Skip to main content

Table 5 Average timesa to ID and AST for AXDX versus SOC in hours

From: Evaluation of the impact of the Accelerate Pheno™ system on time to result for differing antimicrobial stewardship intervention models in patients with gram-negative bloodstream infections

 

AXDX (instrument time)

AXDX (time from positivity)

SOC (time from positivity)

Time Difference (SOC-AXDX)

ID

1.3

5.7

17.5

11.8

AST

6.6

11.1

47.8

36.7

  1. aThe average time for ID was calculated only with samples where ID probe results were in agreement between SOC and AXDX. The average time for AST was calculated only with samples where microbe and drug were both reported by SOC and AXDX
  2. SOC standard of care, AXDX Accelerate Pheno™ system