Skip to main content

Table 3 Examination using the anti-HIV-1 Low Titer Performance Panel PRB107

From: Evaluation of Geenius HIV-1/2 Confirmatory Assay for the confirmatory and differential diagnosis of HIV-1/HIV-2 in Japan and reliability of the Geenius Reader in the diagnosis of HIV-2

(A) Geenius HIV 1/2 Confirmatory Assay
  Genscreen ULTRA HIV Ag-Ab
POS* NEG¶
Geenius HIV-1 POS* 5 0
HIV-1 IND‡ 3 0
HIV NEG¶ 6 1
Others 0 0
(B) NEW LAV BLOT I (WHO criteria)
  Genscreen ULTRA HIV Ag-Ab
POS* NEG¶
LAV I
(WHO)
POS* 0 0
IND‡ 8 0
NEG¶ 6 1
(C) NEW LAV BLOT I (CDC criteria)
  Genscreen ULTRA HIV Ag-Ab
POS* NEG¶
LAV I
(CDC)
POS* 2 0
IND‡ 6 0
NEG¶ 6 1
  1. POS*: Positive; IND‡: Indeterminate; NEG¶: Negative