Table 4 Assessment of surface contamination by cultural technique (RODAC) and ATP-bioluminescence assay. Comparison of the different sampled surfaces

From: ATP bioluminescence assay for evaluating cleaning practices in operating theatres: applicability and limitations

Sampled surfaces

TVC (RODAC) (CFU/plate)

ATP-Bioluminescence assay (RLU/100 cm2)

Geometric mean

CI 95%

Median

Range

Geometric mean

CI 95%

Median

Range

medical anaesthesia trolley

2.1

1.4–3.0

1

0–14

42.2

24.7–72.3

38

6–510

nurse’s computer touch screen

1.1

1.0–1.4

0

0–3

29.9

17.8–50.2

30

4–480

operating table

1.7

1.2–2.5

0

0–13

35.7

21.6–58.9

31

5–468

vitals monitor

1.4

1.1–1.7

0

0–3

29.6

19.7–44.2

32

5–172

anaesthetist’s computer touch screen

1.6

1.1–2.4

0

0–16

27.2

18.2–40.8

26

5–102

surgical lighting

2.6

1.6–4.1

1

0–41

60.5

37.4–98.1

42.5

13–357

Instrument table

2.4

1.7–3.4

1

0–10

18.4

12.4–27.2

13.5

4–106

  1. CI: Confidence Interval