Table 3 Summary accuracy measures for commercial and in-house tests

From: Nucleic acid amplification tests in the diagnosis of tuberculous pleuritis: a systematic review and meta-analysis

Measure of test accuracy

Pooled summary measurea (95% CI)

Test for heterogeneityb p value

All Commercial Tests [N = 14]

  

Sensitivity

0.62 (0.43, 0.77)

<0.001

Specificity

0.98 (0.96, 0.98)

0.12

Positive Likelihood Ratio (LR+)

25.4 (16.2, 40.0)

0.46

Negative likelihood Ratio (LR-)

0.40 (0.24, 0.67)

<0.001

Diagnostic odds ratio (DOR)

80.9 (34.4, 190.4)

0.09

Amplicor [N = 4]

  

Sensitivity

0.37(0.24, 0.52)

0.18

Specificity

0.99 (0.97, 1.0)

0.12

Positive Likelihood Ratio (LR+)

52.8 (11.8, 236.2)

0.14

Negative likelihood Ratio (LR-)

0.59 (0.37, 0.95)

0.20

Diagnostic odds ratio (DOR)

105.5 (15.1, 733.9)

0.10

LCx [N = 4]

  

Sensitivity

0.72 (0.54, 0.84)

0.12

Specificity

0.97 (0.95, 0.98)

0.53

Positive Likelihood Ratio (LR+)

25.6 (14.1, 46.4)

0.57

Negative likelihood Ratio (LR-)

0.32 (0.14, 0.68)

0.09

Diagnostic odds ratio (DOR)

93.2 (35.6, 243.7)

0.38

MTD [N = 6]

  

Sensitivity

0.77 (0.62, 0.88)

<0.001

Specificity

0.98 (0.96, 0.99)

0.66

Positive Likelihood Ratio (LR+)

17.4 (6.1, 49.7)

0.26

Negative likelihood Ratio (LR-)

0.31 (0.09, 1.03)

<0.001

Diagnostic odds ratio (DOR)

72.9 (9.9, 533.4)

0.03

In-house Tests [N = 26]

  

Sensitivity

0.71 (0.63, 0.78)

<0.001

Specificity

0.93 (0.88, 0.96)

<0.001

Positive Likelihood Ratio (LR+)

9.7 (5.7, 16.6)

<0.001

Negative likelihood Ratio (LR-)

0.32 (0.23, 0.43)

<0.001

Diagnostic odds ratio (DOR)

42.4 (22.2, 81.1)

<0.001

  1. aRandom effects model; bChi-square or Fisher's exact test for heterogeneity