Skip to main content

Table 5 Performance of dual HIV/syphilis test for detection of HIV antibodies at POCT sites

From: External quality assessment to support the WHO ProSPeRo study for the evaluation of two dual HIV/syphilis point-of-care tests in seven countries

Site

Chembio DPP

Abbott SD BIOLINE

Agreement

False + 

False -

Indeterminate

Total

Agreement

False + 

False -

Indeterminate

Total

1

0

0

1

55

98.2

0

0

0

55

100.0

2

0

0

0

60

100.0

0

0

0

60

100.0

3

0

0

0

75

100.0

0

0

0

75

100.0

4

0

0

0

90

100.0

0

0

2

90

97.8

5

1

3

0

120

96.7

1

3

0

120

96.7

6

0

0

0

120

100.0

0

0

0

120

100.0

7

0

3

0

50

94.0

0

0

0

50

100.0

8

0

0

0

60

100.0

0

0

0

60

100.0

9

0

0

0

10

100.0

0

0

0

10

100.0

10

0

1

0

60

98.3

0

0

0

60

100.0

Total

1

7

1

700

98.7

1

3

2

700

99.1

Kappa

    

0.96

    

0.98