Table 2 Summary of the kit performance for the blood based kits

From: Evaluation of HIV-1 rapid tests and identification of alternative testing algorithms for use in Uganda

Rapid Test Kit

Sensitivity

Specificity

PPV

NPV

FPR

FNR

With WP as POS

After removal of WP

With WP as POS

After removal of WP

With WP as POS

After removal of WP

With WP as POS

After removal of WP

With WP as POS

After removal of WP

With WP as POS

After removal of WP

Determine

99.3

99.2

89.6

99.0

90.1

99.0

99.2

99.2

10.4

1.0

0.7

0.8

HIVSav

99.1

98.2

97.6

99.6

97.5

99.2

99.1

99.1

2.4

0.4

0.9

1.8

Acon

99.0

97.7

97.5

99.7

97.4

99.3

99.0

99.0

2.5

0.3

1.0

2.3

Unigold

98.9

98.7

98.6

99.6

98.6

99.5

98.9

98.9

1.4

0.4

1.1

1.3

SD Bioline

99.0

98.6

98.3

99.7

98.2

99.6

99.0

99.0

1.7

0.3

1.0

1.4

First Response

99.0

96.7

93.4

98.9

93.4

96.4

98.6

98.6

6.6

1.1

1.0

3.3

Carestart

99.0

97.0

92.1

98.4

92.3

95.2

99.0

99.0

7.9

1.6

1.0

3.0

Doublecheck

98.7

97.8

96.2

99.4

95.9

98.8

98.8

98.8

3.8

0.6

1.3

2.2

Statpak

98.8

98.7

98.4

99.5

98.3

99.4

98.9

98.9

1.6

0.5

1.2

1.3

Medinostics

98.9

98.1

99.1

99.8

99.1

99.6

98.9

98.9

0.9

0.2

1.1

1.9

Insti

99.2

98.9

98.2

99.8

98.9

99.7

99.2

99.2

1.1

0.2

0.8

1.1

  1. Figures represent percentages, with weak positives taken as positives and after their removal