Skip to main content

Table 3 The comparison of reproducibility between technicians, machines and conditions changes

From: Development and implementation of the quality control panel of RT-PCR and real-time RT-PCR for avian influenza A (H5N1) surveillance network in mainland China

Comparison

RT-PCR

rRT-PCR

 

n

Kappa ± SEa

n

Kappa ± SE

Technician

   T1&T2

132

0.976 ± 0.022

129

0.908 ± 0.040

Machines

   M1&M2

88

1.000 ± 0.000

/

/

   M1&M3

88

1.000 ± 0.000

/

/

   M4&M5

/

/

72

0.870 ± 0.073

Conditions

   C1&C2

64

1.000 ± 0.000

64

1.000 ± 0.000

   C1&C3

64

1.000 ± 0.000

64

1.000 ± 0.000

   C1&C4

64

0.925 ± 0.052

64

0.804 ± 0.083

Labs

   Lab1 & Lab2

84

1.000 ± 0.000

72

1.000 ± 0.000

  1. aSE is the abbreviation of Standard Error.