Skip to main content

Table 2 Clinical sensitivity and specificity, predictive values of BD Veritor System for Flu AB test

From: Analytical performance of the BD veritor™ system for rapid detection of influenza virus A and B in a primary healthcare setting

 

RT-PCR

Flu rapid test

Clinical sensitivity (%)

Clinical specificity (%)

Positive predictive value (%)

Negative predictive value (%)

Influenza (A + B) in the whole study population

78 (95/122)

93 (108/116)

92 (96/104)

81 (108/134)

Influenza (A + B) in children (age < 19 years)

76 (28/37)

94 (33/35)

97 (29/31)

80 (33/41)

Influenza (A + B) in adults (age = or >19 years)

79 (67/85)

93 (75/81)

92 (67/73)

81 (75/93)

Influenza A in the whole study population (n = 55)

82 (45/55)

97 (177/183)

88 (45/51)

95 (177/187)

Influenza A in children (age < 19 years)

87 (14/16)

96 (54/56)

87 (14/16)

96 (54/56)

Influenza A in adult (= or >19 years)

79 (31/39)

97 (123/127)

89 (31/35)

94 (123/131)

Influenza B in the whole study population

74 (50/68)

99 (168/170)

96 (50/52)

90 (168/186)

Influenza B in children (age < 19 years)

67 (14/21)

100 (51/51)

100 (14/14)

88 (51/58)

Influenza B in adults (age = or >19 years)

77 (36/47)

97 (117/119)

95 (36/38)

91 (117/128)