Skip to main content

Table 5 Diagnostic accuracy of CLIA test kit in reference to QRBD and level of agreement

From: Test concordance and diagnostic accuracy of three serological assays for detection of anti-SARS-CoV-2 antibody: result from a population-based sero-epidemiological study in Delhi

Test name

QRBD

 

Test accuracy of CLIA

(95% CI)

Total agreement

Kappa

(SE, 95% CI)

Positive

Negative

Total

CLIA

Positive

1084 (99.2)

8 (0.8)

1092 (100.0)

Sensitivity: 42.9% (41.0–44.9%)

Specificity: 99.0% (98.1–99.5%)

56.8%

0.26

(SE = 0.012)

(95% CI 0.23 -0.28)

Negative

1440 (63.6)

824 (36.4)

2264 (100.0)

  1. *SE = standard error, CI = Confidence interval