Skip to main content

Table 2 Intra-assay and inter-assay analysis of Ct mean value of standard curves obtained with scalar dilution of MT2 cell line (from 105 to 1).

From: Human T-lymphotropic virus type 1 (HTLV-1) prevalence and quantitative detection of DNA proviral load in individuals with indeterminate/positive serological results

MT2 cell line N° copies/reaction

Ct mean values ± SD

CV

Intra-assaya

  

105

23.35 ± 0.7

3.0

104

25.91 ± 0.8

3.1

103

28.10 ± 1.0

3.5

102

31.78 ± 1.2

3.7

10

36.18 ± 1.3

3.5

1

ND

-

Inter-assay b

  

105

23.90 ± 0.7

2.9

104

26.12 ± 0.9

3.4

103

28.30 ± 1.1

3.8

102

32.10 ± 1.3

4.0

10

36.50 ± 1.4

3.8

1

ND

-

  1. SD : standard deviation; CV: coefficient of variation; ND: not detectable
  2. a For each sample, the Ct value is the average of results from three replicates.
  3. b For each sample, the Ct value is the average of results from three different experiments performed in triplicate.