Skip to main content

Table 4 Comparison of HPV detection and typing results

From: Bioelectronic DNA detection of human papillomaviruses using eSensor™: a model system for detection of multiple pathogens

Target HPV Type(s) by Reverse Line Probe HPV Type(s) by eSensor™ Comments on eSensor™
6 6 6  
11 11, 35, 40 11, 35 HPV 40 ND1
16 16 16  
18 18 18  
26 18, 26, 42 18, 26 HPV 42 ND
31 31 31, 84 weak 84 signal
35 35 35  
40 11, 35, 40 11, 35 HPV 40 ND
42 42, 68 42, 68  
45 45 ND no globin signal
51 51 51  
52 52 52  
54 weak 54 ND HPV 54 ND
56 56 56  
58 58 58  
59 59 59  
66 66 66  
68 42, 68 42, 68  
82 82, 31, weak 51 82, 31 HPV 51 ND
83 83 83  
84 84 84  
  1. 1 ND, not detected