Skip to main content

Table 4 Effect of cycle duration upon the degradation of Adenovirus DNA

From: Use of UV-C radiation to disinfect non-critical patient care items: a laboratory assessment of the Nanoclave Cabinet

Exposure time

Mean Cycle Threshold (CT) valuea(n = 4)

Stainless steel sheet

Ceramic tile

0 min (control)

17

18

1 min

22

22

2 min

25

27

3 min

27

27

4 min

33

34

5 min

31

45

6 min

45

45

  1. aA 3.3 CT increase equates to a 1 log reduction in detectable viral genome. A CT value of 45 is the assay end-point and when DNA is considered undetectable.