Table 3 Pooled sensitivity and specificity data based on manufacturer of LFD

From: A systematic review of the sensitivity and specificity of lateral flow devices in the detection of SARS-CoV-2

Type of LFD test Sample size Positive sample size LFD detected Negative sample size Number of negatives detected by LFD Sensitivity % Sensitivity 95% CI Low Sensitivity 95% CI High Specificity % Specificity 95% CI Low Specificity 95% CI High
Panbio Abbott 13,221 2566 2012 10,745 10,703 78.41 76.78% 79.96% 99.61 99.47% 99.71%
Innova 6954 197 155 6757 6735 78.68 72.44% 83.82% 99.67 99.51% 99.78%
STANDARD Q 4402 909 744 3493 3460 81.85 79.21% 84.22% 99.06 98.68% 99.33%
CORIS 1199 167 64 1032 1011 38.32 31.29% 45.88% 97.97 96.91% 98.67%
Bioeasy 856 97 87 759 708 89.69 82.05% 94.30% 93.28 91.27% 94.85%
COVID-VIRO® 572 259 233 313 313 89.96 85.70% 93.06% 100.00 98.79% 100.00%
BD Veritor 352 123 122 334 334 99.19 95.54% 99.86% 100.00 98.86% 100.00%
BIOCREDIT 310 188 81 122 122 43.09 36.21% 50.23% 100.00 96.95% 100.00%