Table 3 Original and reclassified viral load test results between the Biocentric platform and the reference assay

From: Field suitability and diagnostic accuracy of the Biocentric® open real-time PCR platform for plasma-based HIV viral load quantification in Swaziland

Lab

Reason for VL testing

Time to freezing (hours)a

VL results during first round of VL quantification

VL results after re-quantification

Reference method (log10 copies/mL-l)

Biocentric platform (log10 copies/mL-l)

Misclassification

Biocentric platform (log10 copies/mL-l)

Misclassification

2.62 log10 copies/mL

3.0 log10 copies/mL

2.62 log10 copies/mL

3.0 log10 copies/mL

Lab-1

ART

2.7

0

3.14

upward

upward

*

*

*

Lab-1

ART

3.8

0

2.65

upward

CON

0

CON

CON

Lab-1

ART

1.1

0

3.03

upward

upward

0

CON

CON

Lab-1

ART

0.9

0

4.00

upward

upward

0

CON

CON

Lab-1

ART

0.5

0

2.70

upward

CON

*

*

*

Lab-1

ART

0.3

2.25

2.92

upward

CON

*

*

*

Lab-1

ART

1.5

2.37

2.68

upward

CON

0

CON

CON

Lab-2

Pre-ART

2.2

2.48

3.34

upward

upward

2.78

upward

CON

Lab-2

ART

4.4

2.72

0

downward

CON

0

downward

CON

Lab-1

ART

1.6

2.73

0

downward

CON

*

*

*

Lab-1

ART

2.4

2.78

3.03

CON

upward

*

*

*

Lab-2

Pre-ART

2.6

2.84

0

downward

CON

2.84

CON

CON

Lab-2

Pre-ART

1.2

2.90

0

downward

CON

0

downward

CON

Lab-2

Pre-ART

1.4

2.91

0

downward

CON

3.00 (999b)

CON

upwardb

Lab-2

Pre-ART

1

3.13

0

downward

downward

0

downward

downward

Lab-2

ART

2.6

3.21

0

downward

downward

3.51

CON

CON

Lab-2

Pre-ART

3

3.31

0

downward

downward

0

downward

downward

Lab-2

Pre-ART

3.2

3.40

2.96

CON

downward

3.84

CON

CON

Lab-2

Pre-ART

4.7

3.63

2.89

CON

downward

3.60

CON

CON

Lab-2

Pre-ART

4.1

3.77

0

downward

downward

3.06

CON

CON

Lab-2

ART

0.3

3.82

0

downward

downward

3.95

CON

CON

Lab-2

Pre-ART

2.5

4.17

0

downward

downward

4.29

CON

CON

  1. *Re-quantification on the Biocentric platform was not possible as no leftover plasma samples were available due to contamination. Reclassification of test results was not performed
  2. Zero values indicate that the VL results were below the detection limit of the VL assays
  3. CON Concordant
  4. aTime from sample collection to freezing at − 20 °C before testing
  5. bDue to rounding, the 3.00 log10 copies/mL values represent a false-positive test result at the 3.0 log10 copies/mL threshold but a concordant result according to the non-log10 values