Skip to main content

Table 2 the sensitivity validation of the quality control system

From: Development and implementation of the quality control panel of RT-PCR and real-time RT-PCR for avian influenza A (H5N1) surveillance network in mainland China

Variation

RT-PCR

rRT-PCR

 

AM

H5

N1

FluA

H5

Virus lysate

H5N1-1

8/8a

8/8

8/8

7/7

7/7

H5N1-2

8/8

8/8

8/8

7/7

7/7

H5N1-3

8/8

8/8

2/8

5/7

5/7

H5N1-4

3/8

1/8

1/8

1/7

2/7

H5N1-5

0/8

0/8

0/8

0/7

0/7

H5N1-6

0/8

0/8

0/8

0/7

0/7

H2O

0/8

0/8

0/8

0/7

0/7

H3N2

0/8

0/8

0/8

7/7

0/7

H1N1

0/8

0/8

0/8

7/7

0/7

FluB

0/8

0/8

0/8

0/7

0/7

In vitro transcribed RNA

H5-1

0/8

8/8

0/8

0/7

7/7

H5-2

0/8

8/8

0/8

0/7

7/7

H5-3

0/8

8/8

0/8

0/7

7/7

H5-4

0/8

7/8

0/8

0/7

3/7

H5-5

0/8

0/8

0/8

0/7

0/7

H5-6

0/8

0/8

0/8

0/7

0/7

IPC

0/8

8/8

0/8

0/7

0/7

H2O

0/8

0/8

0/8

0/7

0/7

  1. aPositive tests/total tests