Skip to main content

Table 2 Performance characteristics of Revogene compared to Xpert GBS LB Assays

From: Comparative study of Revogene GBS LB assay and GeneXpert GBS LB assay for the detection of group B Streptococcus in prenatal screening samples

 

Xpert

Positive

Negative

Total

Revogene

Positive

50

7

57

Negative

1

192

193

Total

51

199

250

  

95% CI

Positive Percent Agreement

98.0%

89.6 to 99.7%

Negative Percent Agreement

96.5%

92.9 to 98.3%

Overall agreement

96.8%

93.8 to 98.4%