Skip to main content

Table 2 Detection limits of human NoVs GII.4 in RT-qPCR and RT-RAA assays

From: Development of a recombinase-aided amplification assay for rapid detection of human norovirus GII.4

Copies/reaction

Times of positive sample tested by two different assays for NoV (n = 8)

RT-qPCR

RT-RAA

2.67 × 107

8

8

2.67 × 106

8

8

2.67 × 105

8

8

2.67 × 104

8

8

2.67 × 103

8

7

2.67 × 102

8

6

2.67 × 101

6

1

2.67 × 100

2

0