Table 1 Ability of the Nanoclave Cabinet to disinfect non-critical patient care items

From: Use of UV-C radiation to disinfect non-critical patient care items: a laboratory assessment of the Nanoclave Cabinet

 

minimum and maximum log10reduction after exposure to two 30-second UV-C cyclesa

Blood pressure gauge (n = 8)

Patient call button (n = 8)

Infusion pump (n = 5)

Tympanic thermometer (n = 9)

Oximeter (base unit) (n = 7)

Computer keyboard/mouse (n = 4)

TV remote control (n = 4)

Blood pressure cuff (n = 6)

Pathogen b

min

max

min

max

min

max

min

max

min

max

min

max

min

max

min

max

MRSA

4.40

>5.29

>4.74

>5.17

>4.94

>5.08

2.16

>5.45

>5.25

>5.48

>4.97

>5.10

>5.05

>5.32

1.93

>5.13

VRE

>5.11

>5.23

>5.05

>5.21

>4.93

>5.30

1.49

>5.44

>4.91

5.27

4.28

>5.03

4.93

>5.16

2.13

>5.00

A. baumannii

3.44

>5.54

5.32

>5.59

>5.39

>5.56

2.29

>5.64

>5.13

>5.48

4.90

>5.74

>5.33

>5.75

3.46

>5.39

Kleb. pneum

2.76

>5.19

>4.84

>5.25

4.04

5.07

1.02

>5.11

>4.33

>5.08

>5.11

>5.21

>5.05

>5.12

3.24

>5.34

  1. ainitial inoculum: 106 cfu/cm2.
  2. btest organism suspended in ¼-strength Ringer’s solution.