Table 1 Test characteristics

From: Laboratory evaluation of four HIV/syphilis rapid diagnostic tests

 

SD Bioline HIV/Syphilis Duo

DPP HIV-Syphilis Assay

Multiplo Rapid TP/HIV Antibody Test

Insti Multiplex HIV-1/HIV-2/Syphilis Antibody Test

Type of assay

Lateral flow immunochromatographic assay

Lateral flow immunochromatographic assay (dual path platform)

Vertical flow immunofiltration immunoassay (flow through)

Vertical flow immunofiltration immunoassay (flow through)

Specimen

Serum, plasma, whole blood

Serum, plasma, whole blood

Serum, plasma, whole blood

Serum, plasma, whole blood

Volume required

10 μl of serum/plasma

10 μl of serum/plasma

1 drop of whole blood/serum/plasma (35-40 μl)

50 μl of whole blood/serum/plasma

20 μl of whole blood

1 sample loop or 10 μl of whole blood

Time to results

15–20 min

15–25 min

Once all fluid is absorbed

Immediately after adding the Clarifying Solution

(+/− 3 min)

(+/− 1–2 min)

Equipment

Timer

Timer

None

None

HIV component

HIV-1/2 recombinant antigens (gp41, gp36 and HIV-1 group O)

HIV-1/2 recombinant antigens (not specified)

HIV-1/2 synthetic peptides (gp36, gp41, gp120 and HIV-1 group O)

HIV-1/2 recombinant antigens (gp41, gp36)

TP component

Recombinant antigen (17 kDa)

IgG/IgM recombinant antigen (not specified)

IgG/IgM recombinant antigens (15 kDa, 17 kDa, 47 kDa)

Recombinant fusion proteins derived from p17 and p47 domains

Steps to perform

Add the specimen.

Add the specimen to the Sample Tainer (with buffer), mix.

Add 3 drops of buffer to the device.

Add the specimen to the Sample diluent vial, mix.

Add 3 drops of diluent.

Add the specimen.

Add 2 drops to the test device at Well 1.

Place the InstantGoldCap.

Add the content of this vial to the device.

Add 4 drops of buffer to the test device at Well 2.

Add 12 drops of buffer.

Add the Colour Developer.

Remove the InstantGoldCap.

Add the Clarifying Solution.

Add 3 drops of buffer.

Result interpretation

Clear

Clear

Clear

There is no marker on the devices to indicate the position of the HIV, syphilis and control dot.