safe, easy-to-operate, and effective system in decreasing the number of pathogens [7]. PX-UV uses a xenon flash lamp to generate broad-spectrum, high-intensity ultraviolet light to deactivate and kill bacteria, spores, and viruses on high-touch surfaces in 5 min or less [7]. Two studies have shown that the use of PX-UV in addition to standard end-of-day manual cleaning helped reduce bacterial contamination levels on surfaces in the operating rooms (ORs) by 62% and 81% [8, 9].

Furthermore, it was shown that contamination in the OR increases with sequential cases, leading to a more contaminated environment for each subsequent patient during operative hours [9]. Hence, rapid and effective between-case cleaning could reduce environmental contamination, protecting subsequent patients during the same day of operation. While improved patient outcomes have been observed after PX-UV during nightly terminal cleaning practices [8, 9], no data are available on the impact of this technology when applied between surgical cases.

In this study, we aimed to determine the sufficient time required by the PX-UV device to reach environmental cleanliness.

Methods

This environmental sampling study was conducted at The University of Texas MD Anderson Cancer Center. The sampling occurred between the last end-of-case cleaning and the nightly standard terminal cleaning practices. Cleaning efficacy was assessed after 1, 2, and 8 min of PX-UV cycles using a PX-UV device (Xenex Disinfection Services). These cycle times were chosen based on proof-of-concept experiments conducted in the laboratory setting (data not shown). For each OR, high-touch surfaces were sampled at two distinct time points: after standard end-of-case cleaning and after PX-UV disinfection.

At the conclusion of surgical cases each day, the room was cleaned by OR staff according to standard end-of-case protocols (manual cleaning with ready-to-use germicidal wipes or diluted solution). Following this cleaning by not more than 1.5 h, samples from 5 high-touch surfaces (computer monitor, electrocautery unit, anesthesia cart, chair, and bed table controls) were collected for quantitative culturing using Tryptic Soy Agar contact plates. For non-flat surfaces, the plates were rolled so that their entire area came in contact with the high-touch surface. The ORs were then disinfected with a PX-UV device for 1, 2, or 8 min (10 rooms for each cycle time) at the head of the table, ensuring direct line of sight for the UV light for high-touch surfaces (Fig. 1). Following PX-UV disinfection, the same 5 high-touch surfaces were sampled at sites adjacent to the first sites using Tryptic Soy Agar contact plates. After 48-h incubation at 37 °C of the plates, colony counts were recorded. We sampled 30 ORs, generating 150 samples before PX-UV use and 150 samples after PX-UV. Table 1 gives a detailed description of all cases in each sampled OR (Table 1).
Fig. 1

Schematic design of an operating room showing the accurate position of the PX-UV device (purple) to ensure direct line of sight of the UV light to the high-touch surfaces (identified as a to h)

Table 1

Description of the case flow and case types of the 30 sampled operating rooms

OR

PX-UV cycle time (minutes)

Mean c.f.u. before PX-UV

Mean c.f.u. after PX-UV

Total number of cases during the day

Consecutive case type

Last case type of the day before PX-UV

1

1

0.6

0.8

3

Laparotomy; Incision and Drainage; Mastectomy

Mastectomy

2

1

4.2

1.4 ↓

2

Wide Local Excision; Neck Dissection

Neck Dissection

3

1

1.2

1.6

3

Laparoscopic Nephrectomy; Laparoscopic Adrenalectomy

Partial Nephrectomy

4

1

2.3

1.4 ↓

3

Mastectomy, Wide Local Excision

Excision of lesion on back

5

1

3.2

1.4 ↓

1

Laminectomy with stabilization

Laminectomy with stabilization

6

1

2.8

0.4 ↓

2

Nephrectomy

Diagnostic Laparostomy

7

1

6.2

2.2 ↓

2

Thyroid with Neck Dissection

Closure of enterostomy

8

1

2.2

2.6

2

Parotidectomy

Examination under Anaesthesia with biopsies

9

1

4.2

3.6 ↓

1

Removal of tibial nail hardware and complex closure

Removal of tibial nail hardware & complex closure

10

1

1.0

1.6

3

Insertion of Port-A-Cath

Insertion of Port-A-Cath

11

2

0.0

0.0 ↓

1

Partial Excision of genitalia with reconstruction

Partial Excision of genitalia with reconstruction

12

2

1.0

1.2

2

closure of enterostomy

Partial colectomy

13

2

1.8

2.2

1

Laparoscopic Gastrectomy

Laparoscopic Gastrectomy

14

2

1.4

0.6 ↓

2

Tandem and Ovoid insertion

Tandem and Ovoid insertion

15

2

1.0

0.6 ↓

1

Partial Lobectomy Liver and Hysterectomy

Partial Lobectomy Liver and Hysterectomy

16

2

1.4

0.8 ↓

2

Diagnostic Lap with Central Venous Catheter insertion

Diagnostic Laparotomy with CVC insertion

17

2

2.2

0.4 ↓

2

Experimental Laparoscopy with bowel anastomosis

Excision of groin lymph node

18

2

2.6

1.6 ↓

1

Proctectomy, Hysterectomy and reconstruction

Proctectomy, Hysterectomy and reconstruction

19

2

11.8

0.8 ↓

2

Lap Hysterectomy

Lap Hysterectomy

20

2

1.4

0.4 ↓

3

Insertion of central line

Thoracoscopy with segmental lung resection

21

8

1.2

0.6 ↓

1

Craniotomy

Craniotomy

22

8

3.2

1.0 ↓

2

Laparoscopic liver resection

Closure of enterostomy and Laparoscopic liver

23

8

3.2

0.8 ↓

2

Port-A-Cath insertion

Diagnostic Laparotomy with liver biopsies

24

8

4.0

0.6 ↓

2

Mastectomy with reconstruction

Segmental Mastectomy

25

8

10.8

1.6 ↓

1

Mastectomy with reconstruction

Mastectomy with reconstruction

26

8

1.4

0.0 ↓

3

Diagnostic Laparostomy with Hysterectomy; Incision and Drainage cyst

Cystoscopy with biopsies

27

8

0.8

0.2 ↓

2

Laparoscopic Salpingo-Oophorectomy

Diagnostic Laparotomy with biopsies

28

8

2.2

0.6 ↓

2

Thyroidectomy

Mastectomy

29

8

7.8

2.6 ↓

1

Partial colectomy with nephrectomy

Partial colectomy with nephrectomy

30

8

7.6

0.5 ↓

1

Thoracotomy with Lobectomy and Pulmonary Arterial reconstruction

Thoracotomy with Lobectomy and Pulmonary Arterial reconstruction

Abbreviations: OR, Operating Room; PX-UV, pulsed xenon ultraviolet; c.f.u., Colony-forming units. The symbol “↓” indicates a decrease in c.f.u. after PX-UV

The pre-PX-UV samples were combined for analysis to remove any variance issue. Means, medians, and ranges of colony counts were recorded at each sampling period for statistical analysis. As the data were nonparametric, a Wilcoxon rank sum test was used to examine the differences between groups.

Results

A total of 147 pre-PX-UV samples and 148 post-PX-UV samples measuring bacterial load obtained for the 5 high-touch surfaces were included in the analysis. Five plates (3 in the pre-PX-UV group and 2 in post-PX-UV groups) were discarded from the analysis as outliers because of counts that were too numerous to count (TNTC) and attributed to lab error, such as a dislodged cover plate. If included in the analysis, the outliers would have had undue leverage on the data for the intervention group that had no outliers (the 1-min group), and therefore the removal of the outliers was deemed conservative.

Table 2 depicts the changes in the colony-forming units (c.f.u.) between pre- and post-PX-UV use at different cycle times. A 1-min cycle of PX-UV did not generate a significant reduction in the level of contamination on the high-touch surfaces (P = 0.594). However, 2- and 8-min cycles showed significant reduction in the level of environmental contamination by decreasing the mean colony counts by 72.5% (P = 0.0328) and 73.1% (P = 0.0075), respectively (Table 2). A 2-min PX-UV cycle was as effective in eliminating an equal load of bacterial contamination when compared to an 8-min cycle.
Table 2

Efficacy of 1-, 2-, and 8-min PX-UV disinfection cycle times in reducing operating room contamination

Timing of sampling

Samples taken (n)

Colony count (c.f.u.)

Reductiona (%)

P-value

Mean

Min

Max

IQR

Pre PX-UV (all cycles combined)

147

3.19

0

55

3

Post 1-min PX-UV

50

1.70

0

14

2

46.7

0.5940

Post 2-min PX-UV

49

0.88

0

9

1

72.5

0.0328

Post 8-min PX-UV

49

0.86

0

7

1

73.1

0.0075

aReduction of mean colony count after PX-UV in comparison with pre-PX-UV mean colony count. PX-UV, pulsed xenon ultraviolet; IQR, Interquartile range; c.f.u., Colony-forming units; min, minimum; max, maximum

Discussion

We found PX-UV disinfection effective in reducing colony counts when performed after standard cleaning. The 2- and 8-min PX-UV cycles produced equivalent and significant reduction of level of contamination when compared to standard OR cleaning alone and were more effective than the 1-min PX-UV cycle. We conclude that a 2-min cycle optimizes efficacy and efficiency.

A recent meta-analysis of financial impact on the United States healthcare system showed that SSIs contribute to 33.7% of the overall annual cost ($9.8 billion) of healthcare-associated infections [2]. By implementing this SSI prevention approach in the OR setting, contamination in the OR could be controlled during sequential cases, leading to a decontaminated environment for subsequent patients and may have positive impact on the rate of SSIs and associated costs.

PX-UV has been successfully used to reduce or eliminate pathogens such as vancomycin-resistant enterococci, Methicillin-resistant Staphylococcus aureus (MRSA), as well as Clostridium difficile on high-touch surfaces in patient rooms [10, 11]. In fact, PX-UV combined with quaternary ammonium removed 95% of C. difficile spores compared to a 70% of spores reduction when disinfecting patient rooms with bleach [10]. Moreover, PX-UV was 16 times more effective than manual cleaning in eliminating MRSA [12] and 100% effective against VRE [11]. The efficacy of this method has also been confirmed against fungi, Bacillus anthracis, and viruses such as Ebola virus [13]. In addition, PX-UV does not damage materials in hospital settings and is not transmitted through glass windows [10].

Another method for decontaminating OR rooms between cases is the use of improved hydrogen peroxide products (IHP) such as Activated Hydrogen Peroxide (Clorox Healthcare). Even though this disinfectant is effective in reducing the contamination level to around 84% of the baseline, it presents a major limitation, i.e., manual cleaning for about 2 to 4.5 h [14]. Manual cleaning is not predictable nor optimal being dependent upon the education of the cleaning personnel and the nurses [15, 16]. In fact, when cleaning, tools such as buckets, mop heads, and wipes can rapidly become contaminated and potentially transfer pathogens to other cleaned surfaces [7]. Also, the continual and recurrent use of the same chemical disinfectant can lead to the emergence of resistant microorganisms [7]. Moreover, the time spent on manually cleaning constitutes an important drawback in ORs where rapid bed turnaround time is crucial and entails operational costs for training specialized personnel. Finally, IHP costs around $175 per room, whereas the PX-UV device costs approximately $3 per room to operate, excluding labor costs in both cases [10].

The present study was limited to 5 high-touch surfaces. Other high-touch surfaces such as floors, light switches, cabinet handles, and doorknobs could be added to future studies. Additional limitations are the somewhat small sample size used in this study and the lack of bacterial identification to the species level by our use of TSA sampling plates, which are limited in detection to aerobic bacteria only. Moreover, the impact of PX-UV use between cases on SSIs and identification of bacteria at the species level on ORs surfaces still need to be determined in future studies. Finally, an operational study that investigate the impact of the between-case use of PX-UV on OR case flow would be necessary.

Conclusions

In summary, our results suggest that supplementing standard cleaning procedures using a portable no-touch PX-UV system could be done routinely and rapidly between cases in the OR. A cycle of 2 min was sufficient in eliminating 70% or more of the bacterial load on inanimate high-touch surfaces, thus allowing short time for room turnover and potentially reducing pathogen transmission to patients and possibly SSI rates.

Abbreviations

c.f.u.: 

Colony-forming units

IHP: 

Improved hydrogen peroxide products

IQR: 

Interquartile range

max: 

Maximum

min: 

Minimum

OR: 

Operating rooms

PX-UV: 

Pulsed xenon lamps

SSIs: 

Surgical site infections

TNTC: 

Too numerous to count

Declarations

Acknowledgements

We thank Sunita Patterson of the Department of Scientific Publications, MD Anderson Cancer Center, for copyediting the manuscript.

Funding

Xenex Disinfection Services provided funding for laboratory analysis.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Authors’ contributions

LEH, SSG, and MS conducted the experiments. Xenex Disinfection Services provided assistance with study design. MS provided the statistical analyses. LEH, MS, and RFC wrote and edited the manuscript. All authors read and approved the final manuscript.

Ethics approval and consent to participate

Not Applicable

Consent for publication

Not Applicable

Competing interests

R.F.C. has received research grants from and acts as a consultant to Xenex Disinfection Services. M. S. is employed by Xenex Disinfection Services. All other authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Authors’ Affiliations

(1)
Department of Infectious Diseases, Infection Control and Employee Health, The University of Texas MD Anderson Cancer Center
(2)
Xenex Disinfection Services
(3)
Department of Gastrointestinal Oncology, H. Lee Moffitt Cancer Center & Research Institute
(4)
Department of Surgical Oncology, Division of Surgery, The University of Texas MD Anderson Cancer Center

References

  1. Shepard J, Ward W, Milstone A, Carlson T, Frederick J, Hadhazy E, et al. Financial impact of surgical site infections on hospitals; the hospital management perspective. JAMA Surg. 2013;148:907–14.View ArticlePubMedGoogle Scholar
  2. Zimlichman E, Henderson D, Tamir O, Franz C, Song P, Yamin CK, et al. Health care–associated infections; a meta-analysis of costs and financial impact on the US health care system. JAMA Intern Med. 2013;173:2039–46.View ArticlePubMedGoogle Scholar
  3. Kramer A, Schwebke I, Kampf G. How long do nosocomial pathogens persist on inanimate surfaces? A systematic review. BMC Infect Dis. 2006;6:130.View ArticlePubMedPubMed CentralGoogle Scholar
  4. Munoz-Price LS, Birnbach DJ, Lubarsky DA, Arheart KL, Fajardo-Aquino Y, Rosalsky M, et al. Decreasing operating room environmental pathogen contamination through improved cleaning practice. Infect Control Hosp Epidemiol. 2012;33:897–904.View ArticlePubMedGoogle Scholar
  5. Yavuz SS, Bicer Y, Yapici N, Kalaca S, Aydin OO, Camur G, et al. Analysis of risk factors for Sternal surgical site infection emphasizing the appropriate ventilation of the operating theaters. Infect Control Hosp Epidemiol. 2006;27:958–63.View ArticleGoogle Scholar
  6. Blanchard J. Terminal cleaning. AORN J. 2009;89:409–11.View ArticleGoogle Scholar
  7. Chemaly RF, Simmons S, Dale C Jr, Ghantoji SS, Rodriguez M, Gubb J, et al. The role of the healthcare environment in the spread of multidrug-resistant organisms: update on current best practices for containment. Ther Adv Infect Dis. 2014;2:79–90.View ArticlePubMedPubMed CentralGoogle Scholar
  8. Simmons SE, Stachowiak J, Stibich M, Croteau M. Using pulsed xenon ultraviolet to decrease contamination in operating rooms during terminal cleaning. Am J Infect Control. 2013;41:S3435.Google Scholar
  9. Fridman A, Bruno-Murtha LA, Osgood R, McAllister J. Decreasing operating room contamination of surfaces and air with pulsed xenon ultraviolet disinfection. Am J Infect Control. 2013;41:S36.View ArticleGoogle Scholar
  10. Ghantoji SS, Stibich M, Stachowiak J, Cantu S, Adachi JA, Raad II, et al. Non-inferiority of pulsed xenon ultraviolet light versus bleach versus for reducing environmental Clostridium difficile contamination on high-touch surfaces in Clostridium difficile isolation rooms. J Med Microbiol. 2015;4:191–4.View ArticleGoogle Scholar
  11. Stibich M, Stachowiak J, Tanner B, Berkheiser M, Moore L, Raad I, et al. Evaluation of a pulsed-xenon ultraviolet room disinfection device for impact on hospital operations and microbial reduction. Infect Control Hosp Epidemiol. 2011;32:286–8.View ArticlePubMedGoogle Scholar
  12. Jinadatha C, Quezada R, Huber TW, Williams JB, Zeber JE, Copeland LA. Evaluation of a pulsed-xenon ultraviolet room disinfection device for impact on contamination levels of methicillin-resistant Staphylococcus aureus. BMC Infect Dis. 2014;14:187.View ArticlePubMedPubMed CentralGoogle Scholar
  13. Stibich M, Stachowiak J. The microbiological impact of pulsed xenon ultraviolet disinfection on resistant bacteria, bacterial spore and fungi and viruses. South. Afr J Infect Dis. 2016;31:12–5.Google Scholar
  14. Wiemken TL, Curran DR, Kelley RR, Pacholski EB, Carrico RM, Peyrani P, et al. Evaluation of the effectiveness of improved hydrogen peroxide in the operating room. Am J Infect Control. 2014;42:1004–5.View ArticlePubMedGoogle Scholar
  15. Havill NL, Havill HL, Mangione E, Dumigan DG, Boyce JM. Cleanliness of portable medical equipment disinfected by nursing staff. Am J Infect Control. 2011;39:602–4.View ArticlePubMedGoogle Scholar
  16. Qureshi Z, Mohamed MH. Role of ultraviolet (UV) disinfection in infection control and environmental cleaning. Infect Disord Drug Targets. 2013;13:191–5.View ArticlePubMedGoogle Scholar