Table 1 Accuracy metrics (point estimates and 95%CIs) of three rapid tests compared to results of HBsAg One® and ETI-MAK-4®, enzyme immunoassays

From: Evaluating HBsAg rapid test performance for different biological samples from low and high infection rate settings & populations

Manufacturer

TP

FN

TN

FP

Sensitivity

Specificity

PPV

NPV

K (CI%)

HBsAg non-reactive/HBsAg reactive (n = 393)

 Vikia HBsAg®

101

2

290

0

98.06 % (93.16–99.76)

100.00 % (98.74–100.00)

100.00 % (96.42–100.00)

99.32 % (97.54–99.92)

98.68 % (96.85–100.00)

 Imuno-Rápido HBsAg®

98

5

287

3

95.15 % (89.03–98.40)

98.97 % (97.01–99.79)

97.03 (91.58–99.38)

98.29 % (96.05–99.44)

94.7 % (91.07–98.33)

 HBsAg teste rápido®

96

7

287

3

93.20 % (86.51–97.22)

98.97 % (97.01–99.79)

96.97 % (91.39–99.37)

97.62 % (95.16–99.04)

93.34 % (89.26–97.42)

  1. Legends: TP True positive, FN False negative, TN True negative, FP False positive, PPV Positive Predictive Value, NPV Negative Predictive Value, k Kappa statistics, n number of observations (biological samples), CI confidence interval